文献速递耳科与耳神经

人工耳蜗电极阵列尖端折叠术中矫正后的言语表现

Speech Performance Following Intraoperative Correction of Cochlear Implant Electrode Array Tip Fold-Overs.
PubMed 文献 · Otolaryngol Head Neck Surg · 2025-03-07

该回顾性图表综述分析了2016至2023年接受细长型耳蜗周围电极阵列的663例植入者,其中35例(5.3%)术中发生并矫正了电极尖端折叠。术后3个月言语识别得分在折叠组与无折叠组间无显著差异,听力保留率在两组间亦无显著差异(31.6% vs 52.5%,P = .11)。结论认为若术中识别并矫正折叠,言语表现似乎不受负面影响。

要点

  • 电极尖端折叠发生率为5.3%(35/663),均在术中被识别并矫正。
  • 术后3个月辅音-核-辅音(CNC)词识别得分在折叠组与无折叠组间无显著差异(44.9% vs 46.2%,P = .98)。
  • 听力保留候选者中,折叠组激活时听力保留率为31.6%(6/19),无折叠组为52.5%(31/59),差异无统计学意义(P = .11)。

核心结论

  • 尽管电极尖端折叠仍是临床关注的问题,但若术中识别并矫正,言语表现似乎不受负面影响。

研究设计与样本

  • 回顾性图表综述,纳入2016至2023年间接受细长型耳蜗周围电极阵列的663例人工耳蜗植入者,来自三级转诊中心。

主要发现

  • 663例植入中35例(5.3%)发生电极尖端折叠并在术中被识别和矫正。
  • 术后3个月CNC词识别得分:折叠组44.9%(SD 20.9%)vs 无折叠组46.2%(SD 21.0%),P = .98,无显著差异。
  • 术后3个月安静条件下AzBio得分:折叠组53.1%(SD 21.7%)vs 无折叠组60.8%(SD 28.0%),P = .26,无显著差异。
  • 术后3个月+10 dB信噪比条件下AzBio得分:折叠组19.1%(SD 23.7%)vs 无折叠组31.5%(SD 27.2%),P = .60,无显著差异。
  • 听力保留候选者中,折叠组19例中6例(31.6%)激活时听力保留,无折叠组59例中31例(52.5%)听力保留,P = .11,无显著差异。

关键数值与推荐

  • 低频纯音平均听阈(LFPTA)定义为125、250和500 Hz听阈平均值。
  • 听力保留候选资格定义为术前LFPTA < 60 dB HL。
  • 成功听力保留定义为激活时LFPTA < 80 dB HL。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: Cochlear implant (CI) electrode array tip fold-overs occur at an increased rate with perimodiolar electrode arrays, necessitating removal and re-insertion. The degree to which an intra-operative correction of tip fold-over affects CI performance and hearing preservation has not been previously reported. STUDY DESIGN: Retrospective chart review of CI recipients receiving a slim perimodiolar electrode array from 2016 to 2023. SETTING: Tertiary referral center. METHODS: Low-frequency pure tone average (LFPTA) was defined as the average of thresholds at 125, 250, and 500 Hz. We defined hearing preservation candidacy as LFPTA < 60 dB HL preoperatively, and successful hearing preservation was defined as LFPTA < 80 dB HL at activation. Consonant-nucleus-consonant (CNC) word recognition and AzBio scores in quiet and in +10 dB signal-to-noise ratio (SNR) were collected preoperatively, and at 3- and 6-months postoperatively. RESULTS: From 663 implants, 35 (5.3%) experienced tip fold-over that was identified and corrected intra-operatively. There was no significant difference in 3-month CNC scores between those with fold-overs (44.9%, SD 20.9%) and those without (46.2%, SD 21.0%; P = .98). Similarly, there was no difference in AzBio in quiet (53.1%, SD 21.7% vs 60.8%, SD 28.0%; P = .26) or in AzBio +10 dB SNR (19.1%, SD 23.7% vs 31.5%, SD 27.2%; P = .60). Of 19 hearing preservation candidates that experienced tip fold-over, 6 (31.6%) had preserved hearing at activation compared to 31 of 59 candidates (52.5%; P = .11) without fold-over. CONCLUSION: While tip fold-over remains a clinical concern, speech performance does not appear to be negatively affected if the fold-over is identified and corrected.

来源:Otolaryngol Head Neck Surg,查看原文。版权归原作者所有。

引用

  1. Smetak MR, Shew MA, Varghese J, et al. Speech Performance Following Intraoperative Correction of Cochlear Implant Electrode Array Tip Fold-Overs. Otolaryngol Head Neck Surg. 2025. PMID: 40052328. doi:10.1002/ohn.1199

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。