耳蜗电图潜伏期与人工耳蜗植入术中电极位置的相关性
本研究对30例人工耳蜗植入患者术中记录纯音耳蜗电图,提取耳蜗微音电位潜伏期(CML),并术后CT确定电极位置。结果显示CML与线性插入深度显著相关(p<0.001),但CML与残余听力无显著关联。
要点
- 术中耳蜗电图可提取耳蜗微音电位潜伏期(CML)。
- CML与电极线性插入深度显著相关(p<0.001)。
- CML与残余听力无显著关联。
核心结论
- 客观提取的CML可编码人工耳蜗患者耳蜗内电极位置,但与残余听力无直接关联。
- CML有潜力通过实时追踪电极位置来增强术中耳蜗电图分析。
- 为更好理解CML的个体间变异,未来需要更大患者队列的研究。
研究设计与样本
- 研究纳入30例人工耳蜗患者,术中在最大刺激强度下记录纯音耳蜗电图以提取CML。
- 电极插入过程中,在0.5和0.75 kHz两个刺激频率下连续记录耳蜗电图。
- 完全插入后,在所有偶数编号电极上以0.25、0.5、0.75和1 kHz频率测量耳蜗电图。
- 通过术后计算机断层扫描(CT)确定电极位置(即线性插入深度),并假设恒定插入速度反推插入期间测量电极的位置。
- 采用线性回归模型分析CML与线性插入深度的关系,并评估CML与术前残余听力的关系。
主要发现
- 在电极插入期间和插入后,CML与线性插入深度显著相关(p < 0.001),但0.25 kHz刺激条件下存在限制,推测是因为所用人工耳蜗电极阵列无法到达0.25 kHz特征区域。
- 尽管个体间变异较大,结果与其它研究中观察到的基底膜延迟一致。
- 未发现CML与残余听力之间存在显著关联。
局限性
- 个体间变异较大。
- 0.25 kHz刺激条件下CML与线性插入深度的相关性存在限制,推测是因为所用人工耳蜗电极阵列无法到达0.25 kHz特征区域。
- 为更好理解CML的个体间变异,未来需要更大患者队列的研究。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: Cochlear implant (CI) candidates increasingly exhibit some degree of residual hearing, which should be preserved despite the implantation. Today, cochlear health is monitored during CI surgery by tracking the cochlear microphonic (CM) amplitude from intracochlear electrocochleography (ECochG) measurements. However, recent studies indicate that the insertion depth of the measuring electrode must be considered to accurately interpret these signals. The acoustic path from the cochlear base to the apex induces excitation delays in deeper regions, which should be reflected in the CM measurements. In this study, we analyzed the potential of cochlear microphonic latency (CML) as an objective method for continuously tracking CI electrode position during cochlear implantation. In addition, we examined whether CML can be associated with residual hearing. DESIGN: We recorded intraoperative pure-tone ECochG at maximum stimulation levels from 30 CI patients to derive CML. During CI electrode insertion, ECochG was continuously recorded at the 2 stimulation frequencies of 0.5 and 0.75 kHz. After complete insertion, ECochG was measured on all evenly numbered electrodes at frequencies of 0.25, 0.5, 0.75, and 1 kHz. The electrode locations (i.e., linear insertion depth) were identified by postoperative computed tomography (CT) scans. The location of the measuring electrode during the insertion period was then calculated backward, assuming a constant insertion speed. Finally, we used a linear regression model to relate CML to linear insertion depth. In addition, we evaluated the relationship between CML and preoperative residual hearing. RESULTS: CML is significantly correlated to the linear insertion depth ( p < 0.001) during and after electrode insertion (with restrictions on 0.25 kHz stimulus, presumably since the characteristic 0.25 kHz region is not within reach of the used CI electrode arrays). Despite high inter-individual variability, our results align with documented delays in the basilar membrane observed in other studies. However, we could not identify a significant association between CML and residual hearing. CONCLUSIONS: Our study demonstrates that objectively extracted CML encodes the intracochlear electrode location in CI patients but is not directly linked to residual hearing. Consequently, CML has the potential to enhance intraoperative ECochG analysis by providing real-time tracking of electrode position. To better understand the inter-individual variations in CML, future studies with larger patient cohorts are needed.