通过听觉皮层诱发电位客观识别正常听力者模拟的人工耳蜗设置
研究向12名正常听力者呈现声码器处理的辅音,模拟四种人工耳蜗刺激速率与最大频谱数组合,记录皮层事件相关电位。行为学表现达天花板效应,常规单变量分析未显示强调制,但多变量判别分析构建的神经混淆矩阵识别出1800/8设置能最大化神经层面区分语音,信息传递分析证实该设置最优。
要点
- 行为学识别率与反应时在四种模拟设置间无差异,均达天花板水平。
- N1和P2振幅/潜伏期的方差分析未显示强调制。
- 基于神经指标组合的多变量判别分析识别出1800/8设置最优。
核心结论
- 作者认为,头皮记录的事件相关电位可能有助于从封闭的参数选项中确定最优信号处理设置,并辅助人工耳蜗装置的客观拟合。
研究设计与样本
- 研究向12名正常听力者呈现声码器处理的辅音(aKa、aSHa、aNa),模拟四种人工耳蜗刺激速率与最大频谱数组合:1800 pps/20、1800/8、500/20、500/8,同时记录神经电位。
主要发现
- 语音识别与反应时在最大频谱数或刺激速率变化时无差异,行为学表现达天花板效应。
- N1和P2振幅/潜伏期的常规单变量方差分析未显示跨人工耳蜗处理语音条件的强调制。
- 基于神经指标组合的多变量判别分析构建神经混淆矩阵,识别出1800/8设置能在神经层面最大化区分语音令牌。
- 信息传递分析证实1800/8设置在听者神经和知觉反应中最优传递信息。
临床怎么用
- 作者认为,转化为实际植入患者后,头皮记录的事件相关电位可能有助于从封闭参数选项中确定最优信号处理设置,并辅助人工耳蜗装置的客观拟合。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: Providing cochlear implant (CI) patients the optimal signal processing settings during mapping sessions is critical for facilitating their speech perception. Here, we aimed to evaluate whether auditory cortical event-related potentials (ERPs) could be used to objectively determine optimal CI parameters. DESIGN: While recording neuroelectric potentials, we presented a set of acoustically vocoded consonants (aKa, aSHa, and aNa) to normal-hearing listeners (n = 12) that simulated speech tokens processed through four different combinations of CI stimulation rate and number of spectral maxima. Parameter settings were selected to feature relatively fast/slow stimulation rates and high/low number of maxima; 1800 pps/20 maxima, 1800/8, 500/20 and 500/8. RESULTS: Speech identification and reaction times did not differ with changes in either the number of maxima or stimulation rate indicating ceiling behavioral performance. Similarly, we found that conventional univariate analysis (analysis of variance) of N1 and P2 amplitude/latency failed to reveal strong modulations across CI-processed speech conditions. In contrast, multivariate discriminant analysis based on a combination of neural measures was used to create "neural confusion matrices" and identified a unique parameter set (1800/8) that maximally differentiated speech tokens at the neural level. This finding was corroborated by information transfer analysis which confirmed these settings optimally transmitted information in listeners' neural and perceptual responses. CONCLUSIONS: Translated to actual implant patients, our findings suggest that scalp-recorded ERPs might be useful in determining optimal signal processing settings from among a closed set of parameter options and aid in the objective fitting of CI devices.
引用
- Lee S, Bidelman GM. Objective Identification of Simulated Cochlear Implant Settings in Normal-Hearing Listeners Via Auditory Cortical Evoked Potentials. Ear Hear. 2017. PMID: 28125444. doi:10.1097/AUD.0000000000000403
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。