人工耳蜗植入者电诱发镫骨肌反射阈值测量的优化

Optimizations for the Electrically-Evoked Stapedial Reflex Threshold Measurement in Cochlear Implant Recipients.
PubMed 文献 · Ear Hear · 2017-01-01

该研究采用重复测量设计,在23名成人人工耳蜗植入者中比较226、678和1000 Hz三种探测音频率对电诱发镫骨肌反射阈值(eSRT)的影响。结果显示226 Hz探测音测得的平均eSRT显著高于678和1000 Hz,且4名受试者仅用226 Hz无法测得eSRT而用678和1000 Hz均可测得。作者认为226 Hz标准探测音不理想,高频探测音可获得更低的eSRT水平。

要点

  • 226 Hz探测音的平均eSRT为174电荷单位,显著高于678 Hz的167和1000 Hz的165。
  • 678与1000 Hz探测音的平均eSRT之间无统计学显著差异。
  • 23名受试者中4名用226 Hz探测音无法测得eSRT,但用678和1000 Hz均可测得。

核心结论

  • 使用标准226 Hz探测音测量eSRT并不理想;与226 Hz相比,使用较高频率探测音(678或1000 Hz)可获得更低的eSRT水平。
  • 需要进一步工作以理解这些变化在人工耳蜗编程中的临床意义。

研究设计与样本

  • 采用重复测量设计,在23名成人人工耳蜗植入者中检查3种探测音频率(226、678和1000 Hz)对eSRT的影响。

主要发现

  • 使用226 Hz探测音测得的平均eSRT为174电荷单位,显著高于使用678 Hz探测音的167电荷单位和1000 Hz探测音的165电荷单位。
  • 678 Hz与1000 Hz探测音的平均eSRT之间无统计学显著差异。
  • 23名受试者中20名使用678或1000 Hz探测音时eSRT的电荷单位水平低于226 Hz探测音。
  • 2名受试者使用678或1000 Hz探测音测得的eSRT水平与226 Hz探测音相等。
  • 仅1名受试者使用226 Hz探测音测得的eSRT电荷单位水平低于678和1000 Hz探测音。

关键数值与推荐

  • 23名受试者中4名使用226 Hz探测音无法测得eSRT,但所有受试者使用678和1000 Hz探测音均可测得eSRT。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 需要进一步工作以理解这些变化在人工耳蜗编程中的临床意义。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: The electrically-evoked stapedial reflex threshold (eSRT) has proven to be useful in setting upper stimulation levels of cochlear implant recipients. However, the literature suggests that the reflex can be difficult to observe in a significant percentage of the population. The primary goal of this investigation was to assess the difference in eSRT levels obtained with alternative acoustic admittance probe tone frequencies. DESIGN: A repeated-measures design was used to examine the effect of 3 probe tone frequencies (226, 678, and 1000 Hz) on eSRT in 23 adults with cochlear implants. RESULTS: The mean eSRT measured using the conventional probe tone of 226 Hz was significantly higher than the mean eSRT measured with use of 678 and 1000 Hz probe tones. The mean eSRT were 174, 167, and 165 charge units with use of 226, 678, and 1000 Hz probe tones, respectively. There was not a statistically significant difference between the average eSRTs for probe tones 678 and 1000 Hz. Twenty of 23 participants had eSRT at lower charge unit levels with use of either a 678 or 1000 Hz probe tone when compared with the 226 Hz probe tone. Two participants had eSRT measured with 678 or 1000 Hz probe tones that were equal in level to the eSRT measured with a 226 Hz probe tone. Only 1 participant had an eSRT that was obtained at a lower charge unit level with a 226 Hz probe tone relative to the eSRT obtained with a 678 and 1000 Hz probe tone. CONCLUSIONS: The results of this investigation demonstrate that the use of a standard 226 Hz probe tone is not ideal for measurement of the eSRT. The use of higher probe tone frequencies (i.e., 678 or 1000 Hz) resulted in lower eSRT levels when compared with the eSRT levels obtained with use of a 226 probe tone. In addition, 4 of the 23 participants included in this study did not have a measureable eSRT with use of a 226 Hz probe tone, but all of the participants had measureable eSRT with use of both the 678 and 1000 Hz probe tones. Additional work is required to understand the clinical implication of these changes in the context of cochlear implant programming.