听神经病与重度至极重度感音神经性听力损失人工耳蜗植入者的听觉表现与电刺激参数比较
该研究在三级大学附属儿科医学中心比较孤立性听神经病(AN)与感音神经性听力损失(SNHL)儿童人工耳蜗植入后平均约8年的听觉、言语表现及电刺激参数。16例AN患儿与16例SNHL患儿在安静和噪声下听觉及言语识别测试表现相当;AN组达到主观电刺激阈所需电流电荷更低、舒适响度级和动态范围更低、神经反应遥测阈值更低,且植入前后残余听力更多。
要点
- 孤立性AN与SNHL儿童人工耳蜗植入后听觉及言语识别测试表现相当。
- AN组达到主观电刺激阈所需电流电荷、舒适响度级、动态范围及神经反应遥测阈值均低于SNHL组。
- AN组植入前后残余听力均多于SNHL组。
核心结论
- 孤立性AN儿童人工耳蜗植入后听觉表现测试模式与SNHL儿童相似。
- AN组达到主观和客观电刺激阈所需电流电荷低于SNHL组,作者认为可能归因于残余神经元和毛细胞对电听觉的贡献,也可能与耳蜗植入电刺激叠加了类似声刺激的基底膜机械刺激有关。
研究设计与样本
- 研究在三级大学附属儿科医学中心开展,纳入16例孤立性AN患儿(当前年龄5至12.2岁,使用人工耳蜗至少3.4年)与16例SNHL对照患儿,按耳聋时长、植入年龄、植入体类型及单侧/双侧植入匹配。
- 所有参与者植入前后均接受广泛听觉康复,包括使用传统助听器;多数患者植入Cochlear Nucleus装置,其余为Med-El或Advanced Bionics装置。
- 电刺激参数仅在两组中Cochlear Nucleus植入者中评估,两组采用相似的映射和电测量技术。
主要发现
- AN组与SNHL组儿童在安静和噪声下听觉及言语识别测试表现相当。
- AN组学龄期进入主流教育环境的儿童多于SNHL组,但差异无统计学意义。
- 电测量显示AN组达到主观电刺激阈所需电流电荷更低、舒适响度级和动态范围更低、神经反应遥测阈值更低。
- 基于纯音听力图,AN组植入前后残余听力均多于SNHL组。
- 分析发现基底及耳蜗中部电极的T级与低频声学阈值之间呈高度正相关。
临床怎么用
- 作者认为,孤立性AN的学语前儿童在接受广泛传统助听器康复后未显示预期口语和听觉进步时,应考虑人工耳蜗植入。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: The aim of the study was to compare auditory and speech outcomes and electrical parameters on average 8 years after cochlear implantation between children with isolated auditory neuropathy (AN) and children with sensorineural hearing loss (SNHL). DESIGN: The study was conducted at a tertiary, university-affiliated pediatric medical center. The cohort included 16 patients with isolated AN with current age of 5 to 12.2 years who had been using a cochlear implant for at least 3.4 years and 16 control patients with SNHL matched for duration of deafness, age at implantation, type of implant, and unilateral/bilateral implant placement. All participants had had extensive auditory rehabilitation before and after implantation, including the use of conventional hearing aids. Most patients received Cochlear Nucleus devices, and the remainder either Med-El or Advanced Bionics devices. Unaided pure-tone audiograms were evaluated before and after implantation. Implantation outcomes were assessed by auditory and speech recognition tests in quiet and in noise. Data were also collected on the educational setting at 1 year after implantation and at school age. The electrical stimulation measures were evaluated only in the Cochlear Nucleus implant recipients in the two groups. Similar mapping and electrical measurement techniques were used in the two groups. Electrical thresholds, comfortable level, dynamic range, and objective neural response telemetry threshold were measured across the 22-electrode array in each patient. Main outcome measures were between-group differences in the following parameters: (1) Auditory and speech tests. (2) Residual hearing. (3) Electrical stimulation parameters. (4) Correlations of residual hearing at low frequencies with electrical thresholds at the basal, middle, and apical electrodes. RESULTS: The children with isolated AN performed equally well to the children with SNHL on auditory and speech recognition tests in both quiet and noise. More children in the AN group than the SNHL group were attending mainstream educational settings at school age, but the difference was not statistically significant. Significant between-group differences were noted in electrical measurements: the AN group was characterized by a lower current charge to reach subjective electrical thresholds, lower comfortable level and dynamic range, and lower telemetric neural response threshold. Based on pure-tone audiograms, the children with AN also had more residual hearing before and after implantation. Highly positive coefficients were found on correlation analysis between T levels across the basal and midcochlear electrodes and low-frequency acoustic thresholds. CONCLUSIONS: Prelingual children with isolated AN who fail to show expected oral and auditory progress after extensive rehabilitation with conventional hearing aids should be considered for cochlear implantation. Children with isolated AN had similar pattern as children with SNHL on auditory performance tests after cochlear implantation. The lower current charge required to evoke subjective and objective electrical thresholds in children with AN compared with children with SNHL may be attributed to the contribution to electrophonic hearing from the remaining neurons and hair cells. In addition, it is also possible that mechanical stimulation of the basilar membrane, as in acoustic stimulation, is added to the electrical stimulation of the cochlear implant.
引用
- Attias J, Greenstein T, Peled M, et al. Auditory Performance and Electrical Stimulation Measures in Cochlear Implant Recipients With Auditory Neuropathy Compared With Severe to Profound Sensorineural Hearing Loss. Ear Hear. 2017. PMID: 28225734. doi:10.1097/AUD.0000000000000384
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。