本研究基于30名受试者CT构建的三维模型生成虚拟内镜与深度图视频,由两名耳鼻咽喉科医生使用软件工具估计鼻咽横截面积和鼻瓣膜最小横截面积。鼻咽横截面积中位百分比误差为3.7%至4.6%,鼻瓣膜最小横截面积中位百分比误差为22.7%至33.6%,识别鼻瓣膜狭窄的敏感度为83.3%、特异度≥90.7%、分类准确率≥90.0%。

要点

  • 鼻咽横截面积估计的中位百分比误差为3.7%至4.6%。
  • 鼻瓣膜最小横截面积估计的中位百分比误差为22.7%至33.6%。
  • 识别鼻瓣膜狭窄的敏感度为83.3%,特异度≥90.7%,准确率≥90.0%。

核心结论

  • 三维模型中的气道横截面积可从深度图准确估计。
  • 开发人工智能算法计算深度图可能很快允许从临床内镜量化气腔横截面积。

研究设计与样本

  • 横断面研究,学术三级医疗中心。
  • 基于30名受试者CT扫描构建的解剖学精确三维模型,数字化创建鼻腔虚拟内镜和深度图视频。
  • 开发软件工具从深度图勾勒气道周界并估计气腔横截面积。
  • 两名耳鼻咽喉科医生使用该软件工具估计左右鼻腔的鼻咽横截面积和鼻瓣膜最小横截面积。
  • 进行了模型验证统计。

主要发现

  • 鼻咽横截面积与三维模型真实值相比,中位百分比误差为3.7%至4.6%。
  • 鼻瓣膜最小横截面积相对于真实值的中位百分比误差为22.7%至33.6%。
  • 评分者成功使用软件工具识别鼻瓣膜狭窄(即最小横截面积<0.20 cm2),敏感度为83.3%,特异度≥90.7%,分类准确率≥90.0%。
  • 评分者间和评分者内一致性高。

关键数值与推荐

  • 鼻瓣膜狭窄定义为最小横截面积<0.20 cm2。
  • 鼻咽横截面积中位百分比误差3.7%至4.6%。
  • 鼻瓣膜最小横截面积中位百分比误差22.7%至33.6%。
  • 识别鼻瓣膜狭窄敏感度83.3%,特异度≥90.7%,分类准确率≥90.0%。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVE: Endoscopy is routinely used to diagnose obstructive airway diseases. Currently, endoscopy is only a visualization technique and does not allow quantification of airspace cross-sectional areas (CSAs). This pilot study tested the hypothesis that CSAs can be accurately estimated from depth maps created from virtual endoscopy videos. STUDY DESIGN: Cross-sectional. SETTING: Academic tertiary medical center. METHODS: Virtual endoscopy and depth map videos of the nasal cavity were digitally created based on anatomically accurate three-dimensional (3D) models built from computed tomography scans of 30 subjects. A software tool was developed to outline the airway perimeter and estimate the airspace CSA from the depth maps. Two otolaryngologists used the software tool to estimate the nasopharynx CSA and the nasal valve minimal CSA (mCSA) in the left and right nasal cavities. Model validation statistics were performed. RESULTS: Nasopharynx CSA had a median percent error of 3.7% to 4.6% when compared to the true values measured in the 3D models. Nasal valve mCSA had a median percent error of 22.7% to 33.6% relative to the true values. Raters successfully used the software tool to identify subjects with nasal valve stenosis (ie, mCSA < 0.20 cm2) with a sensitivity of 83.3%, specificity ≥ 90.7%, and classification accuracy ≥ 90.0%. Interrater and intrarater agreements were high. CONCLUSION: This study demonstrates that airway CSAs in 3D models can be accurately estimated from depth maps. The development of artificial intelligence algorithms to compute depth maps may soon allow the quantification of airspace CSAs from clinical endoscopies.

来源:Otolaryngol Head Neck Surg,查看原文。版权归原作者所有。

引用

  1. Garcia GJM, Catalano D, Shum A, et al. Estimation of Nasal Airway Cross-sectional Area From Endoscopy Using Depth Maps: A Proof-of-Concept Study. Otolaryngol Head Neck Surg. 2024. PMID: 38329226. doi:10.1002/ohn.669

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。