基于同步辐射的机器人耳蜗植入电极插入创伤评估
本研究在15例人类颞骨标本上使用OTODRIVE系统以0.1 mm/s恒定速度插入多种电极和INCAT导管,经同步辐射相位对比成像评估耳蜗内创伤。平均线性插入深度24.8±3.5 mm,平均角度插入深度527°±81°,耳蜗覆盖率为74%±7%。仅观察到1例部分蜗管移位和2例基底膜抬高,无严重创伤。作者认为该系统可实现一致且受控的电极放置并减少创伤,但需进一步临床研究。
要点
- 15例人类颞骨标本使用OTODRIVE系统以0.1 mm/s恒定速度插入电极。
- 平均线性插入深度24.8±3.5 mm,平均角度插入深度527°±81°,耳蜗覆盖率74%±7%。
- 仅1例部分蜗管移位和2例基底膜抬高,无严重创伤。
核心结论
- OTODRIVE系统可实现一致且受控的电极放置,同时减少对耳蜗结构的创伤。
- 作者认为机器人辅助技术可能改善听力和结构保留结局,但需进一步临床研究以确定其相对于手动插入的长期获益。
研究设计与样本
- 在15例人类颞骨标本上使用OTODRIVE系统以0.1 mm/s恒定速度插入多种电极(FLEX26、FLEX28、FLEXSOFT、FLEX34)和INCAT导管,随后用同步辐射相位对比成像评估耳蜗内创伤。
主要发现
- 平均线性插入深度为24.8±3.5 mm,平均角度插入深度为527°±81°,耳蜗覆盖率为74%±7%。
- 仅观察到1例部分蜗管移位和2例基底膜抬高。
- 未观察到电极尖端折叠、前庭阶移位或骨螺旋板骨折等严重创伤。
临床怎么用
- 原文未据此给出临床推荐。
局限性
- 原文未报告局限性。
- 原文结论指出需进一步临床研究以确定机器人辅助技术相对于手动插入在常规耳蜗植入中的长期获益。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: Minimizing intraoperative trauma during cochlear implant electrode insertion is crucial for preserving residual hearing. Manual insertions are subject to variability due to surgeon experience and anatomical differences, often leading to inconsistent outcomes. Robotic-assisted systems have been developed to improve insertion precision and reduce trauma. This study evaluates the safety and efficacy of the OTODRIVE system for cochlear implant electrode insertion. METHODS: Fifteen cadaveric human temporal bones were implanted with a variety of electrodes (FLEX26, FLEX28, FLEXSOFT, FLEX34) and an inner ear catheter (INCAT). The OTODRIVE system was used to maintain a constant insertion speed of 0.1 mm/s. The samples were subsequently imaged using synchrotron radiation phase-contrast imaging (SR-PCI) to assess intracochlear trauma. Parameters such as linear and angular insertion depths and cochlear coverage were measured. The presence of trauma was analyzed using a trauma evaluation plot. RESULTS: The average linear insertion depth was 24.8 ± 3.5 mm, and the average angular insertion depth was 527° ± 81°. Cochlear coverage was 74% ± 7%. Only one case of partial scala media translocation and two cases of basilar membrane elevation were observed, with no instances of severe trauma such as tip fold-over, scala vestibuli translocation, or osseous spiral lamina fracture. CONCLUSION: The findings indicate that the OTODRIVE system enables consistent and controlled electrode placement while minimizing trauma to cochlear structures, suggesting its potential to enhance hearing and structural preservation outcomes. Further clinical studies are required to establish the long-term benefits of robotic-assisted techniques over manual insertions in routine cochlear implantation. LEVEL OF EVIDENCE: N/A.
引用
- Chen JM, Lin VYW, Le TN, et al. Synchrotron-Based Trauma Assessment of Robotic Electrode Insertions in Cochlear Implantation. Laryngoscope. 2025. PMID: 40387265. doi:10.1002/lary.32254
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。