人工耳蜗与电灼术:活体动物模型研究
结果显示最大绝对温度变化平均为头皮4.1°C、圆窗2.9°C,时间和距离对头皮温度无显著影响;所有设备均通过制造商深度分析,挑战了当前禁忌证。
要点
- 头皮最大绝对温度变化平均4.1°C,圆窗2.9°C,未见显著时间或距离效应。
- 结果支持重新审视制造商关于人工耳蜗患者使用单极电灼的禁忌证。
研究设计与样本
- 经IACUC批准,将12个制造商捐赠的CI植入两只麻醉猪,其中6个进行现场阻抗测试,6个进行电灼后分析。
主要发现
- 三个位点平均温度为32.9°C至38.8°C。
- 头皮最大绝对温度变化平均4.1°C(95%CI:1.4-6.8),圆窗平均2.9°C(95%CI:1.0-4.9)。
- 头皮温度变化与时间(p=0.09)和距离(p=0.64)均无显著关联。
- 设备按距离分层,平均最大温度变化呈现距离梯度,但设备邻近度差异无统计学意义。
- 6个设备进行植入后分析,全部通过制造商深度设备分析。
临床怎么用
- 本实验证据提示单极电灼在测试条件下未损害CI设备,可支持对现行禁忌证进行循证指南修订。
局限性
- 研究为活体动物模型,样本量有限(12个CI,两只猪),结果外推至临床需谨慎。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: The use of monopolar cautery is regarded as a contraindication in cochlear implant (CI) recipients, despite a large amount of contradicting data. The objectives are as follows: (1) Evaluate the effects of high-powered electrocautery on CI devices. (2) Record any high-powered electrocautery induced temperature changes. METHODS: After IACUC approval, 12 manufacturer-donated CIs were implanted in two anesthetized pigs. All devices underwent pre-study integrity checks; six had onsite impedance testing, while six received post-cautery analysis. Fifty watts coagulation was applied for 20 min at sites 1-15 cm from the CI. Bluetooth thermometers monitored three sites of temperatures. Devices were explanted for analysis. Stata 17.0 was used for statistical analysis. A p-value < 0.05 was considered statistically significant. Outcome measurements were as follows: (1) Changes in the temperature. (2) Changes to the implanted CIs during electrocautery. (3) CI integrity analysis following electrocautery. RESULTS: The mean temperatures at the three sites (a*, b*, c*) was 32.9°C-38.8°C. Maximum absolute temperature changes averaged 4.1°C (95% CI: 1.4-6.8) for scalp measurements and 2.9°C (95% CI: 1.0-4.9) for round window measurements. Neither time (p = 0.09) nor distance (p = 0.64) generated significant scalp temperature changes. Devices were stratified by proximity. Mean maximum temperature changes showed a gradient by distance. Differences in device proximity were not statistically significant. Postimplantation device analysis was performed on six devices, and all passed the manufacturer in-depth device analysis. CONCLUSIONS: This controlled experimental evidence, demonstrating 100% device integrity across all tested conditions, challenges current manufacturer contraindications and supports evidence-based guideline revision. LEVEL OF EVIDENCE: N/A.
引用
- Jeyakumar A, Lu Q, Johnson R, et al. Cochlear Implants and Electrocautery: A Live Animal Model. Laryngoscope. 2026. PMID: 42242924. doi:10.1002/lary.70649
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。