文献速递耳科与耳神经

感染人工耳蜗植入体取出后保留的耳蜗内电极微生物群研究

Microflora of Retained Intracochlear Electrodes from Infected Cochlear Implants.
PubMed 文献 · Otolaryngol Head Neck Surg · 2017-02-14

这项前瞻性单队列研究纳入6例难治性人工耳蜗感染患者,取出接收-刺激器和耳蜗外电极治疗感染,耳蜗内电极保留至再植入时取出,用培养和16S DNA基因测序分析微生物。结果显示6例耳蜗内电极培养均未检出微生物,2例测序检出细菌、1例检出真菌;6例再植入后1例发生二次感染(原发和二次感染均为金黄色葡萄球菌)。

要点

  • 6例感染人工耳蜗的耳蜗内电极培养均未检出微生物。
  • 基因测序显示2例耳蜗内电极有细菌、1例有真菌。
  • 感染耳蜗外植入体与保留耳蜗内电极的微生物无相关性。

核心结论

  • 感染人工耳蜗的耳蜗内电极携带的微生物负荷低。
  • 取出感染人工耳蜗时保留耳蜗内电极似乎污染替换人工耳蜗的风险低。

研究设计与样本

  • 前瞻性单队列研究,纳入6例难治性人工耳蜗感染患者。
  • 取出接收-刺激器和耳蜗外电极以治疗感染,耳蜗内电极在(延迟)再植入时取出。
  • 植入体标本进行微生物培养和16S DNA基因测序分析。

主要发现

  • 金黄色葡萄球菌是最常检出的微生物。
  • 6例患者的耳蜗内电极培养均未检出微生物。
  • 2例耳蜗内电极经基因测序检出细菌物种,1例检出真菌物种。
  • 感染耳蜗外植入体与保留耳蜗内电极上的微生物之间无相关性。
  • 所有受试者在感染消退后接受再植入,6例中1例再植入后发生二次感染,原发和二次感染均为金黄色葡萄球菌。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

Objectives Cochlear implant infections may be refractory to medical management and require device removal with subsequent reimplantation. During device removal, the intracochlear electrode array is commonly left in place to prevent obliteration of the cochlear lumen. If the electrode is colonized with pathogens, this risks contaminating the replacement implant. In this study, we compare the microorganisms detected on infected cochlear implants against those on the retained electrode using culture and microbial gene-sequencing techniques. Study Design Prospective single-cohort study. Setting Tertiary medical center. Subjects and Methods Six patients with refractory cochlear implant infections had the receiver-stimulator and extracochlear electrode removed to facilitate treatment of the infection. The intracochlear electrode was removed at (delayed) reimplantation. Implant specimens were analyzed by microbial culture and 16S DNA gene sequencing. Results Staphylococcus aureus was the organism most commonly identified. None of the 6 patients' intracochlear electrodes yielded microbes by culture. Two intracochlear electrodes revealed bacterial species, and 1 revealed fungal species by gene sequencing. There was no correlation between the microbes on the infected extracochlear implants and the retained intracochlear electrodes. All subjects underwent reimplantation after resolution of their infections. One of 6 subjects developed a second infection after reimplantation, with S aureus in the primary and secondary infections. Conclusions The intracochlear electrodes of infected cochlear implants carry a low microbial burden. Preserving intracochlear electrodes upon removal of infected cochlear implants appears to carry a low risk of contaminating a replacement cochlear implant.

来源:Otolaryngol Head Neck Surg,查看原文。版权归原作者所有。

引用

  1. Varadarajan VV, Dirain CO, Antonelli PJ, et al. Microflora of Retained Intracochlear Electrodes from Infected Cochlear Implants. Otolaryngol Head Neck Surg. 2017. PMID: 28195822. doi:10.1177/0194599817693228

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。