利用耳蜗内耳蜗电图预测人工耳蜗植入者术后言语感知与听阈

Predicting Postoperative Speech Perception and Audiometric Thresholds Using Intracochlear Electrocochleography in Cochlear Implant Recipients.
PubMed 文献 · Ear Hear · 2024-05-31

研究纳入113例人工耳蜗植入者,在植入后立即于22电极阵列中的11个耳蜗内电极记录500 Hz短音诱发的耳蜗电图,提取耳蜗微音电位(CM)、总和电位(SP)、复合动作电位(CAP)和听神经神经音(ANN)的最大振幅及电极位置,并逐步加入多因素广义相加模型,预测术后3个月和12个月的纯音听阈(PTA)及言语感知评分(SRT和CVC-P)。结果显示,耳蜗电图因子模型的预测效果优于仅用临床因素的模型,其中SP最大偏转振幅及其在阵列中的位置具有显著预测价值,可能反映耳蜗健康状况。

要点

  • 植入后耳蜗内耳蜗电图记录可解释中等程度的术后言语感知和听阈变异。
  • 耳蜗电图因子模型对术后PTA、CVC-P和SRT的预测均优于临床因素模型。
  • SP最大偏转振幅及CM、SP峰值电极位置是预测术后结果的关键耳蜗电图因素。

核心结论

  • 植入后全阵列耳蜗内耳蜗电图记录可解释中等程度的术后言语感知和听阈变异。
  • SP最大偏转及其在阵列中的位置似乎具有显著预测价值,可能反映耳蜗健康状况。

研究设计与样本

  • 研究纳入113名人类受试者,在人工耳蜗植入后立即于22电极阵列中的11个耳蜗内电极记录500 Hz短音诱发的反应。
  • 对凝音和疏音刺激的反应进行相减以突出CM,相加以突出SP、ANN和CAP,并记录各成分的最大振幅和耳蜗外电极位置。
  • 将这些耳蜗电图因素逐步加入多因素广义相加模型,构建预测术后3个月和12个月PTA、SRT和CVC-P的最佳拟合模型,并与仅使用临床因素(术前评分、年龄和性别)的广义相加模型(作为零模型代理)进行比较。

主要发现

  • 耳蜗电图因子模型在预测术后3个月和12个月的PTA、CVC-P和SRT结局方面均优于临床因素模型。
  • 临床因素模型解释了中等程度的PTA方差(3个月时r²=45.9%,12个月时r²=31.8%,均p<0.001),对CVC-P和SRT的解释方差较小(r²范围=6%至13.7%,p=0.008至0.113)。
  • 年龄不是显著的预测因素。
  • 耳蜗电图模型在12个月时间点解释的方差(PTA的r²=52.9%,CVC-P的r²=39.6%,SRT的r²=36.4%)大于3个月时间点(PTA的r²=49.4%,CVC-P的r²=26.5%,SRT的r²=22.3%)。
  • 耳蜗电图模型基于三个因素:SP最大偏转振幅以及CM和SP峰值的电极位置;加入神经(ANN和/或CAP)因素未改善方差解释。

关键数值与推荐

  • 较大的负向SP偏转与较差结局相关,较大的正向SP偏转与较好的术后结局相关。
  • SP和CM的阵列中部峰值均与较差结局相关。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: Electrocochleography (ECochG) appears to offer the most accurate prediction of post-cochlear implant hearing outcomes. This may be related to its capacity to interrogate the health of underlying cochlear tissue. The four major components of ECochG (cochlear microphonic [CM], summating potential [SP], compound action potential [CAP], and auditory nerve neurophonic [ANN]) are generated by different cochlear tissue components. Analyzing characteristics of these components can reveal the state of hair and neural cell in a cochlea. There is limited evidence on the characteristics of intracochlear (IC) ECochG recordings measured across the array postinsertion but compared with extracochlear recordings has better signal to noise ratio and spatial specificity. The present study aimed to examine the relationship between ECochG components recorded from an IC approach and postoperative speech perception or audiometric thresholds. DESIGN: In 113 human subjects, responses to 500 Hz tone bursts were recorded at 11 IC electrodes across a 22-electrode cochlear implant array immediately following insertion. Responses to condensation and rarefaction stimuli were then subtracted from one another to emphasize the CM and added to one another to emphasize the SP, ANN, and CAP. Maximum amplitudes and extracochlear electrode locations were recorded for each of these ECochG components. These were added stepwise to a multi-factor generalized additive model to develop a best-fit model predictive model for pure-tone audiometric thresholds (PTA) and speech perception scores (speech recognition threshold [SRT] and consonant-vowel-consonant phoneme [CVC-P]) at 3- and 12-month postoperative timepoints. This best-fit model was tested against a generalized additive model using clinical factors alone (preoperative score, age, and gender) as a null model proxy. RESULTS: ECochG-factor models were superior to clinical factor models in predicting postoperative PTA, CVC-P, and SRT outcomes at both timepoints. Clinical factor models explained a moderate amount of PTA variance ( r2 = 45.9% at 3-month, 31.8% at 12-month, both p < 0.001) and smaller variances of CVC-P and SRT ( r2 range = 6 to 13.7%, p = 0.008 to 0.113). Age was not a significant predictive factor. ECochG models explained more variance at the 12-month timepoint ( r2 for PTA = 52.9%, CVC-P = 39.6%, SRT = 36.4%) compared with the 3-month one timepoint ( r2 for PTA = 49.4%, CVC-P = 26.5%, SRT = 22.3%). The ECochG model was based on three factors: maximum SP deflection amplitude, and electrode position of CM and SP peaks. Adding neural (ANN and/or CAP) factors to the model did not improve variance explanation. Large negative SP deflection was associated with poorer outcomes and a large positive SP deflection with better postoperative outcomes. Mid-array peaks of SP and CM were both associated with poorer outcomes. CONCLUSIONS: Postinsertion IC-ECochG recordings across the array can explain a moderate amount of postoperative speech perception and audiometric thresholds. Maximum SP deflection and its location across the array appear to have a significant predictive value which may reflect the underlying state of cochlear health.

来源:Ear Hear,查看原文。版权归原作者所有。

引用

  1. Panario J, Bester C, O'Leary S, et al. Predicting Postoperative Speech Perception and Audiometric Thresholds Using Intracochlear Electrocochleography in Cochlear Implant Recipients. Ear Hear. 2024. PMID: 38816899. doi:10.1097/AUD.0000000000001506