颞骨研究:中颅窝入路人工耳蜗电极插入的耳蜗内创伤评估
CT显示全部电极进入耳蜗内,蜗轴抱紧组插入电极数更多(15.1 vs 12.3,P=.0001)、中位插入深度更深(14.4 mm vs 12.5 mm,P=.021);仅1例无创插入,14例(70%)为重度创伤(Eshraghi 3–4级),两种电极模型及插入方向间创伤无差异。
要点
- 中颅窝入路可完成耳蜗内插入,但不能保证电极位于鼓阶。
- 70%样本出现Eshraghi 3–4级重度耳蜗微结构创伤。
- 蜗轴抱紧电极插入更深、电极数更多,但创伤与直电极无差异。
核心结论
- 中颅窝入路手术技术可实现电极进入耳蜗内,但不保证鼓阶正确位置,并带来重要耳蜗微结构创伤风险。
主要发现
- 直电极组平均插入电极数12.3,蜗轴抱紧组15.1(U=78,P=.0001)。
- 蜗轴抱紧组中位插入深度14.4 mm,直电极组12.5 mm(U=66,P=.021)。
- 仅1例为非创伤性插入,14例(70%)为重要创伤(Eshraghi 3和4级)。
- 两种电极模型间及不同插入方向间的耳蜗内位置或创伤分级未见差异。
临床怎么用
- 提示该入路植入时需重视鼓阶定位与创伤控制,电极选择不能仅以插入深度为据。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
Objective To evaluate cochlear trauma after cochlear implant insertion through a middle fossa approach by means of histologic and imaging studies in temporal bones. Study Design Prospective cadaveric study. Setting University-based temporal bone laboratory. Subjects and Methods Twenty fresh-frozen temporal bones were implanted through a middle cranial fossa basal turn cochleostomy. Ten received a straight electrode and 10 a perimodiolar electrode. Samples were fixed in epoxy resin. Computed tomography (CT) scans determined direction, depth of insertion, and the cochleostomy to round window distance. The samples were polished by a microgrinding technique and microscopically visualized to evaluate intracochlear trauma. Descriptive and analytic statistics were performed to compare both groups. Results The CT scan showed intracochlear insertions in every bone, 10 directed to the middle/apical turn and 10 to the basal turn. In the straight electrode group, the average number of inserted electrodes was 12.3 vs 15.1 for the perimodiolar group ( U = 78, P = .0001). The median insertion depth was larger for the perimodiolar group (14.4 mm vs 12.5 mm, U = 66, P = .021). Only 1 nontraumatic insertion was achieved and 14 samples (70%) had important trauma (Eshraghi grades 3 and 4). No differences were identified comparing position or trauma grades for the 2 electrode models or when comparing trauma depending on the direction of insertion. Conclusion The surgical technique allows a proper intracochlear insertion, but it does not guarantee a correct scala tympani position and carries the risk of important trauma to cochlear microstructures.
引用
- Cisneros Lesser JC, de Brito R, Queiroz Martins GS, et al. Evaluating Intracochlear Trauma after Cochlear Implant Electrode Insertion through Middle Fossa Approach in Temporal Bones. Otolaryngol Head Neck Surg. 2017. PMID: 29088548. doi:10.1177/0194599817739837
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。