CT中人工耳蜗电极触点自动定位方法研究
本研究提出一种基于植入体几何先验知识的自动方法,用于在高分辨率CT中定位7例成人人工耳蜗(Cochlear Ltd.)的耳蜗内电极触点,并与两位观察者手动标记的位置对照验证。自动检测与手动定位的半径相关系数R=0.98、植入角R=1.00,半径误差0.12 mm、植入角误差1.7°,与观察者间误差相当。作者认为该方法可准确自动定位电极触点,用于研究植入位置与听力结局的关系。
要点
- 基于植入体几何先验自动定位CT中人工耳蜗电极触点
- 自动与手动定位半径R=0.98、植入角R=1.00
- 自动与手动定位误差为半径0.12 mm、植入角1.7°
核心结论
- 以植入体几何为先前知识,高分辨率CT自动分析可准确实现人工耳蜗电极触点定位。
- 该方法输出可用于更详细地研究人工耳蜗位置对听力结局的影响。
研究设计与样本
- 方法学研究,纳入7例植入Cochlear Ltd.人工耳蜗的成人患者的高分辨率CT扫描。
- 自动检测电极触点位置,并与两位经验丰富观察者标记的位置进行验证对照。
主要发现
- 自动检测电极触点位置与手动检测位置的线性相关性高,半径为R=0.98,植入角为R=1.00。
- 自动与手动定位在半径和植入角上的误差分别为0.12 mm和1.7°,与观察者间误差相当。
- 几何测量结果与人类耳蜗常见表现一致。
- 最顶端电极的平均植入角为410°(范围316°至503°)。
- 耳蜗螺旋第一圈内电极触点的平均半径为3.0 mm,第二圈其余部分的平均半径为1.7 mm。
关键数值与推荐
- 按国际共识将估计的触点位置转换至圆柱形耳蜗坐标系,测量植入角、半径和高度。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: Determining the exact location of cochlear implant (CI) electrode contacts after implantation is important, as it helps quantifying the relation between CI positioning and hearing outcome. Unfortunately, localization of individual contacts can be difficult, because the spacing between the electrode contacts is near the spatial resolution limit of high-resolution clinical computed tomography (CT) scanners. This study introduces and examines a simple, automatic method for the localization of intracochlear electrode contacts. CI geometric specifications may provide the prior knowledge that is essential to accurately estimate contact positions, even though individual contacts may not be visibly resolved. DESIGN: The prior knowledge in CI geometry is used to accurately estimate intracochlear electrode contact positions in high-resolution CT scans of seven adult patients implanted with a CI (Cochlear Ltd.). The automatically detected electrode contact locations were verified against locations marked by two experienced observers. The interobserver errors and the errors between the averaged locations and the automatically detected locations were calculated. The estimated contact positions were transformed to a cylindrical cochlear coordinate system, according to an international consensus, in which the insertion angles and the radius and elevation were measured. RESULTS: The linear correlation of the automatically detected electrode contact positions with the manually detected locations was high (R = 0.98 for the radius, and R = 1.00 for the insertion angle). The errors in radius and in insertion angle between the automatically detected locations and the manually detected locations were 0.12 mm and 1.7°. These errors were comparable to the interobserver errors. Geometrical measurements were in line with what is usually found in human cochleae. The mean insertion angle of the most apical electrode was 410° (range: 316° to 503°). The mean radius of the electrode contacts in the first turn of the cochlear spiral was 3.0 mm, and the mean radius of the remainder in the second turn was 1.7 mm. CONCLUSIONS: With implant geometry as prior knowledge, automatic analysis of high-resolution CT scans enables accurate localization of CI electrode contacts. The output of this method can be used to study the effect of CI positioning on hearing outcomes in more detail.
引用
- Bennink E, Peters JPM, Wendrich AW, et al. Automatic Localization of Cochlear Implant Electrode Contacts in CT. Ear Hear. 2017. PMID: 28379904. doi:10.1097/AUD.0000000000000438
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。