人工耳蜗植入者残余听力中的耳蜗电图:与听力学阈值的比较

Electrocochleography in Cochlear Implant Recipients With Residual Hearing: Comparison With Audiometric Thresholds.
PubMed 文献 · Ear Hear · 2017-01-01

在21例有残余听力的Advanced Bionics人工耳蜗植入者中,于125至4000 Hz测量纯音听阈与耳内电极记录的耳蜗电图波形,差响应见于全部21耳、和响应仅见于18耳;差响应与和响应阈值均与听阈强相关(r=0.87,n=150;r=0.82,n=72),平均差值分别为-3.2(±9.0)dB与-14(±11)dB。

要点

  • 差响应见于全部21耳,和响应波形仅见于18耳。
  • 差响应阈值与听阈强相关(r=0.87,n=150),平均差值-3.2(±9.0)dB。
  • 和响应阈值与听阈强相关(r=0.82,n=72),平均差值-14(±11)dB。

核心结论

  • 耳蜗电图阈值或可作为无法进行行为听力学测试的人工耳蜗植入者评估残余听力的有用指标。

研究设计与样本

  • 研究纳入21例植入耳有残余听力的Advanced Bionics人工耳蜗植入者,在125至4000 Hz测试频率测量纯音听阈与耳蜗电图波形,并从两个交替刺激相位计算差响应与和响应。

主要发现

  • 差响应见于全部21耳,和响应波形仅见于18耳。
  • 差响应阈值与听阈强相关(r=0.87,n=150)。
  • 和响应阈值与听阈强相关(r=0.82,n=72)。
  • 差响应与听阈的平均差值为-3.2(±9.0)dB,和响应与听阈的平均差值为-14(±11)dB。
  • 37次测量中有4次差响应测得于无行为阈值存在的频率。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: To determine whether electrocochleography (ECoG) thresholds, especially cochlear microphonic and auditory nerve neurophonic thresholds, measured using an intracochlear electrode, can be used to predict pure-tone audiometric thresholds following cochlear implantation in ears with residual hearing. DESIGN: Pure-tone audiometric thresholds and ECoG waveforms were measured at test frequencies from 125 to 4000 Hz in 21 Advanced Bionics cochlear implant recipients with residual hearing in the implanted ear. The "difference" and "summation" responses were computed from the ECoG waveforms measured from two alternating phases of stimulation. The interpretation is that difference responses are largely from the cochlear microphonic while summating responses are largely from the auditory nerve neurophonic. The pure-tone audiometric thresholds were also measured with same equipment used for ECoG measurements. RESULTS: Difference responses were observed in all 21 implanted ears, whereas summation response waveforms were observed in only 18 ears. The ECoG thresholds strongly correlated (r = 0.87, n = 150 for difference response; r = 0.82, n = 72 for summation response) with audiometric thresholds. The mean difference between the difference response and audiometric thresholds was -3.2 (±9.0) dB, while the mean difference between summation response and audiometric thresholds was -14 (±11) dB. In four out of 37 measurements, difference responses were measured to frequencies where no behavioral thresholds were present. CONCLUSIONS: ECoG thresholds may provide a useful metric for the assessment of residual hearing in cochlear implant subjects for whom it is not possible to perform behavioral audiometric testing.

来源:Ear Hear,查看原文。版权归原作者所有。