该研究通过两项预注册双盲被试内实验,评估12名耳蜗植入者对时域整合处理策略(TIPS)与临床策略(CIS或ACE)的言语识别表现。结果显示,在言语形噪声和多说话人噪声中,TIPS与基线策略的言语接收阈中位数相当,无统计学差异;TIPS可降低设备功耗21%至42%,但相较CIS策略音质评分较低(p=0.044),与ACE策略相比无差异。

要点

  • TIPS在噪声中言语识别与临床策略相当,未重现此前组水平显著改善。
  • TIPS可降低设备功耗21%至42%,不损害噪声下言语可懂度或音质。
  • 相较CIS策略,TIPS音质评分较低(p=0.044),与ACE策略无差异。

核心结论

  • TIPS可大幅节省功耗,且不损害噪声下言语可懂度或音质。
  • 未观察到Lamping等(2020)此前报告的组水平言语接收阈显著改善。

研究设计与样本

  • 两项预注册双盲被试内实验,纳入12名耳蜗植入听者。
  • 实验1比较TIPS与CIS基线策略,实验2比较TIPS与ACE基线策略。

主要发现

  • 实验1和2中,基线策略与TIPS的言语接收阈中位数在言语形噪声和多说话人噪声中均相当,无统计学差异。
  • TIPS相较CIS策略音质评分较低(p=0.044),与ACE策略相比无差异。
  • 所有条件下,TIPS可降低设备功耗21%至42%。

关键数值与推荐

  • 言语接收阈在言语形噪声(实验1)和多说话人噪声(实验1和2)中测量。
  • 音质采用平均意见评分在安静条件下获取(实验1和2)。
  • 所有策略使用受试者临床脉冲率、相位间隔、相位时长和接地电极,每种策略测试前适应10分钟。

临床怎么用

  • TIPS可在不损害噪声下言语可懂度或音质的前提下显著降低设备功耗。
  • 对个体而言,部分患者可能从TIPS获得言语接收阈改善,但组水平改善未重现。

局限性

  • 样本量较小(12名听者),组水平效应可能受限。
  • 音质评分在安静条件下获取,未评估噪声下音质。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: Speech perception remains challenging for cochlear-implant recipients in conditions containing background noise. The sound processing strategy of the cochlear implant transforms the acoustic signal into a pattern of electrical stimulation pulses, which are transmitted by the electrode array. Modifications to the sound processing strategy could improve the transmission of information by the cochlear implant, alleviating speech perception difficulties in noise. One such modification, the temporal integrator processing strategy (TIPS), uses a model of temporal masking to identify and remove stimulation pulses that are unlikely to be perceived. In a previous investigation, TIPS significantly improved the perception of continuous interleaved sampling-processed speech in the presence of speech-shaped noise while reducing the power required for stimulation ( Lamping et al. 2020 ). This study extends the previous investigation to include conditions that better reflect everyday listening for cochlear-implant recipients. DESIGN: Two pre-registered double-blind within-participant experiments measured the speech-recognition performance of 12 cochlear implant listeners for sentences before and after TIPS processing. Experiments 1 and 2 investigated TIPS performance against the Continuous Interleaved Sampling or Advanced Combination Encoder baseline speech processing strategies, respectively. Speech-reception thresholds were measured in a speech-shaped noise (in experiment 1) and a multi-talker noise (in experiments 1 and 2), and mean-opinion score speech quality ratings were obtained in quiet (in experiments 1 and 2). In both experiments, all strategies used the subjects' clinical pulse rate, inter-phase gap, phase duration, and ground electrodes, and participants were acclimatized for 10 min before testing each strategy. RESULTS: In experiments 1 and 2, the median speech-reception thresholds obtained were comparable across the baseline and TIPS strategies, with no statistical differences noted in either speech-shaped or multi-talker noise. Quality ratings were lower for TIPS when compared to the Continuous Interleaved Sampling strategy ( p = 0.044) but not when compared to the Advanced Combination Encoder strategy, which all participants used as their everyday strategy. In all conditions, TIPS could reduce device power consumption by between 21% and 42%. CONCLUSIONS: TIPS could result in substantial power savings, without compromising speech quality or speech intelligibility in noise. Although some individuals demonstrated improved speech-reception thresholds with TIPS, the significant group-level improvement previously found by Lamping et al. (2020 ) was not observed in this study.