前向压力校准对耳声发射序列监测听力的影响
本研究对20名成人每2个月一次、持续1年进行同一耳内测试,比较传统耳内SPL校准的DPOAE与FPL校准的扫频DPOAE及SFOAE的重测信度。结果显示FPL校准的扫频DPOAE重测信度显著优于传统DPOAE,且在中高频优于扫频SFOAE。在5名听力轻度下降者中,FPL校准的DPOAE和SFOAE能检出多数听力下降事件,而传统DPOAE未能检出。
要点
- FPL校准可降低DPOAE重测变异,提升序列监测稳定性。
- FPL校准的扫频DPOAE重测信度优于传统SPL校准DPOAE。
- FPL校准的DPOAE/SFOAE或可更早检出轻微听力变化。
核心结论
- FPL校准刺激水平相比传统耳内SPL方法可减少DPOAE水平的重测变异。
- FPL校准通过收紧OAE水平自然波动分布,可能改善OAE序列监测并更早发现轻微听力变化。
研究设计与样本
- 对20名成人同一耳每2个月测试一次,持续1年。
- 比较传统SPL校准的离散音DPOAE与FPL校准的扫频DPOAE及SFOAE的重测信度。
- 通过绝对水平差、测量标准误和95%正常偏移限评估信度。
主要发现
- FPL校准的扫频DPOAE整体重测信度显著优于传统DPOAE。
- 扫频DPOAE在中高频信度优于扫频SFOAE,低频两者测试间变异相当。
- 5名听力轻度下降者中,FPL校准的DPOAE和SFOAE检出多数听力下降事件,传统DPOAE未检出。
关键数值与推荐
- 重测信度评估指标包括绝对水平差、测量标准误和95%正常偏移限。
临床怎么用
- 在耳声发射序列监测中,可优先考虑采用FPL校准的扫频DPOAE以降低测试间变异。
- 对疑似轻微听力变化者,FPL校准的DPOAE和SFOAE可能提供更早的检出提示。
局限性
- 听力下降亚组仅5人,样本量小,需更大规模且听力变化确认的人群进一步验证。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: To use otoacoustic emissions (OAEs) for the serial monitoring of hearing or the tracking of hearing loss, test-retest reliability must be known, and normal OAE level variation must be established. Past studies have defined the reliability of distortion-product OAEs (DPOAEs) evoked with stimuli calibrated in a conventional manner (in-the-ear, dB sound pressure level), which can produce inaccuracies due to ear-canal standing waves. In the present study, we make within-subject comparisons of test reliability for conventionally measured 2 f1 - f2 DPOAEs compared with DPOAEs evoked by stimulus tones calibrated in dB forward pressure level (dB FPL) to mitigate the effect of standing waves. We also define the test-retest reliability of the stimulus-frequency OAE (SFOAE). DESIGN: Conventional DPOAEs evoked by discrete tones calibrated in dB SPL as well as DPOAEs and SFOAEs evoked by swept tones calibrated in dB FPL were measured in the same ear of 20 adult subjects every 2 mo over the course of 1 yr. OAE test reliability was assessed for each OAE paradigm by calculating absolute level differences, standard errors of the measurement, and 95% normative shift limits. Comparisons in reliability were made between conventional DPOAEs and those using dB FPL calibration and between DPOAEs and SFOAEs, both evoked by stimuli calibrated in dB FPL. RESULTS: Results indicate that overall DPOAEs evoked by swept tones calibrated in dB FPL have significantly better test-retest reliability than do conventional DPOAEs. Swept-tone DPOAEs also showed better reliability than swept-tone SFOAEs at mid-to-high frequencies, though at low frequencies test-to-test variation was comparable. In a subgroup of 5 subjects showing mild declines in audiometric hearing during the study, we found that DPOAEs and SFOAEs evoked by stimuli calibrated in dB FPL detected most instances of audiometric decline while conventional DPOAEs did not. CONCLUSIONS: FPL-based calibration of stimulus level reduces test-to-test variation in DPOAE level compared with conventional in-the-ear SPL methods. By creating a tighter distribution of natural or typical OAE level fluctuations on repeated testing, forward pressure calibration appears to improve the serial monitoring of OAEs and could provide earlier detection of slight hearing changes. Our preliminary look at this question should be further explored with a large group of individuals showing confirmed changes in hearing over time.
引用
- Abdala C, Benjamin T, Stiepan S, et al. Serial Monitoring of Hearing With Otoacoustic Emissions: Effects of Forward Pressure Level Calibration. Ear Hear. 2026. PMID: 42647103. doi:10.1097/AUD.0000000000001883
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。