耳蜗植入中单极与幻影电极刺激的纯音掩蔽模式
研究纳入8例Advanced Bionics耳蜗植入者(共9耳)且植入耳有残余低频听力,测量单极与幻影电极刺激(电流补偿固定50%)下125–2000 Hz纯音掩蔽阈值。结果显示两种电刺激均显著提高纯音阈值,且幻影刺激产生的掩蔽多于最顶端电极单极刺激。
要点
- 幻影电极刺激(电流补偿50%)在部分耳蜗植入者中可产生比最顶端单极刺激更多的掩蔽。
- 单极与幻影电刺激均显著提高残余听力耳的纯音阈值。
- 结果提示幻影刺激可将神经刺激模式向耳蜗更顶端偏移,与更低音高感知一致。
核心结论
- 对部分耳蜗植入者,幻影电极刺激可将神经刺激模式向耳蜗更顶端偏移,与幻影刺激产生更低音高感知的报道一致。
研究设计与样本
- 8例Advanced Bionics耳蜗植入者(共9耳)且植入耳有残余低频听力参与。
- 测量125、250、500、750、1000、2000 Hz纯音在单极与幻影电极刺激(电流补偿固定50%)下的未掩蔽与掩蔽阈值,两种电掩蔽器响度平衡。
主要发现
- 单极与幻影电掩蔽器存在时纯音阈值均显著升高。
- 掩蔽模式显示幻影刺激产生的掩蔽多于最顶端电极单极刺激。
关键数值与推荐
- 幻影电极刺激电流补偿固定为50%。
- 纯音测试频率为125、250、500、750、1000、2000 Hz。
- 两种电掩蔽器进行了响度平衡。
临床怎么用
- 幻影电极刺激可能用于使部分耳蜗植入者的神经刺激模式更偏向耳蜗顶端。
局限性
- 原文未述局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: Monopolar stimulation of the most apical electrode produces the lowest pitch sensation in cochlear implants clinically. A phantom electrode that uses out-of-phase electrical stimulation between the most apical and the neighboring basal electrode can produce a lower pitch sensation than that associated with the most apical electrode. However, because of the absence of contacts beyond the apical tip of the array, the ability to assess the spread of electrical excitation associated with phantom stimulation is limited in the typical cochlear implant subject with no residual hearing. In the present study, the spread of electrical excitation associated with monopolar and phantom stimulation of the most apical electrode was assessed using electrical masking of acoustic thresholds in cochlear implant subjects with residual, low-frequency, acoustic hearing. DESIGN: Eight subjects with an Advanced Bionics cochlear implant and residual hearing in the implanted ear participated in this study (nine ears in total). Unmasked and masked thresholds for acoustic pure tones were measured at 125, 250, 500, 750, 1000, and 2000 Hz in the presence of monopolar and phantom electrode stimulation presented at the apical-most end of the array. The current compensation for phantom electrode stimulation was fixed at 50%. The two electrical maskers were loudness balanced. Differences between the unmasked and masked acoustic thresholds can be attributed to (1) the electrical stimulus-induced interference in the transduction/conduction of the acoustic signal through cochlear periphery and the auditory nerve and/or (2) masking at the level of the central auditory system. RESULTS: The results show a significant elevation in pure-tone thresholds in the presence of the monopolar and phantom electrical maskers. The unmasked thresholds were subtracted from the masked thresholds to derive masking patterns as a function of the acoustic probe frequency. The masking patterns show that phantom stimulation was able to produce more masking than that associated with the monopolar stimulation of the most apical electrode. CONCLUSION: These results suggest that for some cochlear implant subjects, phantom electrode stimulation can shift the neural stimulation pattern more apically in the cochlea, which is consistent with reports that phantom electrode stimulation produces lower pitch sensations than those associated with monopolar stimulation of the most apical electrode alone.
引用
- Saoji AA, Koka K, Litvak LM, et al. Pure-Tone Masking Patterns for Monopolar and Phantom Electrical Stimulation in Cochlear Implants. Ear Hear. 2018. PMID: 28700446. doi:10.1097/AUD.0000000000000471
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。