文献速递耳科与耳神经

多频耳蜗电图在人工耳蜗植入中减少下降警报

Multi-Frequency Electrocochleography Results in Fewer Drop Alarms During Cochlear Implant Insertion.
PubMed 文献 · Laryngoscope · 2026-05-05

结果显示多频算法较单频算法显著减少下降警报次数(0.42 vs 1.72,p<0.001),并改变植入轨迹模式,可能更准确评估耳蜗微环境。

要点

  • 多频耳蜗电图较单频显著减少下降警报次数(0.42 vs 1.72)。
  • 多频算法下Type C和Type D轨迹模式减少,响应更多。
  • 多频耳蜗电图可能提供更准确的耳蜗微环境评估。

核心结论

  • 新型多频耳蜗电图算法与更少下降警报和改变的植入轨迹模式相关,可能更准确评估耳蜗微环境。

主要发现

  • 单频算法平均下降警报次数为1.72(95% CI: 1.52, 1.92;中位数1),多频算法为0.42(95% CI: 0.31, 0.53;中位数0),p < 0.001。
  • 多频耳蜗电图算法较单频算法减少了Type C模式(植入中振幅上升后下降)的数量。
  • Type D模式(无响应)数量减少,表明多频耳蜗电图在耳蜗内产生更多响应。

关键数值与推荐

  • 耳蜗电图振幅下降6 dB定义为下降警报。
  • 多频刺激同时使用125至4000 Hz之间的四个频率。

临床怎么用

  • 在人工耳蜗植入过程中,采用多频耳蜗电图算法可减少下降警报,可能提高对耳蜗微环境评估的准确性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: To evaluate intracochlear electrocochleography (ECochG) amplitude parameters during cochlear implantation (CI) using a novel multi-frequency ECochG algorithm. METHODS: A multi-institutional, prospective cohort study was performed at 18 high-volume CI centers. The inclusion criteria were adults with sensorineural hearing loss and audiometric thresholds of ≤ 90 dB hearing level at 500 Hz undergoing CI with Advanced Bionics (Valencia, CA) Ultra 3D devices between 2024 and 2025. ECochG recordings were performed with simultaneous multi-frequency stimulation of four frequencies between 125 and 4000 Hz during cochlear implant insertion. Concurrent multi-frequency recording allowed extraction of amplitude and phase of each frequency individually. Post hoc analysis was performed to determine the difference in the number of drop alarms between single- and multi-frequency ECochG. An ECochG amplitude drop of 6 dB was defined as a drop alarm. Insertion track patterns were compared between single- and multi-frequency ECochG. RESULTS: One hundred ninety-five ears were included. Mean number of drop alarms for the single-frequency algorithm was 1.72 (95% CI: 1.52, 1.92; median 1) compared to 0.42 (95% CI: 0.31, 0.53; median 0) for multi-frequency; p < 0.001. The number of Type C patterns (rise in amplitude during insertion followed by a drop) decreased with the multi-frequency ECochG algorithm compared to the single-frequency ECochG algorithm. The number of Type D patterns (no-response) decreased, indicating that multi-frequency ECochG generated more responses across the cochlea than single-frequency ECochG. CONCLUSIONS: A novel multi-frequency ECochG algorithm during CI is associated with fewer drop alarms and altered insertion track patterns, which may provide a more accurate assessment of the cochlear microenvironment.

来源:Laryngoscope,查看原文。版权归原作者所有。

引用

  1. Espahbodi M, Syed A, Tu A, et al. Multi-Frequency Electrocochleography Results in Fewer Drop Alarms During Cochlear Implant Insertion. Laryngoscope. 2026. PMID: 42084107. doi:10.1002/lary.70592

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。