骨导并行听觉脑干反应范式的可行性研究
本研究在听力正常的年轻成人中比较了气导插入式耳机与骨导振动器并行听觉脑干反应(pABR)的波形特征。结果显示两种换能器在相同dB nHL下波V潜伏期与幅值相似,残余噪声相近,8 kHz骨导反应较弱,行为阈值高于纯音刺激且偏移量一致,提示骨导pABR可行。
要点
- 骨导与气导pABR在相同dB nHL下波V潜伏期和幅值相似。
- 骨导8 kHz反应较弱,与骨导物理特性和振动器输出有关。
- 行为阈值pABR高于纯音,两种换能器偏移量相同。
核心结论
- 气导与骨导pABR波形相似,表明骨导pABR可行。
- 结合dB nHL值建立,该工作为pABR诊断所有类型听力损失迈出重要一步。
研究设计与样本
- 招募听力正常年轻成人队列,测量双通道ABR强度序列。
- 使用ER-2插入式耳机和B-81骨导振动器(对侧掩蔽)呈现pABR刺激。
- 仅一耳进行骨导测试,骨导耳别和换能器顺序在受试者间平衡。
主要发现
- 相同dB nHL下两种换能器反应对齐良好,波V幅值和潜伏期相似。
- 骨导刺激伪迹可通过气导pABR的缓解措施有效控制,残余噪声相似。
- 8 kHz骨导反应较弱。
- 行为阈值pABR高于纯音刺激,两种换能器偏移量相同。
- dB nHL值因插入式耳机与骨导振动器的参考等效阈值水平不同而异。
临床怎么用
- 骨导pABR可用于评估传导性与感音神经性听力损失的相对贡献。
- 临床解读需注意8 kHz骨导反应较弱及dB nHL校正因子差异。
局限性
- 仅纳入听力正常年轻成人,未涉及听力损失人群。
- 仅一耳进行骨导测试,未双侧验证。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: Auditory brainstem responses (ABRs) are used to diagnose hearing loss across multiple frequencies in infants and others who cannot participate in behavioral testing. The parallel ABR (pABR) paradigm measures response waveforms to all frequencies from 500 to 8000 Hz in both ears simultaneously, rather than serially in one ear at a time. Work so far has been conducted with air-conducted stimuli through insert earphones, but bone-conducted stimuli are also an essential part of diagnosing the type of hearing loss by determining the relative contributions of conductive and sensorineural hearing loss. This study aimed to confirm the feasibility of bone-conduction pABR. DESIGN: We recruited a cohort of young adults with normal hearing. An intensity series of two-channel ABRs was measured to pABR stimuli played through ER-2 insert earphones, as well as with a B-81 bone vibrator with contralateral masking. Behavioral thresholds were also measured to compare dB nHL correction factors. To keep sessions under 3.5 hr only one ear was tested with bone conduction. The ear with bone conduction and the order of transducers were counterbalanced across participants. Waveform morphologies for each transducer were compared by their residual noise and their wave V latencies and amplitudes. RESULTS: For the same dB nHL levels, responses to both transducers aligned well, and wave V amplitudes and latencies were similar. While stimulus artifacts can be an issue with bone conduction, the steps taken to mitigate them in air conduction pABR were also effective for bone conduction: clear bone-conducted responses were measured, and residual noise was similar for both transducers. Bone-conducted responses to 8 kHz were weak, which was not surprising given the physics of bone conduction and the output capabilities of the bone vibrator. As expected, behavioral thresholds were higher for pABR than pure-tone stimuli, and by the same offset for both transducers. The dB nHL values differed due to the difference in reference equivalent threshold levels for inserts and the bone vibrator. CONCLUSIONS: Air and bone conduction provided similar waveforms, indicating that bone-conduction pABR is feasible. Together with establishing dB nHL values, this work represents an important step toward translating the pABR for diagnosing all types of hearing loss.
引用
- Polonenko MJ, Maddox RK. Bone-Conducted Responses Using the Parallel Auditory Brainstem Response Paradigm. Ear Hear. 2026. PMID: 42129598. doi:10.1097/AUD.0000000000001833
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。