文献速递耳科与耳神经

cVEMP潜伏期与振幅在患者中的信度及测量误差:Meta分析与重测信度

Agreement, Reliability, and Measurement Error of Cervical Vestibular Evoked Myogenic Potentials in Patients: II. Meta-Analysis and Reliability of Latency and Amplitude.
PubMed 文献 · Ear Hear · 2026-08-13

本研究在27例成人双侧极重度感音神经性聋(人工耳蜗候选者)中开展cVEMP重测信度研究,并嵌入Meta分析汇总既往信度证据。结果显示cVEMP潜伏期与振幅参数信度中等、置信区间宽、测量误差明显,MDC95%提示需数毫秒差异才超出随机变异,当前测量流程尚不足以支持个体化临床监测。

要点

  • cVEMP潜伏期与振幅重测信度仅为中等,个体内变异大
  • p13潜伏期MDC95%为3.5毫秒,需数毫秒差异才超出随机误差
  • 当前cVEMP测量流程尚不支持个体化临床诊断与监测

核心结论

  • 临床重测数据与汇总证据一致显示cVEMP潜伏期和振幅测量信度中等,置信区间宽且测量误差显著。
  • cVEMP潜伏期MDC95%提示需数毫秒差异才能超出重复测试的随机变异,限制了个体监测的可解释性。
  • 当前cVEMP测量流程未达到个体患者广泛临床应用所需的信度,其诊断或监测角色应有限制,直至方法学改进可证明重复性提升。

研究设计与样本

  • 在成人双侧极重度听力损失患者中开展重测信度研究,采用相同短纯音刺激(500 Hz,2毫秒时长)和记录参数进行两次独立cVEMP检查。
  • 共27例患者纳入完整分析,采用随机效应模型估计受试者、侧别和测试会话的方差成分。
  • 嵌入随机效应Meta分析,检索既往cVEMP信度研究,将报告的ICC转换为单次测量形式并计算置信区间。

主要发现

  • 阈值水平试验中,ICC(95%CI)为p13=0.61(0.41至0.81)、n23=0.32(0.10至0.54)、IPI=0.29(0.07至0.51)、cAMP=0.16(0.00至0.36)。
  • 对应的SEM分别为1.27、1.87、1.47和0.16毫秒,MDC95%分别为3.5、5.2、4.1和0.45毫秒。
  • Meta分析共识别7项研究,每个结局最多5项研究纳入分析;合并随机效应估计(95%CI)为p13=0.52(0.38至0.64)、n23=0.57(0.39至0.71)、IPI=0.47(0.08至0.73)、cAMP=0.63(0.44至0.76)。
  • 研究间异质性因结局而异,从可忽略到高不等,反映方法学和分析差异;临床数据与合并数据一致,均显示仅中等信度。

临床怎么用

  • 解读个体cVEMP潜伏期变化时需考虑MDC95%阈值,差异未超过随机变异范围不宜视为真实改变。
  • cVEMP结局可用于群体水平或研究比较,但需明确考虑测量精度。
  • 在方法学改进证明重复性提升前,cVEMP作为个体诊断或监测工具应保持限制。

局限性

  • 研究样本为成人双侧极重度听力损失患者,结论外推至其他人群需谨慎。
  • Meta分析中每个结局纳入研究数量有限(最多5项),且研究间异质性从可忽略到高不等。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: This study assessed the reliability and measurement error of cervical vestibular evoked myogenic potential (cVEMP) latency and amplitude parameters in cochlear implant candidates and synthesized previous reliability evidence through an embedded meta-analysis. Together with a preceding paper on cVEMP activation and threshold reliability, this investigation extends the evaluation of clinically used cVEMP measurement properties to latency and amplitude outcomes. DESIGN: A test-retest study was conducted in adults with profound bilateral hearing loss. Each participant underwent two independent cVEMP examinations using identical tone-burst stimulation (500 Hz, 2 msec duration) and recording parameters. Reliability was assessed for the p13 and n23 latencies, the inter-peak interval (IPI), and the corrected peak-to-peak amplitude (cAMP). Variance components were estimated using random-effects models accounting for subject, side, and test session. Single-measure, absolute-agreement intraclass correlation coefficients (ICCs), SEs of measurement (SEM), and minimal detectable change (MDC95%) were derived from the variance components. A random-effects meta-analysis of previous cVEMP reliability studies was conducted after transformation of reported ICCs to single-measure form and calculation of confidence intervals. RESULTS: Twenty-seven patients contributed to the full analysis. For threshold-level trials, ICCs (95% confidence interval) were: p13 = 0.61 (0.41 to 0.81), n23 = 0.32 (0.10 to 0.54), IPI = 0.29 (0.07 to 0.51), and cAMP = 0.16 (0.00 to 0.36). Corresponding SEMs were 1.27, 1.87, 1.47, and 0.16 msec, with MDC95% values of 3.5, 5.2, 4.1, and 0.45 msec, respectively. The wide confidence intervals and SEMs revealed considerable within-subject variability across sessions. For the meta-analysis, seven studies were identified, and up to five contributed to each outcome-specific analysis because sufficient methodological clarity permitted conversion to single-measure ICCs. Pooled random-effects estimates (95% confidence interval) were: p13 = 0.52 (0.38 to 0.64), n23 = 0.57 (0.39 to 0.71), IPI = 0.47 (0.08 to 0.73), and cAMP = 0.63 (0.44 to 0.76). Between-study heterogeneity varied across outcomes, ranging from negligible to high, reflecting methodological and analytical differences. Despite these differences, the clinical and pooled data were concordant, each showing only a moderate degree of reliability. CONCLUSIONS: Across the clinical test-retest data and the pooled evidence, cVEMP latency and amplitude measures showed moderate reliability with wide confidence intervals and notable measurement error. The MDC95% values for cVEMP latencies indicate that differences of several milliseconds are required to exceed random variability expectable from test repetition, limiting their interpretability for individual monitoring. Considered together with the meta-analytic findings and the preceding results on cVEMP activation and threshold detection, these findings suggest that current cVEMP measurement procedures do not achieve the reliability required for widespread clinical use in individual patients. Although cVEMP outcomes may be informative for group-level or research comparisons when measurement precision is explicitly considered, their role as diagnostic or monitoring tools should remain limited until methodological refinements demonstrably improve reproducibility.