人工耳蜗植入者磁共振成像的疼痛、安全性与图像质量:一项前瞻性试验
这项前瞻性队列研究在两个三级学术中心纳入带人工耳蜗(CI)的成年患者,评估可旋转磁铁能否改善磁共振成像(MRI)时的舒适度并保持扫描可用性。结果显示,可旋转磁铁装置的疼痛评分变化显著低于固定磁铁装置(0.7 vs 4.0,P = .02),未发生磁铁移位或退磁;虽产生可测量的伪影,但多数情况下未明显影响扫描可用性。
要点
- 可旋转磁铁装置MRI相关疼痛评分变化显著低于固定磁铁装置(0.7 vs 4.0,P = .02)。
- 研究期间未发生磁铁移位或退磁。
- 旋转磁铁装置在6个脑部MRI序列轴位图像上的平均最大信号缺失伪影为6.1 cm(标准差1.2 cm)。
核心结论
- 含可旋转磁铁的人工耳蜗较固定磁铁可减少MRI相关疼痛。
- 磁铁虽产生可测量的伪影,但多数情况下未明显影响扫描可用性。
研究设计与样本
- 前瞻性队列研究,在两个三级学术中心开展。
- 纳入接受任何部位MRI的带人工耳蜗成年患者。
- 共分析23次MRI扫描(21例患者),其中11次扫描(47.8%)涉及设计为可与磁场自对准的内置磁铁植入体。
主要发现
- 与头巾使用无关的疼痛评分平均差值,可旋转磁铁装置显著低于固定磁铁装置(0.7 vs 4.0,P = .02)。
- 未发生磁铁移位或退磁。
- 在6个MRI脑部序列中,可旋转磁铁装置轴位图像的平均最大信号缺失伪影为6.1 cm(标准差1.2 cm)。
- 评估MRI能否回答开具扫描所要解决的问题时,100%的应答(n = 12)介于略微同意至强烈同意之间。
- 多数患者(70.0%)表示如有医疗需要愿意再次接受MRI。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: To evaluate whether rotating cochlear implant (CI) magnets improve patient comfort and preserve scan usability during magnetic resonance imaging (MRI). STUDY DESIGN: Prospective cohort study. SETTING: Two tertiary academic centers. METHODS: Adult patients with CIs who underwent MRI of any body region were included. The primary outcome was the mean change in pain score from baseline (before any study procedure) to the highest reported pain level during MRI with a CI magnet in place. Secondary outcomes included patient-reported survey responses regarding their MRI experience and the interpretability of the brain MRI scans involving rotating magnets, as assessed by two independent, blinded neuroradiologists. RESULTS: Twenty-three MRI scans (21 patients) with CIs were analyzed. Eleven scans (47.8%) involved implants containing internal magnets designed to self-align with the magnetic field. The mean pain-score difference unrelated to headwrap use was significantly lower for rotating magnet devices than fixed magnets (0.7 vs 4.0, P = .02). No magnet displacement or depolarization occurred. In six MRI brain sequences, the mean maximal signal-loss artifact measured 6.1 cm (SD 1.2 cm) on axial images for rotating magnet devices. When assessing if the MRI could answer the question for which the scan was ordered, 100% of responses (n = 12) ranged from slightly agree to strongly agree. The majority of patients (70.0%) indicated they would undergo another MRI if medically necessary. CONCLUSION: CIs containing rotatable magnets reduce MRI-related pain compared to those with fixed magnets. Although the magnets produce a measurable artifact, it does not substantially compromise scan utility in the majority of cases.
引用
- Patel EJ, Wu X, McLouth J, et al. Pain, Safety, and Image Quality of Magnetic Resonance Imaging With Cochlear Implants: A Prospective Trial. Otolaryngol Head Neck Surg. 2025. PMID: 40376713. doi:10.1002/ohn.1301
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。