残余声听人工耳蜗使用者中相位间间隔偏移效应作为神经健康指标的研究

The Inter-Phase Gap Offset Effect as a Measure of Neural Health in Cochlear Implant Users With Residual Acoustic Hearing.
PubMed 文献 · Ear Hear · 2024-07-26

该前瞻性研究纳入17名在苏黎世大学医院接受人工耳蜗手术、500 Hz存在残余声听的成人,于术中及术后4至12周记录电诱发复合动作电位阈值、相位间间隔偏移效应、耳蜗电图及阻抗。结果显示相位间间隔偏移效应与术中电诱发复合动作电位阈值正相关(p < 0.001),与术前纯音平均听阈无显著相关(p = 0.999);耳蜗电图振幅下降的电极组相位间间隔偏移效应术后减小更显著,t(17)=2.76,p=0.014;电极阻抗变化对相位间间隔偏移效应无系统性影响(p = 0.263)。作者认为临床解读该指标需谨慎。

要点

  • 相位间间隔偏移效应与术中电诱发复合动作电位阈值正相关(p < 0.001)。
  • 相位间间隔偏移效应与术前纯音平均听阈无显著相关(p = 0.999)。
  • 耳蜗电图振幅下降的电极组相位间间隔偏移效应术后减小更显著(p = 0.014)。

核心结论

  • 研究未发现相位间间隔偏移效应与电诱发复合动作电位阈值或术前声听之间存在预期关系。
  • 电极阻抗变化未对相位间间隔偏移效应表现出直接影响,但该影响可能在人工耳蜗使用者间存在个体化差异。
  • 作者认为,鉴于相位间间隔偏移效应与其他耳蜗及神经健康指标关系复杂,临床解读和应用该指标应谨慎。

研究设计与样本

  • 前瞻性研究,纳入17名在苏黎世大学医院接受人工耳蜗手术、500 Hz存在残余声听的成人。
  • 于术中及术后4至12周在蜗顶、蜗中、蜗底电极组记录电诱发复合动作电位阈值、相位间间隔偏移效应、500 Hz短音耳蜗电图反应和单极电阻抗。
  • 术前3周内及术后约6周进行纯音测听,采用线性混合回归分析和t检验评估各指标间关系。

主要发现

  • 相位间间隔偏移效应与术中电诱发复合动作电位阈值正相关(p < 0.001)。
  • 相位间间隔偏移效应与术前纯音平均听阈无显著相关(p = 0.999)。
  • 耳蜗电图振幅下降的电极组术后相位间间隔偏移效应减小,且减小幅度显著大于耳蜗电图振幅未下降的电极组,t(17)=2.76,p=0.014。
  • 线性混合回归分析显示电极阻抗变化对相位间间隔偏移效应无系统性影响(p = 0.263),但提示电极阻抗对相位间间隔偏移效应存在参与者依赖性影响。

临床怎么用

  • 作者认为,鉴于相位间间隔偏移效应与其他耳蜗及神经健康指标关系复杂,临床解读和应用该指标应谨慎。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: The inter-phase gap (IPG) offset effect is defined as the dB offset between the linear parts of electrically evoked compound action potential (ECAP) amplitude growth functions for two stimuli differing only in IPG. The method was recently suggested to represent neural health in cochlear implant (CI) users while being unaffected by CI electrode impedances. Hereby, a larger IPG offset effect should reflect better neural health. The aims of the present study were to (1) examine whether the IPG offset effect negatively correlates with the ECAP threshold and the preoperative pure-tone average (PTA) in CI recipients with residual acoustic hearing and (2) investigate the dependency of the IPG offset effect on hair cell survival and intracochlear electrode impedances. DESIGN: Seventeen adult study participants with residual acoustic hearing at 500 Hz undergoing CI surgery at the University Hospital of Zurich were prospectively enrolled. ECAP thresholds, IPG offset effects, electrocochleography (ECochG) responses to 500 Hz tone bursts, and monopolar electrical impedances were obtained at an apical, middle, and basal electrode set during and between 4 and 12 weeks after CI surgery. Pure-tone audiometry was conducted within 3 weeks before surgery and approximately 6 weeks after surgery. Linear mixed regression analyses and t tests were performed to assess relationships between (changes in) ECAP threshold, IPG offset, impedance, PTA, and ECochG amplitude. RESULTS: The IPG offset effect positively correlated with the ECAP threshold in intraoperative recordings ( p < 0.001) and did not significantly correlate with the preoperative PTA ( p = 0.999). The IPG offset showed a postoperative decrease for electrode sets that showed an ECochG amplitude drop. This IPG offset decrease was significantly larger than for electrode sets that showed no ECochG amplitude decrease, t (17) = 2.76, p = 0.014. Linear mixed regression analysis showed no systematic effect of electrode impedance changes on the IPG offset effect ( p = 0.263) but suggested a participant-dependent effect of electrode impedance on IPG offset. CONCLUSIONS: The present study results did not reveal the expected relationships between the IPG offset effect and ECAP threshold values or between the IPG offset effect and preoperative acoustic hearing. Changes in electrode impedance did not exhibit a direct impact on the IPG offset effect, although this impact might be individualized among CI recipients. Overall, our findings suggest that the interpretation and application of the IPG offset effect in clinical settings should be approached with caution considering its complex relationships with other cochlear and neural health metrics.