植入舌下神经刺激器患者使用单极电灼的安全性
该回顾性研究分析2012年10月至2023年10月一家三级学术中心连续85例舌下神经刺激器(HGNS)植入病例,其中38例(45%)后续手术使用单极电灼(MPE)。MPE暴露后1个月内无器械故障或失灵,11例有暴露前后睡眠研究的患者AHI无显著差异(P=1.00)。作者认为,在HGNS植入物邻近区域以外使用MPE是安全的。
要点
- 85例HGNS植入患者中38例(45%)后续手术使用MPE,躯干上下手术结局无显著差异(P=0.478)。
- MPE暴露后1个月内未发生器械故障或失灵。
- 11例有暴露前后睡眠研究的患者AHI无显著差异(P=1.00)。
核心结论
- 这是首项评估HGNS植入患者使用MPE安全性的研究,提示在HGNS植入物邻近区域以外使用MPE是安全的,不损害HGNS功能。
研究设计与样本
- 回顾性综述,纳入2012年10月至2023年10月一家三级学术中心连续85例HGNS病例,按后续手术是否使用MPE分为MPE组(n=38)和无MPE组(n=47)。
主要发现
- 85例患者中38例(45%)接受MPE手术,其中头颈部7例(18%)、躯干和上肢20例(53%)、躯干以下11例(29%)。
- 躯干以上与躯干以下手术的结局无显著差异(P=0.478)。
- MPE暴露后至1个月,未发生器械故障或失灵。
- 11例有MPE暴露前后睡眠研究的患者中,AHI无显著差异(P=1.00)。
临床怎么用
- 作者认为,在HGNS植入物邻近区域以外使用MPE是安全的,不损害HGNS功能。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: Monopolar electrocautery (MPE) is traditionally avoided in patients with implanted hardware due to concerns about electromagnetic interference. While MPE use in cardiac, spinal cord, and cochlear implants has been documented, no study exists on its impact on hypoglossal nerve stimulators (HGNS). Here we evaluate the impact of MPE use in patients with HGNS implants. METHODS: This is a retrospective review of 85 consecutive HGNS cases at a tertiary academic center from October 2012 to October 2023. Post-HGNS implantation surgical procedures were categorized into surgeries with MPE (n = 38) and without MPE (n = 47). Data on demographics, polysomnography, and device outcomes were analyzed using descriptive statistics, Wilcoxon signed-rank tests, and chi-squared tests. RESULTS: Of the 85 patients, 38 (45%) underwent MPE-involved surgeries, with 7 (18%) head and neck, 20 (53%) torso and upper extremities, and 11 (29%) below the torso. No significant difference in outcomes was found between surgeries above or below the torso (p = 0.478). Up to 1 month post-MPE exposure, no device failure or malfunction occurred. Among 11 patients with sleep studies before and after MPE exposure, there was no significant difference in AHI (p = 1.00). CONCLUSION: This is the first study to assess the safety of MPE in patients with HGNS implants, suggesting that MPE use outside the immediate vicinity of HGNS implants is safe and does not compromise HGNS function.
引用
- Abdul-Rahman NH, Choi AY, Ravuri S, et al. Safety of Monopolar Electrocautery Use in Patients With Implanted Hypoglossal Nerve Stimulators. Laryngoscope. 2025. PMID: 40407268. doi:10.1002/lary.32291
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。