文献速递耳科与耳神经

自验配助听器:起始响应与现场体验的影响

Self-Fitting Hearing Aids: Effects of Starting Response and Field Experience.
PubMed 文献 · Ear Hear · 2025-02-11

20名轻至中度听力损失的成年助听器使用者通过智能手机界面在安静条件下以65 dB SPL聆听预录语音,调整可穿戴主助听器的输出水平和频谱倾斜。比较基于个人听力图的处方起始响应与基于典型轻至中度听力图的通用起始响应在实验室自验配及短暂现场体验后的效果。结果显示起始响应和现场体验对实验室平均自验配输出无显著影响,但现场体验后高频自验配输出的个体起始响应效应标准差降低约50%,通用起始响应的调整步数更多,现场体验使自验配时间减少50%。

要点

  • 起始响应和现场体验对实验室平均自验配输出无显著影响(p=0.506和0.149)。
  • 现场体验后高频自验配输出的个体起始响应效应标准差降低约50%(p=0.006)。
  • 通用起始响应的调整步数多于处方起始响应(p=0.014),现场体验使自验配时间减少50%(p<0.001)。

核心结论

  • 研究结果支持以下结论:对于轻至中度听力损失人群,通用起始响应可作为非处方自验配助听器的一种可行选择。
  • 研究结果强调,新型自验配助听器需要练习和体验,且自验配可能并不适合所有人。

研究设计与样本

  • 研究纳入20名轻至中度听力损失的成年助听器使用者。
  • 参与者使用智能手机界面调整可穿戴主助听器的输出水平和频谱倾斜,聆听安静条件下以65 dB SPL呈现的预录语音。
  • 处方起始响应基于参与者自己的听力图,通用起始响应基于典型轻至中度听力损失的听力图且对所有参与者相同。
  • 两种起始响应的初始自验配在实验室进行;经过涉及对话、自听和环境噪声且按需重新调整的短暂现场体验后,在实验室重复两种起始响应的自验配。
  • 起始响应、自验配响应和调整步骤记录在主助听器中,用于后续评估真耳输出频谱和自验配行为。

主要发现

  • 起始响应和现场体验对实验室平均自验配输出均无显著影响(p=0.506和p=0.149)。
  • 现场体验后,个体起始响应对高频自验配输出效应的标准差降低约50%(p=0.006)。
  • 起始响应对自验配行为的影响仅限于调整步数,通用起始响应的调整步数更多(p=0.014)。
  • 现场体验对自验配行为的影响仅限于自验配时间减少50%(p<0.001)。
  • 自验配时间减少主要归因于每次调整步骤后聆听时间减少(p=0.019)。

临床怎么用

  • 作者认为,对于轻至中度听力损失人群,通用起始响应可作为非处方自验配助听器的一种可行选择。
  • 作者指出,新型自验配助听器需要练习和体验,且自验配可能并不适合所有人。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: To determine the effects of changing from a prescribed to a generic starting response on self-fitting outcome and behavior before and after a brief field experience. METHOD: Twenty adult hearing-aid users with mild-to-moderate hearing loss used a smartphone interface to adjust level and spectral tilt of the output of a wearable master hearing aid while listening to prerecorded speech, presented at 65 dB SPL, in quiet. A prescribed starting response was based on the participant's own audiogram. A generic starting response was based on an audiogram for a typical mild-to-moderate hearing loss and was the same for all participants. Initial self-fittings from the two starting responses took place in the lab. After a brief field experience, involving conversation, self-hearing, and ambient noise, with readjustment as needed, self-fittings from the two starting responses were repeated in the lab. Starting responses, self-fitted responses, and adjustment steps were logged in the master hearing aid for subsequent evaluation of real-ear output spectra and for assessment of self-fitting behavior. RESULTS: Neither starting response nor field experience had a significant effect on mean self-fitted output in the lab ( p = 0.506 and 0.149, respectively). However, the SD of individual starting-response effects on high-frequency self-fitted output fell by around 50% after the field experience ( p = 0.006). The effect of starting response on self-fitting behavior was limited to number of adjustment steps, which was higher for the generic start ( p = 0.014). The effect of field experience on self-fitting behavior was limited to a 50% reduction in self-fitting time ( p < 0.001). This reduction was attributable mainly to less time spent listening after each adjustment step ( p = 0.019). CONCLUSIONS: The findings support the conclusion that, for a population with mild-to-moderate hearing loss, a generic starting response can be a viable option for over-the-counter self-fitting hearing aids. They highlight, however, the need for practice and experience with novel self-fitting hearing aids and the fact that self-fitting may not be suitable for all.

来源:Ear Hear,查看原文。版权归原作者所有。

引用

  1. Boothroyd A, Sengupta D, Painter S, et al. Self-Fitting Hearing Aids: Effects of Starting Response and Field Experience. Ear Hear. 2025. PMID: 39930570. doi:10.1097/AUD.0000000000001642

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。