文献速递耳科与耳神经

多极刺激配置在响度与兴奋扩散方面的比较

Comparison of Multipole Stimulus Configurations With Respect to Loudness and Spread of Excitation.
PubMed 文献 · Ear Hear · 2017-01-01

研究纳入15例语后聋人工耳蜗使用者,通过心理物理实验比较部分三相脉冲、三相脉冲及相控阵等多极配置与单极参考刺激在响度与兴奋扩散上的差异。结果显示各多极配置响度增长函数相似但电流消耗差异明显,部分三相脉冲与16电极相控阵能效最高,兴奋扩散未见组间差异。

要点

  • 部分三相脉冲与16电极相控阵在多极配置中能效最高,三相脉冲能效最低。
  • 多极配置电动态范围更大,但可辨别响度级数未显著增加。
  • 兴奋扩散实验未能显示各刺激策略间存在差异。

核心结论

  • 所测试的五种多极配置响度特性相似,部分三相脉冲与16电极相控阵因能效更高,是未来临床语音编码策略测试中最有前景的候选方案。

研究设计与样本

  • 15例语后聋人工耳蜗使用者完成心理物理实验,评估兴奋扩散、响度缩放、响度阈值、响度平衡与响度辨别。
  • 比较部分三相脉冲(σ=0.75)、三相脉冲、16电极相控阵、5电极限制性相控阵和3电极限制性相控阵与单极参考刺激。

主要发现

  • 各配置响度增长函数相似,但电流消耗存在明显差异。
  • 能效最高的多极配置为部分三相脉冲,其次为16电极相控阵和5/3电极限制性相控阵,三相脉冲能效最低。
  • 多极配置的电动态范围更大,但可辨别响度级数未显著增加。
  • 兴奋扩散实验未能显示各刺激策略之间存在差异。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: Current spread is a substantial limitation of speech coding strategies in cochlear implants. Multipoles have the potential to reduce current spread and thus generate more discriminable pitch percepts. The difficulty with multipoles is reaching sufficient loudness. The primary goal was to compare the loudness characteristics and spread of excitation (SOE) of three types of phased array stimulation, a novel multipole, with three more conventional configurations. DESIGN: Fifteen postlingually deafened cochlear implant users performed psychophysical experiments addressing SOE, loudness scaling, loudness threshold, loudness balancing, and loudness discrimination. Partial tripolar stimulation (pTP, σ = 0.75), TP, phased array with 16 (PA16) electrodes, and restricted phased array with five (PA5) and three (PA3) electrodes was compared with a reference monopolar stimulus. RESULTS: Despite a similar loudness growth function, there were considerable differences in current expenditure. The most energy efficient multipole was the pTP, followed by PA16 and PA5/PA3. TP clearly stood out as the least efficient one. Although the electric dynamic range was larger with multipolar configurations, the number of discriminable steps in loudness was not significantly increased. The SOE experiment could not demonstrate any difference between the stimulation strategies. CONCLUSIONS: The loudness characteristics all five multipolar configurations tested are similar. Because of their higher energy efficiency, pTP and PA16 are the most favorable candidates for future testing in clinical speech coding strategies.

来源:Ear Hear,查看原文。版权归原作者所有。

引用

  1. Vellinga D, Briaire JJ, van Meenen DMP, et al. Comparison of Multipole Stimulus Configurations With Respect to Loudness and Spread of Excitation. Ear Hear. 2017. PMID: 28301391. doi:10.1097/AUD.0000000000000416

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。