内镜距离对量化年龄相关声带萎缩指标的影响
本研究纳入23例年龄相关声带萎缩(ARVA)患者,在柔性视频喉频闪喉镜下比较近距与远距拍摄对声带弓形指数(BI)和标准化声门间隙面积(NGGA)的影响。近距BI均值较远距高1.94(95%CI:0.92-2.96,p=0.001);NGGA变化为-0.24(95%CI:-0.48至0.01,p<0.05)。结论认为BI和NGGA受内镜距离影响,BI在近距条件下显著更高,声带弓形程度越大差异越明显。
要点
- 近距BI均值较远距高1.94(95%CI:0.92-2.96,p=0.001)。
- NGGA随距离变化为-0.24(95%CI:-0.48至0.01,p<0.05)。
- BI高于中位数的患者近距BI显著高于远距(p<0.05)。
核心结论
- 柔性视频喉频闪喉镜下,BI和NGGA受内镜距离影响,近距BI显著高于远距。
- 声带弓形程度越大,近距与远距的BI差异越明显。
研究设计与样本
- 纳入接受柔性视频喉频闪喉镜检查的ARVA患者23例,平均年龄77±7岁。
- 图像分为近距(靠近会厌茎)和远距(鼻咽下方,可见舌根和整个会厌)两种条件。
- BI用移动应用程序计算,NGGA用ImageJ测量。
主要发现
- 近距BI均值较远距高1.94(95%CI:0.92-2.96,p=0.001)。
- NGGA随距离变化为-0.24(95%CI:-0.48至0.01,p<0.05)。
- 按BI中位数分组,高于中位数者近距BI较远距显著增高(p<0.05),低于中位数者近距与远距无差异。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
INTRODUCTION: The bowing index (BI) and normalized glottal gap area (NGGA) are used to quantify vocal fold morphology in ARVA; however, the influence of the distance between the flexible laryngoscope lens and the target area is not known. The goal is to test whether the endoscopic distance impacts vocal fold morphology measurements in patients with ARVA during flexible video laryngostroboscopy (VLS). METHOD: Patients with ARVA who underwent VLS were included. Images were classified into near (close to the petiole of the epiglottis) and far (below nasopharynx, with tongue base and entire epiglottis visible) conditions. BI was calculated using a mobile application, and NGGA was measured using ImageJ. RESULTS: This study included 23 patients; the mean age was 77 ± 7 years. Mean BI measured at the near distance was higher than far distances with a mean difference of 1.94 (95% CI: 0.92-2.96, p = 0.001). NGGA showed difference with changed distance -0.24 (95% CI: -0.48 to 0.01, p < 0.05).When stratifying patients into two groups based on median BI measurement, there was a statistically significant difference between near and far conditions, with increased BI in the near condition for patients above the median (p < 0.05), but no difference between the near and far condition for patients with BI below the median. CONCLUSION: The BI and NGGA were impacted by the endoscopic distance during flexible VLS. BI was significantly higher in the near condition compared with the far condition. The difference in BI between the near and far conditions was more pronounced when the vocal fold bowing was greater. These findings call for heightened awareness of measurement discrepancies secondary to the endoscopic distance during flexible laryngostroboscopy. LEVEL OF EVIDENCE: 2 Laryngoscope, 134:4649-4655, 2024.
引用
- Aboueisha MA, Sauder C, Jaleel Z, et al. Endoscopic Distance and its Impact on Quantified Age-related Vocal Fold Atrophy Measures. Laryngoscope. 2024. PMID: 38877827. doi:10.1002/lary.31579
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。