耳蜗内电图音调定位测量评估人工耳蜗性能:侧壁与蜗轴电极阵列比较
这项前瞻性队列研究在57例人工耳蜗植入者中,术中通过耳蜗内电图(ECochG)测量多频短音诱发的最大幅度响应总和(MA-ECochG-TR),并与术后6个月CNC词得分相关分析。结果显示MA-ECochG-TR对蜗轴阵列(Mid-Scala)有中度强正相关(r=0.73),对侧壁阵列(SlimJ)无显著相关(r=0.04)。结论认为ECochG可独立预测蜗轴阵列性能,对侧壁阵列预测力较弱。
要点
- MA-ECochG-TR与蜗轴阵列6个月CNC词得分显著相关(r=0.73)。
- MA-ECochG-TR与侧壁阵列6个月CNC词得分无显著相关(r=0.04)。
- 侧壁阵列中排除EAS者后相关性仍不显著(r=0.44)。
核心结论
- MA-ECochG-TR可独立估算蜗轴阵列的术后性能。
- 耳蜗内电图对侧壁阵列的预测力较弱,可能与EAS混杂或电极-蜗轴距离增大导致信号衰减有关。
研究设计与样本
- 前瞻性队列研究,纳入57例接受人工耳蜗植入并完成术中ECochG记录的患者,其中43例完成6个月术后测试。
- 术中采用Advanced Bionics的Active Insertion Monitoring系统,经声管给予250 Hz至2 kHz多频短音(95-105 dB HL),记录每个耳内电极的响应。
主要发现
- MA-ECochG-TR与蜗轴阵列(Mid-Scala,n=23)6个月CNC词得分呈中度强正相关(r=0.73,95% CI [0.46-0.88],P < .001)。
- MA-ECochG-TR与侧壁阵列(SlimJ,n=20)6个月CNC词得分无显著相关(r=0.04,95% CI [-0.41 to 0.47],P = .87)。
- 13例未使用电声刺激(EAS)的侧壁阵列受者中,MA-ECochG-TR与6个月CNC词得分呈中度但不显著正相关(r=0.44,95% CI [-0.14 to 0.80],P = .13)。
临床怎么用
- 对于使用蜗轴阵列(如Mid-Scala)的成人人工耳蜗植入者,可考虑术中MA-ECochG-TR辅助预估术后言语识别表现。
- 对于侧壁阵列(如SlimJ),MA-ECochG-TR的预测价值有限,需结合其他因素综合判断。
局限性
- 侧壁阵列样本量较小(n=20),且EAS混杂因素未达统计学显著。
- 仅随访至6个月,长期预测效度未知。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: To assess intracochlear electrocochleography (ECochG) for estimating cochlear implant (CI) performance for lateral wall and perimodiolar electrode arrays. STUDY DESIGN: Prospective cohort study. SETTING: Tertiary referral center. METHODS: From December 2021 to June 2024, ECochG was measured intraoperatively using the Active Insertion Monitoring (AIM) system by Advanced Bionics. After insertion, multifrequency tone bursts were delivered via a sound tube from 250 Hz to 2 kHz at 95 to 105 dB HL with recordings at each internal electrode. Using spectral analysis, the maximal amplitude (MA) response electrode was identified by frequency and summed into a single value of total response (MA-ECochG-TR). Postoperative CI performance was routinely monitored. RESULTS: Of 57 participants with ECochG recordings, 43 completed 6-month testing. Of these, 23 were perimodiolar (Mid-Scala [MS]), and 20 were lateral wall (SlimJ [SJ]). MA-ECochG-TR had a moderately strong positive correlation to 6-month consonant-nucleus-consonant (CNC) word scores for MS cases (r = 0.73, 95% CI [0.46-0.88], P < .001), but showed no significant correlation for SJ cases (r = 0.04, 95% CI [-0.41 to 0.47], P = .87). For the 13 SJ recipients without electroacoustic stimulation (EAS), there was a moderate, non-significant positive correlation between MA-ECochG-TR and 6-month CNC (r = 0.44, 95% CI [-0.14 to 0.80], P = .13). CONCLUSION: MA-ECochG-TR independently estimates performance for perimodiolar arrays. Intracochlear ECochG is less predictive for lateral wall arrays; this may be due to confounding from EAS (though this was not statistically significant) or from signal degradation with a larger electrode-modiolar distance.