应用透视与耳蜗内压力在尸体标本中表征电极尖端折叠
研究在尸体人头标本上经圆窗插入围蜗轴电极(Cochlear Slim-Modiolar, CI532),用光纤压力传感器测量耳蜗内压力并在透视下观察,发现尖端折叠至少分三型:前后向C形、内外向C形和S形翻转;前后向与S形翻转伴随最大瞬态压力,与电极在耳蜗内旋转或扭转相关。
要点
- 尖端折叠至少分三型:前后向C形、内外向C形和S形翻转。
- 前后向与S形翻转时出现最大瞬态压力,与电极在耳蜗内扭转相关。
- 作者认为将尖端折叠分型可能有助于未来术中识别尖端折叠事件。
核心结论
- 结果证明尖端折叠至少存在三种亚型,电极扭转相关的瞬态压力升高出现在尖端折叠事件之前和期间。
- 作者认为将尖端折叠分型是新颖的,可能有助于未来术中识别尖端折叠事件。
- 作者认为识别尖端折叠事件仍然重要,应使用多模式验证系统早期识别。
- 作者认为仍需进一步研究以确定这些变化的意義及其他可能的耳蜗内电极运动模式。
研究设计与样本
- 尸体研究:在尸体人头标本上行乳突切开术和面隐窝入路,将光纤压力传感器置入前庭阶和鼓阶测量耳蜗内压力。
- 经圆窗在透视下插入围蜗轴人工耳蜗电极(Cochlear Slim-Modiolar, CI532)。
- 证据等级:NA(尸体研究)。
主要发现
- 观察到三种尖端折叠事件:前后向C形、内外向C形和S形翻转。
- 最大瞬态压力出现在前后向和S形翻转时,并与耳蜗内旋转或扭转相关。
- 瞬态压力升高出现在尖端折叠事件之前和期间,与电极扭转相关。
临床怎么用
- 作者认为识别尖端折叠事件仍然重要,应使用多模式验证系统早期识别。
局限性
- 作者认为仍需进一步研究以确定这些变化的意義及其他可能的耳蜗内电极运动模式。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: Cochlear implant array malpositioning is associated with impaired speech perception, vertigo, and facial nerve stimulation. Tip fold-over is a subset of malpositioning that occurs more often with perimodiolar electrodes, but historically it has not been characterized due to lack of knowledge regarding electrode movements of the electrode within the cochlea. The aim of this study was to characterize the mechanics of tip fold-over events and their associated insertion pressure profiles. METHODS: Cadaveric human heads were surgically prepared with a mastoidectomy and facial recess. Fiberoptic pressure sensors were inserted into the scala vestibuli and tympani to measure intracochlear pressures. Perimodiolar CI electrodes (Cochlear Slim-Modiolar, CI532) were inserted via round window under fluoroscopy. RESULTS: Three types of tip fold-over events were observed: anterior-posterior C-shaped, medial-lateral C-shaped, and S-shaped roll-overs. The largest transient pressures occurred with anterior-posterior and S-type roll-over, and were associated with rotation or twisting inside the cochlea. CONCLUSIONS: Results demonstrate at least three subtypes of tip fold-overs. Elevated pressure transients were noted before and during the tip fold-over event related to electrode twisting. The characterization of tip fold-over into subtypes is novel and may aid identification of tip fold-over events intraoperatively in the future. It remains important to identify tip fold-over events, and they should be recognized early using a multimodal verification system. Further investigation is still required to determine the significance of these changes and other possible patterns of intracochlear electrode movement. LEVEL OF EVIDENCE: NA: Cadaver study Laryngoscope, 135:1795-1802, 2025.
引用
- Chabuz CA, Banakis Hartl RM, Rodriguez K, et al. Characterization of Tip Fold-Over Using Fluoroscopy and Intracochlear Pressure in Cadaver Specimens. Laryngoscope. 2024. PMID: 39719824. doi:10.1002/lary.31977