两种电极类型引起的人工耳蜗内压力变化:人工模型实验
该前瞻性对照研究在人工耳蜗模型上比较直电极与蜗轴电极植入时的耳蜗内液体压力变化,发现蜗轴高体积电极引起的压力峰值和振幅均显著高于直电极。
要点
- 蜗轴电极压力峰值1.12±0.15 mmHg,显著高于直电极0.86±0.05 mmHg(P=.006)。
- 蜗轴电极压力振幅0.38±0.07 mmHg,显著高于直电极0.09±0.07 mmHg(P<.001)。
- 电极体积对人工耳蜗植入过程中耳蜗内压力变化有显著影响。
核心结论
- 在人工模型实验中客观证明电极体积对人工耳蜗植入过程中耳蜗内压力变化有显著影响。
研究设计与样本
- 前瞻性对照研究,在三级转诊中心的人工耳蜗模型上于顶区放置压力传感器进行实验。
主要发现
- 使用直电极和蜗轴电极阵列,以相同植入速度进行5次植入,并在相同时间内进行5次植入。
- 蜗轴高体积电极引起的耳蜗内液体压力变化显著大于直电极。
- 蜗轴电极压力峰值为1.12±0.15 mmHg,直电极为0.86±0.05 mmHg(P=.006)。
- 蜗轴电极压力振幅为0.38±0.07 mmHg,直电极为0.09±0.07 mmHg(P<.001)。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
Objective To preserve residual hearing in cochlear implant surgery, the electrode design has been refined, and an atraumatic insertion has become one aspect of cochlear implant research. Previous studies have described the effect of insertion speed and opening of the round window membrane on intracochlear pressure changes. The aim of our current study was to observe intracochlear pressure changes due to different cochlear implant electrodes in an artificial cochlear model with stable surrounding factors. Study Design Prospective controlled study. Setting Tertiary referral center. Subjects and Methods The experiments were performed in an artificial cochlear model with a pressure sensor in the apical area. With straight and perimodiolar electrode arrays, 5 insertions with the same insertion speed and 5 insertions over the same time were performed. Results With the perimodiolar high-volume electrode, significantly greater intracochlear fluid pressure changes were observed than with the straight electrode. Compared with the straight electrode, the perimodiolar electrode induces significantly higher pressure peaks (1.12 ± 0.15 vs 0.86 ± 0.05 mm Hg, P = .006) and significantly higher amplitudes (0.38 ± 0.07 vs 0.09 ± 0.07 mm Hg, P < .001). Conclusion The reliable preservation of residual hearing is an important multifactorial challenge in modern cochlear implant surgery. Insertion speed, handling, and electrode design are known to influence the preservation of residual hearing. In our artificial model experiments, we could prove objectively that the volume of the electrodes has a significant influence on the intracochlear pressure changes during cochlear implantation.
引用
- Mittmann P, Mittmann M, Ernst A, et al. Intracochlear Pressure Changes due to 2 Electrode Types: An Artificial Model Experiment. Otolaryngol Head Neck Surg. 2016. PMID: 28025904. doi:10.1177/0194599816684104
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。