文献速递耳科与耳神经

人工耳蜗是否应禁忌经颅电刺激?一例具有安全性启示的病例研究

Should Cochlear Implant Contraindicate Transcranial Stimulation? A Case Study with Safety Implications.
PubMed 文献 · Laryngoscope · 2024-11-20

本文报告一例人工耳蜗(CI)患者连续两次脊柱侧凸矫正手术的病例,第二次手术中经颅电刺激神经监测提示疑似即将发生的神经损伤;术后听力学与电学检查分别显示听觉功能保留、CI功能正常。结合文中复习的既往报告,经颅电刺激神经监测和电休克治疗后未见CI故障或听力下降,作者认为亟需对现行禁忌证进行审查并更新数据驱动的风险分析。

要点

  • CI患者两次脊柱侧凸矫正术中行经颅电刺激神经监测,术后听觉功能保留、CI功能正常。
  • 文中复习的既往报告显示经颅电刺激神经监测和电休克治疗后无CI故障或听力下降。
  • 作者认为亟需制造商与监管机构主动对现行禁忌证开展全面审查并更新风险分析。

核心结论

  • 本病例与文中复习的既往报告一致,经颅电刺激神经监测和电休克治疗后未见CI故障或听力下降。
  • 鉴于不断积累的证据,作者认为亟需制造商和监管机构主动对现行禁忌证进行全面审查,并更新数据驱动的风险分析以权衡阻止这些禁忌工具和治疗的后果及其对患者的潜在获益。

研究设计与样本

  • 本文为病例研究,报告一例CI患者连续两次因脊柱侧凸矫正接受手术,第二次手术中经颅电刺激神经监测提示疑似即将发生的神经损伤。

主要发现

  • 第二次手术中经颅电刺激神经监测提示疑似即将发生的神经损伤。
  • 每次术后听力学检查显示听觉功能保留,电学检查显示CI功能正常。
  • 文中复习的既往报告显示经颅电刺激神经监测和电休克治疗后无CI故障或听力下降。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

Electrical stimulation in proximity to a cochlear implant (CI) is contraindicated due to the potential risk of damaging the implant and thermal insult to the inner ear. Yet, during surgical procedures that pose a risk to the nervous system, the use of transcranial electrical stimulation for neuromonitoring is crucial for assessing the integrity of the motor pathways and preventing neurological insults. The current paper depicts a case of a patient with CI undergoing two consecutive surgeries for scoliosis correction, revealing a suspected pending neurological insult in the second operation. An audiological and electrical examination after each operation demonstrated preserved auditory function and normal functioning of the CI, respectively. The current case is consistent with previous reports reviewed in the paper, showing no CI failures nor hearing deterioration following transcranial electrical stimulation for neuromonitoring and electroconvulsive therapy. In light of this accumulating evidence, there is an urgent need for manufacturers and regulatory bodies to proactively undertake a comprehensive review of this declared contraindication. Subsequently, an updated data-driven risk analysis is essential to weigh the consequences of preventing the contraindicated tools and treatments, as well as their potential benefits to these patients. Laryngoscope, 135:1493-1495, 2025.

来源:Laryngoscope,查看原文。版权归原作者所有。

引用

  1. Zarchi O, Amitai A, Attias J, et al. Should Cochlear Implant Contraindicate Transcranial Stimulation? A Case Study with Safety Implications. Laryngoscope. 2024. PMID: 39564733. doi:10.1002/lary.31882

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。