文献速递耳科与耳神经

双侧人工耳蜗同步自动增益控制可减轻独立双侧压缩引起的动态定位缺陷

Synchronizing Automatic Gain Control in Bilateral Cochlear Implants Mitigates Dynamic Localization Deficits Introduced by Independent Bilateral Compression.
PubMed 文献 · Ear Hear · 2024-03-13

研究纳入10例双侧人工耳蜗患者,比较同步与独立自动增益控制(AGC)下,70 dBA(高于AGC阈值)和50 dBA(低于AGC阈值)时有无头部运动的声源定位表现。结果显示独立AGC在70 dBA时前后翻转率高于50 dBA基线,同步AGC结合头部运动可将前后翻转率降至接近50 dBA基线水平。

要点

  • 独立AGC在70 dBA时前后翻转率高于50 dBA基线,提示其损害前后定位。
  • 静止时同步AGC未显著降低前后翻转率。
  • 同步AGC结合头部运动可将前后翻转率恢复至接近50 dBA基线水平。

核心结论

  • 允许头部运动时,同步AGC可使听者减轻不同步AGC引起的前后混淆,使个体表现恢复至约50 dBA基线水平。
  • 同步AGC未能克服AGC未启用时观察到的定位缺陷,这些缺陷与AGC压缩无关。

研究设计与样本

  • 研究为在10例双侧人工耳蜗患者中进行的声源定位测试,比较同步与独立AGC条件。
  • 宽带噪声刺激来自环绕听者的6个等间距扬声器,测试70 dBA(高于AGC阈值)和50 dBA(低于AGC阈值)两种水平。
  • 每种水平下测试两种条件:患者保持静止,以及鼓励在±30°范围内自由头部运动。

主要发现

  • 70 dBA独立AGC时,无论有无头部运动,平均前后翻转率均高于50 dBA,提示独立AGC启用时导致更差的前后定位。
  • 听者静止时,同步AGC未显著降低前后翻转率。
  • 70 dBA独立AGC时,头部运动对前后翻转率无显著影响。
  • 50 dBA AGC未启用时和70 dBA同步AGC时,头部运动对前后翻转率有显著组效应。
  • 同步AGC对听者总体定位正确率有显著组效应。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: The independence of left and right automatic gain controls (AGCs) used in cochlear implants can distort interaural level differences and thereby compromise dynamic sound source localization. We assessed the degree to which synchronizing left and right AGCs mitigates those difficulties as indicated by listeners' ability to use the changes in interaural level differences that come with head movements to avoid front-back reversals (FBRs). DESIGN: Broadband noise stimuli were presented from one of six equally spaced loudspeakers surrounding the listener. Sound source identification was tested for stimuli presented at 70 dBA (above AGC threshold) for 10 bilateral cochlear implant patients, under conditions where (1) patients remained stationary and (2) free head movements within ±30° were encouraged. These conditions were repeated for both synchronized and independent AGCs. The same conditions were run at 50 dBA, below the AGC threshold, to assess listeners' baseline performance when AGCs were not engaged. In this way, the expected high variability in listener performance could be separated from effects of independent AGCs to reveal the degree to which synchronizing AGCs could restore localization performance to what it was without AGC compression. RESULTS: The mean rate of FBRs was higher for sound stimuli presented at 70 dBA with independent AGCs, both with and without head movements, than at 50 dBA, suggesting that when AGCs were independently engaged they contributed to poorer front-back localization. When listeners remained stationary, synchronizing AGCs did not significantly reduce the rate of FBRs. When AGCs were independent at 70 dBA, head movements did not have a significant effect on the rate of FBRs. Head movements did have a significant group effect on the rate of FBRs at 50 dBA when AGCs were not engaged and at 70 dBA when AGCs were synchronized. Synchronization of AGCs, together with head movements, reduced the rate of FBRs to approximately what it was in the 50-dBA baseline condition. Synchronizing AGCs also had a significant group effect on listeners' overall percent correct localization. CONCLUSIONS: Synchronizing AGCs allowed for listeners to mitigate front-back confusions introduced by unsynchronized AGCs when head motion was permitted, returning individual listener performance to roughly what it was in the 50-dBA baseline condition when AGCs were not engaged. Synchronization of AGCs did not overcome localization deficiencies which were observed when AGCs were not engaged, and which are therefore unrelated to AGC compression.