术中视觉确认耳蜗外电极:刺激电流诱导非刺激电极电压的应用
本研究为前瞻性术中研究,纳入成人人工耳蜗候选者或使用者,在电极插入过程中记录刺激电流诱导非刺激电极电压(SCINSEVs),并由不知手术结果的专家识别耳蜗外电极(EE)数量。结果显示SCINSEVs识别EE存在的总体灵敏度为95.9%、特异度为40.7%,与视觉确认的EE数量相关性r=0.63(p<0.0001),可作为术中快速、无辐射的EE检测工具。
要点
- SCINSEVs可在术中快速识别耳蜗外电极,灵敏度95.9%、特异度40.7%。
- EE数量与基于EFI的EE排名相关(r=0.63,p<0.0001)。
- 该方法为无辐射替代术后CT,结果术中可得,便于立即修正。
核心结论
- SCINSEVs是术中检测耳蜗外电极的快速可靠工具,高灵敏度、中等特异度,可无辐射替代术后CT并在术中提供结果。
研究设计与样本
- 前瞻性术中研究,纳入单中心接受初次或翻修手术的成人人工耳蜗候选者或使用者,术中随机将1至6个电极置于耳蜗外并由手术显微镜视觉确认,记录模拟术后条件(血液或筋膜包围EE)的EFI数据,由3名不知手术结果的专家审查EFI以识别EE数量。
主要发现
- 一般线性模型显示术者未插入的EE数量(0至6个)和审查者(1至3名)对EE排名误差有显著影响,手术类型(翻修或初次)和材料(血液或筋膜)无影响。
- 未插入EE数量与基于EFI的EE排名相关(r=0.63,p<0.0001)。
- 基于SCINSEVs视觉检查识别EE存在与否的总体灵敏度为95.9%、特异度为40.7%;在±2 EE审查者-术者一致容差下为97.9%和65.7%。
关键数值与推荐
- 识别EE存在与否的总体灵敏度95.9%、特异度40.7%(±2 EE一致容差下97.9%和65.7%)。
- 未插入EE数量与EFI排名相关r=0.63(p<0.0001)。
临床怎么用
- 可在人工耳蜗植入术中记录并解读SCINSEVs以快速识别耳蜗外电极,便于立即手术修正,减少对术中或术后影像的依赖。
- 作者建议培训相关临床人员记录和解读这些测量以用于临床实践。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: Extra-cochlear electrodes (EEs) are a frequent complication in cochlear implant (CI) surgery, potentially leading to reduced speech-perception outcomes. Standard intraoperative tools such as contact impedances and electrically evoked compound action potentials often fail to detect these misplaced electrodes. Stimulation Current-Induced Non-Stimulating Electrode Voltages (SCINSEVs), or electrical field imaging (EFI) in the advanced bionics device, are presented here as a faster and more accurate intraoperative measure for identifying EE. Adopting this method clinically could reduce reliance on intraoperative or postoperative imaging and allow EE identification in the critical window of surgery, when immediate correction can occur. DESIGN: This prospective intraoperative study tested adult CI candidates or users undergoing new or revision surgery at a single institution. EFI measurements were recorded during electrode insertion, where the surgeon was asked to hold a randomly selected number of 1 to 6 electrodes outside of the cochlea before full insertion. Visual confirmation of the number of EE electrode positions was achieved with the surgical microscope. Recordings were taken simulating postoperative conditions (with EE surrounded by blood or fascia). Three expert reviewers blinded to surgical outcomes reviewed EFI data to identify the number of EEs. Agreement between reviewers and surgeons was analyzed using correlations, receiver operating characteristic plots, and sensitivity and specificity calculations. RESULTS: General linear models revealed significant effects of the number of EEs not inserted by the surgeon (0 to 6 EEs) and the reviewer (1 to 3) on the EE ranking error, and no effects of the surgery type (revision or first-time cochlear implantation), nor of the material used (blood or fascia). A correlation was observed between the number of not-inserted EEs and EEs ranked based on the EFIs (r = 0.63, p < 0.0001). For identification of the presence or absence of EEs based on visual inspection of SCINSEVs, an overall sensitivity of 95.9% and a specificity of 40.7% were achieved (97.9% and 65.7% with ±2 EE reviewer-surgeon agreement tolerance). CONCLUSIONS: SCINSEVs are a rapid, reliable tool for intraoperative EE detection. With high sensitivity and moderate specificity, they offer a non-radiative alternative to postoperative computerized tomography, with results available during the surgical procedure. Use of this tool may improve CI outcomes through immediate surgical correction. The authors recommend training of relevant clinical staff to record and interpret these measurements for use within clinical practice.