文献速递睡眠呼吸障碍

舌下神经刺激患者中家庭睡眠设备与睡眠检查测试的比较

Comparison of Home Sleep Devices and Sleep Study Testing in Hypoglossal Nerve Stimulation Patients.
PubMed 文献 · Laryngoscope · 2024-08-20

本研究前瞻性纳入5个中心60例接受舌下神经刺激(HGNS)治疗的患者,随访6个月,比较睡眠监测垫与NightOwl外周动脉张力(PAT)传感器在植入后1至6个月家庭监测的呼吸暂停低通气指数(AHI)与实验室多导睡眠图(PSG)或家庭睡眠呼吸暂停测试(HSAT)的一致性。睡眠监测垫识别AHI<15的敏感度和特异度分别为61%和82%,识别AHI<30分别为77%和100%;PAT传感器识别AHI<15分别为57%和77%,识别AHI<30分别为81%和80%。结论认为两种家庭设备在HGNS治疗患者中检测轻度和中度OSA具有高敏感度和特异度,可能实现家庭环境下纵向客观监测HGNS疗效。

要点

  • 睡眠监测垫识别AHI<15的敏感度61%、特异度82%,识别AHI<30的敏感度77%、特异度100%。
  • PAT传感器识别AHI<15的敏感度57%、特异度77%,识别AHI<30的敏感度81%、特异度80%。
  • 两种家庭设备在HGNS治疗患者中检测轻度和中度OSA具有高敏感度和特异度。

核心结论

  • 睡眠监测垫和PAT传感器在HGNS治疗患者中检测轻度和中度OSA表现出高敏感度和特异度,可能实现家庭环境下纵向客观监测HGNS疗效。

研究设计与样本

  • 前瞻性纳入5个中心60例接受HGNS治疗的患者,随访6个月。
  • 患者植入后1至6个月使用睡眠监测垫和NightOwl PAT传感器进行家庭AHI纵向监测。
  • 植入后3个月和6个月使用实验室PSG或HSAT评估疗效,并将两种家庭设备的AHI与PSG和HSAT比较以评估OSA严重程度识别的准确性。

主要发现

  • 睡眠监测垫识别AHI<15的敏感度61%、特异度82%,识别AHI<30的敏感度77%、特异度100%。
  • PAT传感器识别AHI<15的敏感度57%、特异度77%,识别AHI<30的敏感度81%、特异度80%。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: Hypoglossal nerve stimulation (HGNS) is an implantable therapy for obstructive sleep apnea (OSA). Therapy efficacy is currently confirmed by a formal sleep study after empiric adjustment by the patient at home based on their subjective experience with the device. Home-based longitudinal apnea hypopnea index (AHI) measurements have the potential to refine HGNS therapeutic amplitude selection with objective data. Our objective was to compare AHI derived from routine sleep studies and two different home sleep devices in new HGNS recipients. METHODS: Prospectively enrolled patients receiving HGNS therapy were provided a Sleep Tracking Mat (Withings, Issy-les-Moulineaux, France) and NightOwl peripheral arterial tonometry (PAT) sensor (Ectosense, Leuven, Belgium) for longitudinal, home AHI monitoring from 1 to 6 months post-implant. Therapy efficacy was assessed at 3 and 6 months post-implant using in-lab polysomnography (PSG) or home sleep apnea test (HSAT). The sleep mat and PAT sensor AHI were compared against PSG and HSAT for accuracy of OSA severity identification. RESULTS: Sixty patients were enrolled across 5 centers and followed for 6 months. The sleep mat had sensitivity and specificity for identifying AHI <15 of 61% and 82% and AHI <30 of 77% and 100%. The PAT device had sensitivity and specificity for identifying AHI <15 of 57% and 77% and AHI <30 of 81% and 80%. CONCLUSIONS: The sleep mat and PAT sensor demonstrated high sensitivity and specificity for detection of mild and moderate OSA in patients with HGNS therapy and may enable longitudinal objective monitoring of HGNS efficacy in the home setting. LEVEL OF EVIDENCE: 3 Laryngoscope, 135:469-477, 2025.