文献速递耳科与耳神经

全植入式人工耳蜗在双侧重度感音神经性聋成人中的可行性研究

Evaluation of a Totally Implantable Cochlear Implant in Adults With Bilateral Sensorineural Hearing Loss: A Feasibility Study.
PubMed 文献 · Ear Hear · 2025-11-17

这项前瞻性单臂重复测量可行性研究纳入10例双侧中重度至极重度感音神经性聋成人,单侧植入全植入式人工耳蜗TICI研究系统。6个月时,外部处理器开启和关闭两种模式的言语识别均较基线显著改善;设备故障2例因麦克风液体侵入而再植入常规装置。

要点

  • 全植入式人工耳蜗在外部处理器关闭时仍可提供言语识别获益。
  • 外部处理器开启模式在安静词识别上优于关闭模式。
  • 2例因麦克风液体侵入导致设备故障并再植入常规装置。

核心结论

  • TICI研究系统在外部处理器开启和关闭两种模式下均可为患者提供获益,支持进一步开发该设计。
  • 未发现经皮麦克风使用中的根本性问题。

研究设计与样本

  • 前瞻性、单治疗臂、重复测量的可行性研究,评估临床获益与安全性。
  • 10例双侧重度至极重度感音神经性聋成人单侧植入TICI研究系统。

主要发现

  • 6个月时,安静词识别较基线平均改善38.5%(95%CI 24.6至52.5)为外部处理器开启模式,25.2%(95%CI 13.7至36.7)为关闭模式(n=9)。
  • 6个月时,噪声中句子识别较基线平均改善74.8%(95%CI 53.4至96.2)为开启模式,64.5%(95%CI 42.9至86.0)为关闭模式。
  • 6个月和12个月时,安静词识别开启模式均优于关闭模式,边际均值差分别为13.3%(p<0.001)和12.0%(p=0.004)。
  • 9个月时,安静音素识别在45、55和65 dB SPL下开启模式优于关闭模式,差异分别为33.9%(p<0.001)、21.7%(p<0.001)和9.8%(p=0.002),75 dB SPL下无显著差异(p=0.384)。
  • 数据记录显示关闭模式使用时间占总聆听时间的22%至88%(中位数49%)。

关键数值与推荐

  • 言语感知测试包括安静词(60 dB SPL)和噪声中句子(60 dB SPL,+10 dB SNR)。
  • 患者报告结局采用言语、空间与听觉质量量表、健康效用指数mark 2/3和患者满意度调查。

临床怎么用

  • 全植入式设计有望让患者在外部处理器关闭时仍获得听觉获益。
  • 需关注经皮麦克风的液体侵入风险及设备故障。
  • 外部处理器开启模式在安静环境下仍提供额外言语识别优势。

局限性

  • 样本量小(10例植入,部分分析n=9)。
  • 单臂设计,缺乏对照组。
  • 随访至植入后2年,长期性能未报告。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: To investigate the feasibility of a newly designed investigational totally implantable cochlear implant, the Cochlear ™ totally implantable cochlear implant (TICI) Research System, which enables hearing without the use of an external sound processor. DESIGN: A prospective, single-treatment arm, repeated measures feasibility study to investigate clinical benefit and safety. Ten adult participants with bilateral moderately severe-to-profound sensorineural hearing loss were implanted unilaterally with the TICI Research System. Co-primary endpoints evaluated external hearing (EH) mode (i.e., sound processor on) and invisible hearing (IH) mode (i.e., sound processor off) performance using words in quiet (60 dB SPL) and sentences in babble noise (60 dB SPL, +10 dB SNR) from the preoperative baseline to 6-month post-activation. The same speech perception tests were re-measured at 12-month post-activation. At 9-month post-activation, performance using phonemes in quiet was measured at a wider range of speech input levels (45, 55, 65, and 75 dB SPL), such as would be encountered in real-life settings. Patient-reported outcomes were evaluated using the Speech, Spatial, and Qualities of Hearing scale, Global Health Utilities Index mark 2/3, and Patient Satisfaction Survey. Device data logging captured time of use statistics in each hearing mode. Safety outcomes (adverse events and device deficiencies) were reported for the study duration (up to 2 years post-implantation). RESULTS: Speech performance with the TICI Research System for both EH and IH modes at 6 months was significantly improved over baseline (n = 9). For words in quiet, there was a mean improvement of 38.5% (95% confidence interval [CI], 24.6 to 52.5) and 25.2% (95% CI, 13.7 to 36.7) for EH and IH, respectively. For sentences in noise, there was a mean improvement of 74.8% (95% CI, 53.4 to 96.2) and 64.5% (95% CI, 42.9 to 86.0) for EH and IH, respectively. For words in quiet, performance was found to be significantly better for EH compared with IH, with marginal mean differences of 13.3% ( p < 0.001) and 12.0% ( p = 0.004) at 6 and 12 months, respectively. Likewise, for phonemes in quiet at 9 months, EH was significantly better than IH with differences of 33.9% ( p < 0.001), 21.7% ( p < 0.001), and 9.8% ( p = 0.002) at 45, 55 and 65 dB SPL, respectively, with no significant difference at 75 dB SPL ( p = 0.384). For sentences in noise, at 6 months, EH had significantly better performance than IH with a marginal mean difference of 10.3% ( p = 0.021), with no difference shown at 12 months ( p = 1.000). Speech, Spatial, and Qualities of Hearing scale showed a significant mean improvement of 1.84 (95% CI, 1.00 to 2.65) compared with baseline. Health Utilities Index mark 2/3 showed variability in scores across the participant cohort, and Patient Satisfaction Survey ratings showed that overall, participants were satisfied with the device. Data logging showed participants' use of IH varied from 22% to 88% of total listening time (median 49%). Device failures occurred in 2 participants due to a microphone-related fluid ingress issue. Both participants were reimplanted with conventional Cochlear Nucleus ® devices. No other safety issues were detected. CONCLUSIONS: This study demonstrated the feasibility of the TICI Research System to provide patient benefit in both IH and EH modes. No fundamental problem with the use of a transcutaneous microphone was detected. These findings support further development of the TICI design and approach.

来源:Ear Hear,查看原文。版权归原作者所有。

引用

  1. Briggs R, Tari S, Rousset A, et al. Evaluation of a Totally Implantable Cochlear Implant in Adults With Bilateral Sensorineural Hearing Loss: A Feasibility Study. Ear Hear. 2025. PMID: 41243136. doi:10.1097/AUD.0000000000001750

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。