插入深度影响侧壁电极的言语感知与听力保存
回顾性分析48例成人人工耳蜗植入,所有电极完全位于鼓阶。中位AID因电极类型而异;AID越大CNC得分越高(r=0.48,P<0.001),但有残余听力者AID与低频PTA偏移相关(r=0.41,P=0.04),提示更深插入短期听力保存更差。
要点
- 所有48例Med-El电极完全位于鼓阶。
- AID越大CNC得分越高,排除有残余听力者后相关性更强。
- 有残余听力者AID越大短期低频听力损失越明显。
核心结论
- 鼓阶插入率100%;AID与CNC得分正相关且不受术后听力状态影响,但更深插入短期听力保存更差。
研究设计与样本
- 回顾性研究,纳入48例成人人工耳蜗植入(Med-El Flex 24、Flex 28、Standard电极),均有术前及术后CT。
主要发现
- Flex 24中位AID 408°(四分位范围373°-449°),Flex 28为575°(465°-584°),Standard为584°(368°-643°)。
- 术后CNC得分均值43.7%±21.9;AID与CNC正相关(r=0.48,P<0.001);排除有残余听力者后r=0.57(P<0.001)。
- 有术前残余听力者低频PTA偏移均值27 dB±14;AID与开机时低频PTA偏移相关(r=0.41,P=0.04)。
临床怎么用
- 原文未据此给出临床推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: 1) Examine angular insertion depths (AID) and scalar location of Med-El (GmbH Innsbruck, Austria) electrodes; and 2) determine the relationship between AID and audiologic outcomes controlling for scalar position. STUDY DESIGN: Retrospective review. METHODS: Postlingually deafened adults undergoing cochlear implantation with Flex 24, Flex 28, and Standard electrode arrays (Med-El) were identified. Patients with preoperative and postoperative computed tomography scans were included so that electrode location and AID could be determined. Outcome measures were 1) speech perception in the cochlear implant (CI)-only condition, and 2) short-term hearing preservation. RESULTS: Forty-eight implants were included; all electrodes (48 of 48) were positioned entirely within the scala tympani. The median AID was 408° (interquartile [IQ] range 373°-449°) for Flex 24, 575° (IQ range 465°-584°) for Flex 28, and 584° (IQ range 368°-643°) for Standard electrodes (Med-El). The mean postoperative CNC score was 43.7% ± 21.9. A positive correlation was observed between greater AID and better CNC performance (r = 0.48, P < 0.001). Excluding patients with postoperative residual hearing, a strong correlation between AID and CNC persisted (r = 0.57, P < 0.001). In patients with preoperative residual hearing, mean low-frequency pure-tone average (PTA) shift was 27 dB ± 14. A correlation between AID and low-frequency PTA shift at activation was noted (r = 0.41, P = 0.04). CONCLUSION: Favorable rates of scala tympani insertion (100%) were observed. In the CI-only condition, a direct correlation between greater AID and CNC score was noted regardless of postoperative hearing status. Deeper insertions were, however, associated with worse short-term hearing preservation. When patients without postoperative residual hearing were analyzed independently, the relationship between greater insertion depth and better performance was strengthened. LEVEL OF EVIDENCE: 4. Laryngoscope, 127:2352-2357, 2017.
引用
- O'Connell BP, Hunter JB, Haynes DS, et al. Insertion depth impacts speech perception and hearing preservation for lateral wall electrodes. Laryngoscope. 2017. PMID: 28304096. doi:10.1002/lary.26467
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。