文献速递耳科与耳神经

人工耳蜗使用者外周与心理物理振幅调制检测的关系

Relationship Between Peripheral and Psychophysical Measures of Amplitude Modulation Detection in Cochlear Implant Users.
PubMed 文献 · Ear Hear · 2017-01-01

本研究在8例Cochlear Ltd.植入者中同步记录电诱发复合动作电位(ECAP)与心理物理调制检测阈值(MDT),发现ECAP振幅呈正弦变化且调制反应幅度(MRA)随调制频率升高而增大,MDT亦升高,两者在低频(250、500 Hz)显著相关,1000 Hz不显著,提示中枢对高调制频率处理的限制。

要点

  • 8例Cochlear Ltd.植入者中,ECAP振幅随调制刺激呈正弦变化,MRA随调制频率升高而增大(至1000 Hz无截止)。
  • MDT随调制频率升高而增大,与生理指标趋势不一致。
  • 除1000 Hz外各频率MDT与MRA相关,仅250和500 Hz达统计学显著。

核心结论

  • 所有测试调制频率均可记录到稳健的ECAP,其振幅呈正弦变化反映调制刺激的周期性。
  • MRA随调制频率升高而增大,作者将其归因于神经适应。
  • 低调制频率下较高MRA与较低心理物理阈值相关,但1000 Hz处不相关,提示中枢对调制刺激处理的限制。

研究设计与样本

  • 研究纳入8例使用Cochlear Ltd.植入体的患者。
  • 以4000 pps脉冲串、125、250、500、1000 Hz正弦振幅调制记录ECAP,并在至少一个调制周期内逐脉冲测量。
  • 在顶端、中间和基底电极上测量ECAP,并以三选一强制选择、两下一下自适应程序测量MDT,调制频率和电极相同。

主要发现

  • ECAP振幅随调制刺激呈正弦变化,调制反应幅度(MRA)定义为调制周期内ECAP最大与最小振幅之差。
  • MRA随调制频率升高而增大,至1000 Hz无 apparent cutoff。
  • MDT随调制频率升高而增大,与生理测量趋势不一致。
  • 固定调制频率下MDT与MRA相关,除1000 Hz外各频率均可见此趋势,但仅250和500 Hz调制速率达统计学显著。
  • 顶端电极的外周反应较大、心理物理阈值较低(相对基底和中间电极),可能反映更好的耳蜗健康和神经存活,证据为术前低频听力阈值较低及ECAP振幅增长函数更陡。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: This study investigates the relationship between electrophysiological and psychophysical measures of amplitude modulation (AM) detection. Prior studies have reported both measures of AM detection recorded separately from cochlear implant (CI) users and acutely deafened animals, but no study has made both measures in the same CI users. Animal studies suggest a progressive loss of high-frequency encoding as one ascends the auditory pathway from the auditory nerve to the cortex. Because the CI speech processor uses the envelope of an ongoing acoustic signal to modulate pulse trains that are subsequently delivered to the intracochlear electrodes, it is of interest to explore auditory nerve responses to modulated stimuli. In addition, psychophysical AM detection abilities have been correlated with speech perception outcomes. Thus, the goal was to explore how the auditory nerve responds to AM stimuli and to relate those physiologic measures to perception. DESIGN: Eight patients using Cochlear Ltd. Implants participated in this study. Electrically evoked compound action potentials (ECAPs) were recorded using a 4000 pps pulse train that was sinusoidally amplitude modulated at 125, 250, 500, and 1000 Hz rates. Responses were measured for each pulse over at least one modulation cycle for an apical, medial, and basal electrode. Psychophysical modulation detection thresholds (MDTs) were also measured via a three-alternative forced choice, two-down, one-up adaptive procedure using the same modulation frequencies and electrodes. RESULTS: ECAPs were recorded from individual pulses in the AM pulse train. ECAP amplitudes varied sinusoidally, reflecting the sinusoidal variation in the stimulus. A modulated response amplitude (MRA) metric was calculated as the difference in the maximal and minimum ECAP amplitudes over the modulation cycles. MRA increased as modulation frequency increased, with no apparent cutoff (up to 1000 Hz). In contrast, MDTs increased as the modulation frequency increased. This trend is inconsistent with the physiologic measures. For a fixed modulation frequency, correlations were observed between MDTs and MRAs; this trend was evident at all frequencies except 1000 Hz (although only statistically significant for 250 and 500 Hz AM rates), possibly an indication of central limitations in processing of high modulation frequencies. Finally, peripheral responses were larger and psychophysical thresholds were lower in the apical electrodes relative to basal and medial electrodes, which may reflect better cochlear health and neural survival evidenced by lower preoperative low-frequency audiometric thresholds and steeper growth of neural responses in ECAP amplitude growth functions for apical electrodes. CONCLUSIONS: Robust ECAPs were recorded for all modulation frequencies tested. ECAP amplitudes varied sinusoidally, reflecting the periodicity of the modulated stimuli. MRAs increased as the modulation frequency increased, a trend we attribute to neural adaptation. For low modulation frequencies, there are multiple current steps between the peak and valley of the modulation cycle, which means successive stimuli are more similar to one another and neural responses are more likely to adapt. Higher MRAs were correlated with lower psychophysical thresholds at low modulation frequencies but not at 1000 Hz, implying a central limitation to processing of modulated stimuli.

来源:Ear Hear,查看原文。版权归原作者所有。

引用

  1. Tejani VD, Abbas PJ, Brown CJ, et al. Relationship Between Peripheral and Psychophysical Measures of Amplitude Modulation Detection in Cochlear Implant Users. Ear Hear. 2017. PMID: 28207576. doi:10.1097/AUD.0000000000000417

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。