豚鼠模型提示:耳蜗植入后人类声学听力保留的客观评估受声诱发电位空间起源沿耳蜗长度偏移的混杂影响

A Guinea Pig Model Suggests That Objective Assessment of Acoustic Hearing Preservation in Human Ears With Cochlear Implants Is Confounded by Shifts in the Spatial Origin of Acoustically Evoked Potential Measurements Along the Cochlear Length.
PubMed 文献 · Ear Hear · 2024-01-05

该研究在豚鼠圆窗附近记录声诱发听神经复合动作电位,植入耳蜗植入体但不行电刺激或记录,并用红藻氨酸从耳蜗顶向底灌注序贯削减神经反应以揭示电位空间起源。结果显示植入使3至11 kHz阈值升高,8 kHz处最大升高33.66 dB,11 kHz成为植入耳最敏感区,而500 Hz阈值及阈上功能不受影响。随声压级升高,植入耳CAP空间起源向最敏感区而非扩展高频区偏移,个体耳间起源部位各异。

要点

  • 豚鼠耳蜗植入体在原位可使3至11 kHz声诱发阈值升高,8 kHz处最大升高33.66 dB。
  • 植入后11 kHz成为豚鼠耳对声刺激最敏感的区域。
  • 植入不影响500 Hz低频阈值及听神经重叠波形所示的阈上功能。

核心结论

  • 原位耳蜗植入体所致的耳蜗底功能改变可能影响植入耳常用高强度低频声刺激进行的客观测量。
  • 豚鼠结果有助于解读用于理解人耳植入后残余声学听力如何及何时丧失的测量。

研究设计与样本

  • 豚鼠模型研究,声诱发听神经复合动作电位(CAP)经置于圆窗附近的电极记录。
  • 耳蜗植入体未用于电刺激或记录,仅原位放置。
  • 采用从耳蜗顶向底耳蜗导水管缓慢灌注红藻氨酸溶液序贯削减精细间隔耳蜗频率区的神经反应。

主要发现

  • 3至11 kHz阈值测量因植入而升高。
  • 同一耳内,插入耳蜗植入体时阈值升高、移除时阈值降低,提示植入体所致「传导性听力损失」。
  • 最大阈值升高出现在 naive 豚鼠耳最敏感区(8 kHz处33.66 dB),使11 kHz成为植入耳对声刺激最敏感区。
  • 急性植入不影响500 Hz低频阈值及听神经重叠波形所示的阈上功能。
  • 随短纯音声压级升高,平均数据显示CAP沿耳蜗长度的空间起源向植入耳最敏感耳蜗区偏移,而非扩展高频耳蜗区。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: Our recent empirical findings have shown that the auditory nerve compound action potential (CAP) evoked by a low-level tone burst originates from a narrow cochlear region tuned to the tone burst frequency. At moderate to high sound levels, the origins shift to the most sensitive audiometric regions rather than the extended high-frequency regions of the cochlear base. This means that measurements evoked from extended high-frequency sound stimuli can shift toward the apex with increasing level. Here we translate this study to understand the spatial origin of acoustically evoked responses from ears that receive cochlear implants, an emerging area of research and clinical practice that is not completely understood. An essential step is to first understand the influence of the cochlear implant in otherwise naive ears. Our objective was to understand how function of the high-frequency cochlear base, which can be excited by the intense low-frequency sounds that are frequently used for objective intra- and postoperative monitoring, can be influenced by the presence of the cochlear implant. DESIGN: We acoustically evoked responses and made measurements with an electrode placed near the guinea pig round window. The cochlear implant was not utilized for either electrical stimulation or recording purposes. With the cochlear implant in situ, CAPs were acoustically evoked from 2 to 16 kHz tone bursts of various levels while utilizing the slow perfusion of a kainic acid solution from the cochlear apex to the cochlear aqueduct in the base, which sequentially reduced neural responses from finely spaced cochlear frequency regions. This cochlear perfusion technique reveals the spatial origin of evoked potential measurements and provides insight on what influence the presence of an implant has on acoustical hearing. RESULTS: Threshold measurements at 3 to 11 kHz were elevated by implantation. In an individual ear, thresholds were elevated and lowered as cochlear implant was respectively inserted and removed, indicative of "conductive hearing loss" induced by the implant. The maximum threshold elevation occurred at most sensitive region of the naive guinea pig ear (33.66 dB at 8 kHz), making 11 kHz the most sensitive region to acoustic sounds for guinea pig ears with cochlear implants. Conversely, the acute implantation did not affect the low-frequency, 500 Hz thresholds and suprathreshold function, as shown by the auditory nerve overlapped waveform. As the sound pressure level of the tone bursts increased, mean data show that the spatial origin of CAPs along the cochlear length shifted toward the most sensitive cochlear region of implanted ears, not the extended high-frequency cochlear regions. However, data from individual ears showed that after implantation, measurements from moderate to high sound pressure levels originate in places that are unique to each ear. CONCLUSIONS: Alterations to function of the cochlear base from the in situ cochlear implant may influence objective measurements of implanted ears that are frequently made with intense low-frequency sound stimuli. Our results from guinea pigs advance the interpretation of measurements used to understand how and when residual acoustic hearing is lost in human ears receiving a cochlear implant.

来源:Ear Hear,查看原文。版权归原作者所有。

引用

  1. Lee C, Hartsock JJ, Salt AN, et al. A Guinea Pig Model Suggests That Objective Assessment of Acoustic Hearing Preservation in Human Ears With Cochlear Implants Is Confounded by Shifts in the Spatial Origin of Acoustically Evoked Potential Measurements Along the Cochlear Length. Ear Hear. 2024. PMID: 38178312. doi:10.1097/AUD.0000000000001457

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。