保留听力的电声联合刺激:人工耳蜗低频截止频率对言语理解和聆听难度的影响
该研究采用受试者内重复测量设计,纳入11名成年人工耳蜗植入者(13耳),在R-SPACE半弥散餐厅噪声环境下系统改变人工耳蜗低频截止频率,评估电声重叠对言语理解的影响。结果显示,多数重叠条件下保留植入耳残余听力有显著获益,临床软件默认推荐频率很少带来最高言语识别率(13耳中仅1耳),大于临床推荐的重叠可显著改善言语理解并降低聆听努力。
要点
- 多数电声重叠条件下,保留植入耳残余听力对言语理解有显著获益。
- 临床软件默认推荐的低频截止频率在13耳中仅1耳获得最高言语识别率。
- 大于临床推荐的电声重叠可显著改善言语理解并降低聆听努力。
核心结论
- 对于保留听力的标准电极人工耳蜗植入者,声电刺激的频谱重叠在实验室餐厅噪声模拟中可显著改善言语理解并降低聆听努力。
- 电声联合刺激患者从大于当前软件拟合推荐(仅基于听阈)的电声重叠中可能获得更多获益。
- 既往研究可能未在优化电声联合刺激参数下评估结果,从而低估了保留听力带来的获益。
研究设计与样本
- 采用受试者内重复测量设计,纳入11名有经验的成年人工耳蜗植入者(13耳),植入耳和非植入耳均有功能性残余听力。
- 辅助声学带宽固定,系统改变人工耳蜗低频截止频率,测试低于和高于原「临床软件推荐」频率的截止频率。
- 在R-SPACE声音模拟系统中评估,包含来自环绕受试者头部八个扬声器的半弥散餐厅噪声。
- AzBio句子以67 dBA呈现,信噪比在+10至0 dB之间个体化调整,以使全带宽人工耳蜗单独条件下正确率约为50%至60%。
- 聆听条件包括人工耳蜗单独、双模式(人工耳蜗+对侧助听器)和双侧辅助电声联合刺激(人工耳蜗+双侧助听器)。
主要发现
- 多数重叠条件下,保留植入耳声学听力有统计学显著获益。
- 默认临床软件推荐频率很少带来最高言语识别率,13耳中仅1耳。
- 大于临床推荐的电声重叠带来言语理解的显著改善。
- 受试者使用视觉模拟量表评估所有条件下的聆听难度。
临床怎么用
- 作者认为,电声联合刺激患者从大于当前软件拟合推荐(仅基于听阈)的电声重叠中可能获得更多获益。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: The primary objective of this study was to assess the effect of electric and acoustic overlap for speech understanding in typical listening conditions using semidiffuse noise. DESIGN: This study used a within-subjects, repeated measures design including 11 experienced adult implant recipients (13 ears) with functional residual hearing in the implanted and nonimplanted ear. The aided acoustic bandwidth was fixed and the low-frequency cutoff for the cochlear implant (CI) was varied systematically. Assessments were completed in the R-SPACE sound-simulation system which includes a semidiffuse restaurant noise originating from eight loudspeakers placed circumferentially about the subject's head. AzBio sentences were presented at 67 dBA with signal to noise ratio varying between +10 and 0 dB determined individually to yield approximately 50 to 60% correct for the CI-alone condition with full CI bandwidth. Listening conditions for all subjects included CI alone, bimodal (CI + contralateral hearing aid), and bilateral-aided electric and acoustic stimulation (EAS; CI + bilateral hearing aid). Low-frequency cutoffs both below and above the original "clinical software recommendation" frequency were tested for all patients, in all conditions. Subjects estimated listening difficulty for all conditions using listener ratings based on a visual analog scale. RESULTS: Three primary findings were that (1) there was statistically significant benefit of preserved acoustic hearing in the implanted ear for most overlap conditions, (2) the default clinical software recommendation rarely yielded the highest level of speech recognition (1 of 13 ears), and (3) greater EAS overlap than that provided by the clinical recommendation yielded significant improvements in speech understanding. CONCLUSIONS: For standard-electrode CI recipients with preserved hearing, spectral overlap of acoustic and electric stimuli yielded significantly better speech understanding and less listening effort in a laboratory-based, restaurant-noise simulation. In conclusion, EAS patients may derive more benefit from greater acoustic and electric overlap than given in current software fitting recommendations, which are based solely on audiometric threshold. These data have larger scientific implications, as previous studies may not have assessed outcomes with optimized EAS parameters, thereby underestimating the benefit afforded by hearing preservation.
引用
- Gifford RH, Davis TJ, Sunderhaus LW, et al. Combined Electric and Acoustic Stimulation With Hearing Preservation: Effect of Cochlear Implant Low-Frequency Cutoff on Speech Understanding and Perceived Listening Difficulty. Ear Hear. 2017. PMID: 28301392. doi:10.1097/AUD.0000000000000418
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。