新人工耳蜗植入者音调定位与默认频率拟合对噪声下言语理解的比较:一项前瞻性、随机、双盲、交叉研究
该研究为前瞻性随机双盲交叉研究,纳入26例新人工耳蜗植入者,比较音调定位图与常规图对噪声和安静下言语识别的影响。结果显示音调定位拟合在所有测试信噪比下噪声下言语识别均显著更优,安静下言语识别和纯音测听无显著差异,92%参与者研究结束后保留音调定位图。
要点
- 音调定位拟合在信噪比+9、+6、+3、0 dB下噪声下言语识别均显著优于常规拟合。
- 安静下言语识别(p=0.66)和纯音测听(p=0.203)在两种拟合间无显著差异。
- 92%的参与者在研究期结束后保留了基于音调定位的图。
核心结论
- 对于新人工耳蜗植入者,音调定位拟合似乎比默认频率拟合更有效,因为它在不损害安静下言语理解的情况下改善了噪声下言语识别。
研究设计与样本
- 前瞻性、随机、双盲、两阶段交叉研究,在法国雷恩大学医院中心进行,为期6个月。
- 纳入26例新人工耳蜗植入者,年龄大于18岁,双侧重度至极重度感音神经性听力损失或完全听力损失病程小于5年。
- 使用术后平板计算机断层扫描和基于Greenwood函数的重建软件创建解剖音调定位图。
- 参与者随机分配至先接受常规图后接受音调定位图或相反顺序,每种设置维持6周,期末进行言语感知任务。
- 主要结局指标为噪声下言语识别;采用区组随机化(区组大小2,比例1:1)分配序列,参与者和结局评估者设盲。
主要发现
- 26例参与者中每序列随机分配13例,因COVID-19大流行排除2例(每序列1例),最终分析24例。
- 噪声下言语识别在音调定位拟合下所有测试信噪比水平均显著更优:信噪比+9 dB,p=0.002,平均效应=12.1%,95%CI=4.9至19.2,标准化效应量=0.71;信噪比+6 dB,p<0.001,平均效应=16.3%,95%CI=9.8至22.7,标准化效应量=1.07;信噪比+3 dB,p<0.001,平均效应=13.8%,95%CI=6.9至20.6,标准化效应量=0.84;信噪比0 dB,p=0.003,平均效应=10.8%,95%CI=4.1至17.6,标准化效应量=0.68。
- 任何信号水平均未观察到阶段效应或交互效应。
- 安静下言语识别(p=0.66)和纯音测听(p=0.203)在两种设置间无显著差异。
- 92%的参与者在研究期结束后保留了基于音调定位的图。
关键数值与推荐
- 噪声下言语识别在信噪比+9 dB时,音调定位拟合平均效应=12.1%,95%CI=4.9至19.2,p=0.002。
- 噪声下言语识别在信噪比+6 dB时,音调定位拟合平均效应=16.3%,95%CI=9.8至22.7,p<0.001。
- 噪声下言语识别在信噪比+3 dB时,音调定位拟合平均效应=13.8%,95%CI=6.9至20.6,p<0.001。
- 噪声下言语识别在信噪比0 dB时,音调定位拟合平均效应=10.8%,95%CI=4.1至17.6,p=0.003。
- 安静下言语识别在两种拟合间无显著差异(p=0.66),纯音测听无显著差异(p=0.203)。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: While cochlear implants (CIs) have provided benefits for speech recognition in quiet for subjects with severe-to-profound hearing loss, speech recognition in noise remains challenging. A body of evidence suggests that reducing frequency-to-place mismatch may positively affect speech perception. Thus, a fitting method based on a tonotopic map may improve speech perception results in quiet and noise. The aim of our study was to assess the impact of a tonotopic map on speech perception in noise and quiet in new CI users. DESIGN: A prospective, randomized, double-blind, two-period cross-over study in 26 new CI users was performed over a 6-month period. New CI users older than 18 years with bilateral severe-to-profound sensorineural hearing loss or complete hearing loss for less than 5 years were selected in the University Hospital Centre of Rennes in France. An anatomical tonotopic map was created using postoperative flat-panel computed tomography and a reconstruction software based on the Greenwood function. Each participant was randomized to receive a conventional map followed by a tonotopic map or vice versa. Each setting was maintained for 6 weeks, at the end of which participants performed speech perception tasks. The primary outcome measure was speech recognition in noise. Participants were allocated to sequences by block randomization of size two with a ratio 1:1 (CONSORT Guidelines). Participants and those assessing the outcomes were blinded to the intervention. RESULTS: Thirteen participants were randomized to each sequence. Two of the 26 participants recruited (one in each sequence) had to be excluded due to the COVID-19 pandemic. Twenty-four participants were analyzed. Speech recognition in noise was significantly better with the tonotopic fitting at all signal-to-noise ratio (SNR) levels tested [SNR = +9 dB, p = 0.002, mean effect (ME) = 12.1%, 95% confidence interval (95% CI) = 4.9 to 19.2, standardized effect size (SES) = 0.71; SNR = +6 dB, p < 0.001, ME = 16.3%, 95% CI = 9.8 to 22.7, SES = 1.07; SNR = +3 dB, p < 0.001 ME = 13.8%, 95% CI = 6.9 to 20.6, SES = 0.84; SNR = 0 dB, p = 0.003, ME = 10.8%, 95% CI = 4.1 to 17.6, SES = 0.68]. Neither period nor interaction effects were observed for any signal level. Speech recognition in quiet ( p = 0.66) and tonal audiometry ( p = 0.203) did not significantly differ between the two settings. 92% of the participants kept the tonotopy-based map after the study period. No correlation was found between speech-in-noise perception and age, duration of hearing deprivation, angular insertion depth, or position or width of the frequency filters allocated to the electrodes. CONCLUSION: For new CI users, tonotopic fitting appears to be more efficient than the default frequency fitting because it allows for better speech recognition in noise without compromising understanding in quiet.