文献速递耳科与耳神经

新型主动经皮骨锚式助听器植入患者接受MRI体验

Patient Experience of Undergoing MRI With a New Active Transcutaneous Bone-Anchored Hearing Implant.
PubMed 文献 · Laryngoscope · 2026-08-10

研究纳入15例植入Sentio Ti(1.5T条件兼容经皮骨锚式助听器)的患者,按制造商说明完成1.5T头部MRI后填写问卷,临床医生评估并发症及缓解措施,并随访2周。15/15例扫描无并发症或需缓解措施,随访期1例可能相关不良事件,患者不适感低,中位压力和疼痛评分分别为3和1(0-10分)。结论认为该植入患者可安全接受1.5T MRI。

要点

  • Sentio Ti植入患者按说明可安全完成1.5T头部MRI
  • 15例扫描均无并发症或需头带包扎、局麻等缓解措施
  • 患者报告不适感低,中位压力3分、疼痛1分(0-10分)

核心结论

  • 植入Sentio Ti的患者按制造商说明可安全接受1.5T头部MRI扫描。

研究设计与样本

  • 前瞻性观察性研究,纳入15例植入Sentio Ti并接受1.5T头部MRI的患者。
  • 扫描后患者填写体验问卷,临床医生评估并发症及缓解措施需求。
  • 随访2周监测不良事件。

主要发现

  • 15/15例MRI扫描完成,无并发症,无需头带包扎或局部麻醉等缓解措施。
  • 随访期记录1例可能与该操作相关的不良事件。
  • 患者整体体验良好,不适感低;植入部位中位压力评分3分、疼痛评分1分(0-10数字评分量表)。

关键数值与推荐

  • 植入体为1.5T条件兼容经皮骨锚式助听器Sentio Ti。
  • 扫描按制造商说明进行1.5T头部MRI。

临床怎么用

  • 对于Sentio Ti植入患者,可按制造商说明安排1.5T头部MRI,无需常规预防性头带包扎或局麻。
  • 扫描后仍应短期随访,关注可能的不良事件。

局限性

  • 样本量仅15例,临床经验有限。
  • 仅评估1.5T头部MRI,未涉及其他场强或部位。
  • 随访期仅2周,长期安全性未评估。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: Hearing implants may be a concern if a patient needs to undergo magnetic resonance imaging (MRI). Despite the development of MRI conditional implants, there are still reports of complications and of patients for whom the MRI needs to be aborted prematurely due to having an implant. The Sentio Ti implant is a new 1.5 T conditional transcutaneous bone-anchored hearing implant. Owing to the limited clinical experience with head MRI procedures in recipients of this recently introduced implant, the present investigation sought to rigorously characterize patient-reported experiences during MRI and to systematically assess the incidence and nature of implant-related complications and adverse events. METHOD: Fifteen implanted patients underwent a 1.5 T head MRI scan according to the manufacturer's instructions and afterwards completed a questionnaire about their experience. Furthermore, the clinicians were asked to rate their experience and report complications or need for mitigating actions, such as head bandaging or local anesthesia. Adverse events were monitored in a 2-week follow-up period. RESULTS: In 15/15 cases, the MRI scan was completed without complications or need for mitigating actions. During the immediate follow-up period, one adverse event possibly related to the procedure was noted. Patients reported overall good experiences, with low levels of discomfort. Median pressure and pain at the implant site were 3 and 1, respectively, assessed on a numerical rating scale of 0-10. CONCLUSION: Patients with the investigated implant can safely undergo a 1.5 T MRI scan according to the manufacturer's instructions.

来源:Laryngoscope,查看原文。版权归原作者所有。

引用

  1. Hajema C, Wesarg T, Tysome J, et al. Patient Experience of Undergoing MRI With a New Active Transcutaneous Bone-Anchored Hearing Implant. Laryngoscope. 2026. PMID: 42576286. doi:10.1002/lary.70772

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。