文献速递耳科与耳神经

智能手机辅助的原位助听器测听用于社区听力检测

Smartphone-Facilitated In-Situ Hearing Aid Audiometry for Community-Based Hearing Testing.
PubMed 文献 · Ear Hear · 2024-03-01

该研究通过四项实验评估智能手机辅助原位测听在社区场景下的有效性:实验1确定覆盖助听器的耳罩最大允许环境噪声级为24至47.3 dB SPL;实验2显示14台同型号助听器设备间差异均小于3 dB;实验3和4中,85.2%和83.3%的自动原位测听阈值与临床纯音测听阈值的差异在10 dB以内,重测信度组内相关系数分别为0.85至0.93和0.83至0.97。结论认为该方法可使社区健康工作者在最少培训下进行可靠的社区听力检测。

要点

  • 覆盖助听器的Peltor 3M耳罩在各频率的最大允许环境噪声级均高于标准耳机,范围为24至47.3 dB SPL。
  • 14台同型号助听器设备间所有强度和频率的输出差异均小于3 dB,频率输出差异小于0.7%。
  • 社区场景下83.3%的自动原位测听阈值与移动纯音测听阈值差异在10 dB以内,重测组内相关系数为0.83至0.97。

核心结论

  • 智能手机辅助的原位测听可实现可靠且有效的社区听力检测。
  • 简单的智能手机用户界面和自动原位测听使经过最少培训的社区健康工作者能够实施测试。
  • 该方法在初始测试后可通过智能手机编程助听器,有潜力支持中低收入国家由社区健康工作者推动的个性化助听器验配广泛可及。
  • 该方法也支持基于原位阈值的自我验配选项,使通过非处方助听器进行测试和验配成为可能。

研究设计与样本

  • 实验1:15名听力正常成人参与者,确定覆盖Lexie Lumen助听器的Peltor 3M耳罩在智能手机辅助原位测听下的最大允许环境噪声级。
  • 实验2:测量14台同型号Lexie Lumen助听器的纯音频率和强度输出,确定设备间可靠性。
  • 实验3:85名成人参与者在隔声室内比较自动原位纯音测听与听力师实施的临床纯音测听(500、1000、2000、3000、4000、6000 Hz)的准确性,其中前39名参与者测试两次以确定重测信度。
  • 实验4:社区场景下144名成人参与者由社区健康工作者使用智能手机应用实施自动原位测听,随后接受自动移动纯音测听(500、1000、2000、4000 Hz);另44名参与者测试两次以确定重测信度。

主要发现

  • 实验1:覆盖Peltor 3M耳罩的助听器最大允许环境噪声级在所有频率均高于标准耳机,范围为24至47.3 dB SPL。
  • 实验2:设备间性能可靠性高,所有强度和频率的设备间差异均小于3 dB,频率输出差异小于0.7%。
  • 实验3和4:分别有85.2%和83.3%的自动原位测听阈值与隔声室和社区场景获得的阈值差异在10 dB以内。
  • 隔声室获得的所有阈值重测组内相关系数为0.85至0.93,社区场景为0.83至0.97。

临床怎么用

  • 原文结论认为智能手机辅助原位测听可实现可靠且有效的社区听力检测。
  • 原文结论认为简单的智能手机用户界面和自动原位测听使经过最少培训的社区健康工作者能够实施测试。
  • 原文结论认为该方法有潜力支持中低收入国家由社区健康工作者推动的个性化助听器验配广泛可及。
  • 原文结论认为该方法也支持基于原位阈值的自我验配选项,使通过非处方助听器进行测试和验配成为可能。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: Hearing loss prevalence is increasing, with an estimated 2.5 billion people affected globally by 2050. Scalable service delivery models using innovative technologies and task-shifting are World Health Organization priorities to improve access to hearing care, particularly in low- and middle-income countries. Smartphone-facilitated audiometry in the community using hearing aids covered by noise-attenuating ear cups ("in-situ") could support more accessible hearing care when provided by less trained individuals such as community health workers (CHWs). This study aimed to determine the validity of this method for potential hearing aid fitting. Study objectives included determining the maximum permissible ambient noise level (MPANL), inter-device reliability, clinical threshold accuracy, reliability, and performance in real-world settings. DESIGN: Experiment 1: 15 normal-hearing adult participants were evaluated to determine MPANLs for circumaural Peltor 3M earcups covering Lexie Lumen hearing aids with smartphone-facilitated in-situ audiometry. MPANLs were calculated by measuring the difference in attenuation between thresholds obtained with standard headphones and in-situ hearing aids. Experiment 2: Pure-tone frequency and intensity output of 14 same-model Lexie Lumen hearing aids were measured to determine inter-device reliability. Pure-tone stimuli were measured and analyzed to determine sound pressure levels in decibels and pure-tone frequency when connected to a test box 2cc coupler. Experiment 3: 85 adult participants were tested in a sound booth to determine the accuracy of automated in-situ pure-tone audiometry (PTA) compared to clinical PTA (500, 1000, 2000, 3000, 4000, 6000 Hz) facilitated by an audiologist. The first 39 participants were tested twice to determine test-retest reliability. Experiment 4: In a community setting, 144 adult participants were tested with automated in-situ audiometry facilitated by CHWs using a smartphone app. These participants were subsequently tested with automated mobile PTA (500, 1000, 2000, 4000 Hz). An additional 44 participants were tested twice to determine test-retest reliability. RESULTS: Experiment 1: MPANLs of the Peltor 3M earcup-covered hearing aids were higher than standard headphones across all frequencies, ranging from 24 to 47.3 dB SPL. Experiment 2: Inter-device performance reliability was high, with all inter-device differences across all intensities and frequencies less than 3 dB. Frequency output was consistent and differed less than 0.7% between devices. Experiments 3 and 4: 85.2% and 83.3% of automated in-situ audiometry thresholds were within 10 dB of thresholds obtained in the sound booth and in a community setting, respectively. Acceptable test-retest intraclass correlation coefficient (ICC) was evident across all thresholds obtained in a sound booth (ICC = 0.85 to 0.93) and in a community setting (ICC = 0.83 to 0.97). CONCLUSIONS: Smartphone-facilitated in-situ audiometry allows for reliable and valid community-based testing. A simple smartphone user interface and automated in-situ audiometry allow CHWs with minimal training to facilitate the testing. With the additional capability to program hearing aids via the smartphone after the initial test, this approach would have the potential to support widespread access to personalized hearing aid fittings facilitated by CHWs in low- and middle-income countries. This approach also supports self-fitting options based on in-situ thresholds, enabling testing and fitting via over the counter hearing aids.

来源:Ear Hear,查看原文。版权归原作者所有。