验证人工耳蜗植入中用于耳蜗电图的诱发电位平台
在三级转诊中心开展术中验证研究,纳入30例成人与儿童人工耳蜗植入患者,比较智能听力系统Duet诱发电位仪与Cochlear公司耳蜗电图平台的术中耳蜗外耳蜗电图。两种平台的总反应呈强相关(r=0.905,95%CI 0.809–0.954);4例经鼓膜耳蜗电图可行性评估获得相似反应。
要点
- 30例患者(平均年龄67岁,SD 18.8)术中同时用IHS与Cochlear平台记录耳蜗外耳蜗电图
- 两平台总反应强相关:r=0.905(95%CI 0.809–0.954)
- 4例经鼓膜IHS耳蜗电图可行性评估获得相似反应
核心结论
- 耳蜗外耳蜗电图已被用于预测人工耳蜗植入者的言语感知表现。
- IHS Duet系统是测量人工耳蜗植入人群耳蜗外耳耳蜗电图的有效方法。
- 未来工作将利用该系统测量经鼓膜耳蜗电图以改善术前对人工耳蜗植入表现的估计。
研究设计与样本
- 术中验证研究,在三级转诊中心开展。
- 纳入成人与儿童人工耳蜗植入患者,共30例,平均年龄67岁(SD 18.8)。
- 术中同时使用智能听力系统(IHS)Duet诱发电位仪与Cochlear公司(CC)耳蜗电图平台记录耳蜗电图。
- 记录电极经标准面隐窝进行耳蜗外测量,给予250 Hz至2 kHz(约110 dB SPL)短纯音。
- 对关键频率耳蜗电图波形行快速傅里叶变换(FFT)并求和为总反应(ECochG-TR)。
主要发现
- 植入耳术前平均纯音平均听阈(PTA;0.5、1、2、4 kHz)为87.4 dB HL(SD 19.3),术前平均言语识别率(WRS)为17.0%正确(SD 19.1)。
- IHS-ECochG-TR中位数2.30 μV(范围0.1–148.26),CC-ECochG-TR中位数3.00 μV(范围0.1–239.63)。
- IHS-ECochG-TR与CC-ECochG-TR呈强相关(r=0.905,95%CI 0.809–0.954)。
- 4例患者使用IHS系统行经鼓膜耳蜗电图可行性评估,获得相似反应。
关键数值与推荐
- 证据等级:3级。
- 发表于Laryngoscope,2025年,135卷308–315页。
临床怎么用
- IHS Duet系统可作为人工耳蜗植入人群耳蜗外耳蜗电图的有效测量工具。
- 作者提出未来将利用该系统测量经鼓膜耳蜗电图,以改善术前对人工耳蜗植入表现的估计。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: To validate electrocochleography (ECochG) between an auditory evoked potential (AEP) machine and an established cochlear implant (CI) manufacturer ECochG system. METHODS: Intraoperative validation study at a tertiary referral center. Patients included adults and children undergoing cochlear implantation. Intraoperative ECochG was measured with both the Intelligent Hearing Systems (IHS) Duet AEP machine and Cochlear Corporation (CC) ECochG platform. Recording electrodes captured extracochlear measurements through a standard facial recess. Tone-bursts were presented from 250 Hz to 2 kHz (~110 dB SPL). A fast Fourier transform (FFT) of ECochG waveforms at key frequencies was summed into a total response (ECochG-TR). Pearson's correlation was utilized to evaluate the relationship between IHS-ECochG-TR and CC-ECochG-TR after confirming normality. RESULTS: Thirty patients were enrolled with an average age of 67 years (SD 18.8). In the ear that was implanted, mean preoperative pure-tone average (PTA; 0.5, 1, 2, and 4 kHz) was 87.4 dB HL (SD 19.3) and mean preoperative word-recognition scores (WRS) was 17.0% correct (SD 19.1). There was strong correlation (r = 0.905, 95% confidence interval: 0.809 to 0.954) between IHS-ECochG-TR (median 2.30 μV, range 0.1-148.26) and CC-ECochG-TR (median 3.00 μV, range 0.1-239.63). Four patients underwent transtympanic ECochG with the IHS system for feasibility evaluation and achieved similar responses. CONCLUSION: Extracochlear ECochG has been predictive of CI speech perception performance. The IHS duet system is a valid measure of extracochlear ECochG for the CI population. Future work will utilize this system for measuring transtympanic ECochG to improve preoperative estimation of CI performance. LEVEL OF EVIDENCE: 3 Laryngoscope, 135:308-315, 2025.
引用
- Varghese JJ, Shew MA, Walia A, et al. Validating an Evoked Potential Platform for Electrocochleography During Cochlear Implantation. Laryngoscope. 2024. PMID: 39189299. doi:10.1002/lary.31724
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。