新型雷达设备诊断阻塞性睡眠呼吸暂停:一项评估其与多导睡眠监测一致性的平行对照研究
2023年7月至9月在上海第六人民医院耳鼻咽喉头颈外科睡眠实验室招募155名受试者,145名纳入最终分析,同时接受QSA600毫米波雷达睡眠监测系统与多导睡眠监测(PSG)。以呼吸暂停低通气指数(AHI)5、15、30次/h为阈值,QSA600诊断轻、中、重度OSA的灵敏度分别为88.68%、89.83%、97.06%,特异度分别为94.87%、98.84%、99.10%,ROC曲线下面积分别为0.973 4、0.990 9、0.999 5;AHI的Pearson相关系数为0.987 2(P<0.001),Bland-Altman差值均数为-1.43次/h(95%CI:-8.74~5.88),ICC为0.985 0(95%CI:0.975 4~0.990 4)。结论认为QSA600与PSG在诊断OSA及严重程度分级方面表现出很高的一致性,有较好临床应用前景。
要点
- 以AHI 5、15、30次/h为阈值,QSA600诊断轻、中、重度OSA的灵敏度分别为88.68%、89.83%、97.06%。
- 对应特异度分别为94.87%、98.84%、99.10%,ROC曲线下面积分别为0.973 4、0.990 9、0.999 5。
- QSA600与PSG的AHI Pearson相关系数0.987 2(P<0.001),ICC 0.985 0(95%CI:0.975 4~0.990 4)。
核心结论
- QSA600作为低负荷新型睡眠监测系统,在诊断OSA及不同严重程度分级方面与PSG表现出很高的一致性,有较好临床应用前景。
研究设计与样本
- 2023年7月至9月在上海市第六人民医院耳鼻咽喉头颈外科睡眠实验室开展平行对照一致性研究,招募155名受试者,145名纳入最终分析,男75例、女70例,年龄(35.30±12.41)岁,身高(168.23±8.08)cm,体重(68.28±13.74)kg。
- 受试者同时接受PSG和QSA600监测,按AHI分为4组:<5.0次/h(NOSA组,39例)、≥5.0~<15.0次/h(轻度OSA组,47例)、≥15.0~<30.0次/h(中度OSA组,25例)、≥30.0次/h(重度OSA组,34例)。
- 采用组内相关系数(ICC)、Pearson相关系数和Bland-Altman分析评价两种监测技术AHI等指标的一致性,并检验QSA600在不同AHI阈值下的诊断灵敏度与特异度,统计学分析采用MATLAB R2022a。
- 临床试验注册号为NCT06038006。
主要发现
- 以AHI 5、15、30次/h为阈值,QSA600诊断轻度、中度、重度OSA的灵敏度分别为88.68%、89.83%、97.06%。
- 对应特异度分别为94.87%、98.84%、99.10%,ROC曲线下面积分别为0.973 4、0.990 9、0.999 5。
- QSA600与PSG的AHI测量值Pearson相关系数为0.987 2(P<0.001)。
- 二者Bland-Altman测定差值均数为-1.43次/h(95%CI:-8.74~5.88),ICC为0.985 0(95%CI:0.975 4~0.990 4)。
临床怎么用
- 原文结论认为QSA600与PSG在诊断OSA及严重程度分级方面一致性很高,有较好临床应用前景,未据此给出具体推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
Objective: This study evaluates the agreement between a new low-load sleep monitoring system, QSA600, based on millimeter-wave radar technology, and polysomnography (PSG) in diagnosing obstructive sleep apnea (OSA). Methods: A total of 155 subjects were recruited for a parallel agreement study in the sleep laboratory of the Department of Otorhinolaryngology Head and Neck Surgery at Shanghai Sixth People's Hospital from July to September 2023. The subjects underwent simultaneous monitoring with both PSG and the QSA600 system. One hundred and forty-five subjects consisting of 75 males and 70 females included in the final analysis, with an average age of (35.30±12.41) years, an average height of (168.23±8.08) cm, and an average weight of (68.28±13.74) kg. The subjects were divided into four groups based on the apnea-hypopnea index (AHI): <5.0 events/h (non-OSA group, 39 cases), ≥5.0-<15.0 events/h (mild OSA group, 47 cases), ≥15.0-<30.0 events/h (moderate OSA group, 25 cases), and≥30.0 events/h (severe OSA group, 34 cases). Intraclass correlation coefficients (ICC), Pearson correlation coefficients (r), and Bland-Altman analysis were employed to assess the agreement between the two monitoring techniques regarding AHI and other parameters. Sensitivity and specificity of the QSA600 in diagnosing OSA were evaluated at different AHI thresholds. Statistical analyses were conducted using MATLAB R2022a. Results: Using AHI 5 events/h, 15 events/h and 30 events/h as thresholds, the sensitivity for diagnosing mild, moderate, and severe OSA was 88.68%, 89.83% and 97.06%, respectively. The specificity was 94.87%, 98.84% and 99.10%, respectively. The areas under the receiver operating characteristic (ROC) curve was 0.973 4, 0.990 9 and 0.999 5, respectively. The comparison of key indicators between QSA600 and PSG diagnostic results revealed:a Pearson correlation coefficient of 0.987 2(P<0.001) between the AHI measurement values. The mean difference between the Bland-Altman measurement values of the two was -1.43(95%CI:-8.74-5.88) events/h and the ICC between the two was 0.985 0(95%CI: 0.975 4-0.990 4). Conclusions: As a new low-load sleep monitoring system, QSA600 demonstrates high concordance with traditional PSG in diagnosing OSA and stratifying its severity, which has promising potential for clinical application. (Clinical trial registration number: NCT06038006). 目的: 评价基于毫米波雷达的低负荷新型睡眠监测系统QSA600与多导睡眠监测(polysomnography,PSG)在诊断阻塞性睡眠呼吸暂停(obstructive sleep apnea,OSA)的一致性。 方法: 2023年7月至9月招募155名受试者,在上海市第六人民医院耳鼻咽喉头颈外科睡眠实验室同时接受PSG和QSA600监测,145名受试者纳入最终分析,其中男75例,女70例,年龄为(35.30±12.41)岁,身高为(168.23±8.08)cm,体重为(68.28±13.74)kg,进行一致性平行对照研究。以呼吸暂停低通气指数(apnea hypopnea index,AHI)将受试者分为4组:<5.0 次/h(NOSA组,39例),≥5.0次/h~<15.0 次/h(轻度OSA组,47例),≥15.0次/h~<30.0 次/h(中度OSA组,25例),≥30.0 次/h(重度OSA组,34例)。采用组内相关系数(intraclass correlation coefficients,ICC)、Pearson相关系数(r)和Bland-Altman检验来评价2种监测技术AHI等指标的一致性。根据AHI不同阈值,检验QSA600的诊断灵敏度、特异度。采用MATLAB R2022a进行统计学分析。 结果: 分别以AHI 5 次/h、15 次/h、30 次/h为阈值,对轻度、中度、重度OSA诊断的灵敏度分别为88.68%、89.83%、97.06%,特异度分别为94.87%、98.84%、99.10%,受试者工作特征(receiver operating characteristic,ROC)曲线下面积分别为0.973 4、0.990 9、0.999 5。QSA600与PSG诊断结果的关键指标(AHI)的一致性对比结果:二者Pearson相关性为0.987 2(P<0.001);二者Bland-Altman测定差值均数为-1.43次/h[95%置信区间(confidence interval,CI)为-8.74~5.88];二者ICC为0.985 0(95%CI为0.975 4~0.990 4)。 结论: 作为低负荷新型睡眠监测系统,QSA600在诊断OSA及不同严重程度OSA分级方面与PSG表现出很高的一致性,有较好临床应用前景。.
引用
- Li CY, Wang W, Huang WJ, et al. [Diagnosis of obstructive sleep apnea by a new radar device: a parallel controlled study evaluating agreement with polysomnographic monitoring]. Zhonghua Er Bi Yan Hou Tou Jing Wai Ke Za Zhi. 2024. PMID: 39193596. doi:10.3760/cma.j.cn115330-20231204-00267
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。