文献速递耳科与耳神经

成人人工耳蜗效果早期进展的障碍因素

Barriers to Early Progress in Adult Cochlear Implant Outcomes.
PubMed 文献 · Ear Hear · 2024-08-12

该单中心横断面回顾性研究纳入32例成人单侧Nucleus人工耳蜗使用者,用简短测试组套评估外周与神经认知限制因素对安静及噪声下句子识别进展的影响。无限制因素者开机1个月安静下句子识别≥90%的可能性高8.5倍(p=0.010)。早期成绩可预测长期噪声下表现。

要点

  • 无外周或神经认知限制因素者开机1个月安静下句子识别≥90%的可能性高8.5倍(p=0.010)。
  • 32例中检出外周限制因素4例、神经认知限制因素14例。
  • 开机1个月安静下句子识别≥90%可定义为进展成功,较低分提示需诊断性测试与持续康复。

核心结论

  • 外周与神经认知加工因素均影响人工耳蜗开机后的早期句子识别。
  • 外周限制因素检出较少,可能因其采用基于样本的标准而神经认知因素采用基于正常人群的标准。
  • 早期表现总体可预测长期表现。
  • 开机1个月安静下句子识别≥90%可用于定义进展成功,较低分提示需诊断性测试与持续康复。
  • 开机后1天的句子测试成绩即可为新使用者提供重要信息并提示康复随访需求。

研究设计与样本

  • 单中心、横断面、回顾性设计,纳入32例27至90岁(均值70,标准差13.5)成人单侧Nucleus人工耳蜗使用者。
  • 术后在1天、1至2个月及最长2年评估安静与不同信噪比下的句子识别成绩。
  • 本诊所标准测试组套包括生理学与神经认知测量:电诱发复合动作电位恢复功能、兴奋扩散与极性效应;蒙特利尔认知评估筛查、斯特鲁普色词测验及16岁以上成人法语阅读技能评估第3和第4项。

主要发现

  • 开机1个月时32例中13例安静下句子识别≥90%,与年龄无显著依赖关系。
  • 无外周或神经认知限制因素者开机1个月安静下句子识别≥90%的可能性高8.5倍(p=0.010)。
  • 32例中检出4例外周限制因素,与顶极和底极位置神经健康或电极-神经界面不良有关。
  • 32例中检出14例神经认知限制因素。
  • 早期句子识别成绩可预测长期噪声下句子识别阈值,限制因素似乎持续产生影响。

关键数值与推荐

  • 生理学评分判定为异常即限制性:顶极和底极位置总神经恢复期与极性效应均>1.65个标准差(相对人群均值)。
  • 兴奋扩散>6个电极单位亦视为限制性。
  • 神经认知测试成绩差于已发表正常人群均值1.65个标准差视为限制性。
  • 开机1个月安静下句子识别≥90%定义为进展成功。

临床怎么用

  • 作者认为开机1个月安静下句子识别≥90%可用于定义进展成功,较低分提示需诊断性测试与持续康复。
  • 作者认为开机后1天的句子测试成绩可为新人工耳蜗使用者提供重要信息并提示康复随访需求。
  • 作者认为了解限制人工耳蜗表现的可测量协变量可能有助于随访与咨询。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: Adult cochlear implant (CI) recipients obtain varying levels of speech perception from their device. Adult CI users adapt quickly to their CI if they have no peripheral "bottom-up" or neurocognitive "top-down" limiting factors. Our objective here was to understand the influence of limiting factors on the progression of sentence understanding in quiet and in noise, initially and over time. We hypothesized that the presence of limiting factors, detected using a short test battery, would predictably influence sentence recognition with practical consequences. We aimed to validate the test battery by comparing the presence of limiting factors and the success criteria of >90% sentence understanding in quiet 1 month after activation. DESIGN: The study was a single-clinic, cross-sectional, retrospective design incorporating 32 adult unilateral Nucleus CI users aged 27 to 90 years (mean = 70, SD = 13.5). Postoperative outcome was assessed through sentence recognition scores in quiet and in varying signal to noise ratios at 1 day, 1 to 2 months, and up to 2 years. Our clinic's standard test battery comprises physiological and neurocognitive measures. Physiological measures included electrically evoked compound action potentials for recovery function, spread of excitation, and polarity effect. To evaluate general cognitive function, inhibition, and phonological awareness, the Montreal Cognitive Assessment screening test, the Stroop Color-Word Test, and tests 3 and 4 of the French Assessment of Reading Skills in Adults over 16 years of age, respectively were performed. Physiological scores were considered abnormal, and therefore limiting, when total neural recovery periods and polarity effects, for both apical and basal electrode positions, were >1.65 SDs from the population mean. A spread of excitation of >6 electrode units was also considered limiting. For the neurocognitive tests, scores poorer than 1.65 SDs from published normal population means were considered limiting. RESULTS: At 1 month, 13 out of 32 CI users scored ≥90% sentence recognition in quiet with no significant dependence on age. Subjects with no limiting peripheral or neurocognitive factors were 8.5 times more likely to achieve ≥90% score in quiet at 1 month after CI switch-on ( p = 0.010). In our sample, we detected 4 out of 32 cases with peripheral limiting factors that related to neural health or poor electrode-neural interface at both apical and basal positions. In contrast, neurocognitive limiting factors were identified in 14 out of 32 subjects. Early sentence recognition scores were predictive of long-term sentence recognition thresholds in noise such that limiting factors appeared to be of continuous influence. CONCLUSIONS: Both peripheral and neurocognitive processing factors affect early sentence recognition after CI activation. Peripheral limiting factors may have been detected less often than neurocognitive limiting factors because they were defined using sample-based criteria versus normal population-based criteria. Early performance was generally predictive of long-term performance. Understanding the measurable covariables that limit CI performance may inform follow-up and improve counseling. A score of ≥90% for sentence recognition in quiet at 1 month may be used to define successful progress; whereas, lower scores indicate the need for diagnostic testing and ongoing rehabilitation. Our findings suggest that sentence test scores as early as 1 day after activation can provide vital information for the new CI user and indicate the need for rehabilitation follow-up.

来源:Ear Hear,查看原文。版权归原作者所有。

引用

  1. Marx M, Laborde ML, Algans C, et al. Barriers to Early Progress in Adult Cochlear Implant Outcomes. Ear Hear. 2024. PMID: 39129126. doi:10.1097/AUD.0000000000001559

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。