文献速递耳科与耳神经小儿耳鼻咽喉

X连锁不完全分隔III型耳聋人工耳蜗植入后对电刺激的敏感性进行性丧失

Progressive Loss of Sensitivity to Electrical Stimulation After Cochlear Implantation in X-Linked Incomplete Partition Type III Deafness.
PubMed 文献 · Ear Hear · 2024-09-12

回顾性分析13例X连锁不完全分隔III型(IP3)畸形患者与正常解剖对照组的阻抗、刺激水平、电极停用和言语感知。IP3组最大舒适响度(MCL)每年增加2.5电荷单位(qu),对照组未见此现象;电极停用更常见,估计25%电极在15岁时停用;言语感知普遍较低且与停用电极数相关。

要点

  • IP3畸形患者MCL水平每年增加2.5 qu,对照组未见此现象。
  • IP3畸形电极停用更常见,估计25%电极在15岁时停用。
  • IP3畸形言语感知普遍较低,且与停用电极数相关。

核心结论

  • IP3畸形耳聋患者较正常解剖者可能面临更高的人工耳蜗停用风险。
  • 对电刺激的敏感性进行性丧失可能提示异常耳蜗内一种神经退化形式。
  • 随着时间推移,即使使用人工耳蜗技术,该组患者也可能出现言语感知逐渐恶化。

研究设计与样本

  • 回顾性分析临床数据库中全部13例IP3畸形患者的阻抗、刺激水平、停用电极和言语感知,并纳入正常解剖对照组。

主要发现

  • MCL水平随时间每年增加2.5电荷单位(qu),对照组未见此现象。
  • 电极停用在IP3畸形中更常见,估计25%的电极在15岁时停用。
  • 阻抗稳定但研究人群较高。
  • IP3畸形言语感知普遍较低,且与停用电极数相关。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: Patients with X-linked incomplete partition type III (IP3) deafness treated with cochlear implants exhibit higher "Most Comfortable Loudness" (MCL) levels of stimulation and more electrode deactivation than patients with normal morphology. We endeavored to analyze the progression of the MCL levels and electrode deactivation over time and assess those factors that could have led to deactivation. Furthermore, we aimed to assess whether speech perception was affected by a progressive loss of neural contact. DESIGN: All 13 patients with the IP3 malformation in our clinical database were analyzed retrospectively with regard to impedance, stimulation levels, deactivated electrodes, and speech perception. A control group of patients with normal anatomy was included. RESULTS: MCL levels increased over time by 2.5 charge units (qu) per year, which was not seen in the control group. Electrode deactivation was more common in IP3 malformation, and it was estimated that 25% of electrodes would be deactivated by 15 years of age. Impedance was stable but higher in the study population. Speech perception was lower in IP3 malformation generally and was correlated to the number of deactivated electrodes. CONCLUSIONS: Patients diagnosed with IP3 malformation deafness may suffer a greater risk of cochlear implant discontinuation compared with those with normal anatomy. A progressive loss of sensitivity to electrical stimulation may indicate a form of neural degradation in the abnormal cochlea. With time, patients in this group, even with cochlear implant technology, may experience gradual deterioration of speech perception. This has clinical implications for the counseling of parents.

来源:Ear Hear,查看原文。版权归原作者所有。

引用

  1. Eklöf M, Smeds H, Karltorp E, et al. Progressive Loss of Sensitivity to Electrical Stimulation After Cochlear Implantation in X-Linked Incomplete Partition Type III Deafness. Ear Hear. 2024. PMID: 39261990. doi:10.1097/AUD.0000000000001584

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。