上气道刺激治疗中呼吸传感导线故障:单中心报告

Respiratory Sensing Lead Malfunction in Upper Airway Stimulation: A Single Institution Report.
PubMed 文献 · Laryngoscope · 2023-09-21

该研究回顾某中心2017年7月至2022年6月植入上气道刺激(UAS)装置的190例成人患者,4例(2.1%)出现需翻修的传感导线故障;同时分析FDA MAUDE数据库中122份传感导线相关不良事件报告,72%(88/122)需翻修手术,最常见为导线尖端与导线主体分离(46%,56/122)。

要点

  • 本中心190例UAS植入者中4例(2.1%)发生传感导线故障
  • MAUDE数据库中122份传感导线不良事件报告,72%(88/122)需翻修手术
  • 最常见不良事件为导线尖端与导线主体分离,占46%(56/122)

核心结论

  • 传感导线组件的功能对UAS装置完整性至关重要,可通过波形分析评估;需全面了解该组件失效的场景、发生率及恰当处理,以在装置植入后提供最佳诊疗。

研究设计与样本

  • 单中心报告:回顾2017年7月至2022年6月在本中心植入UAS装置、年龄≥18岁的患者,分析其中确定存在传感导线故障者。
  • 同时回顾FDA制造商与用户设施器械经验(MAUDE)数据库2014年1月至2022年9月与「Inspire Model 4323 Respiratory」及「Inspire Model 4340 Respiratory」相关的报告。

主要发现

  • 研究期间本中心190例患者接受UAS,4例(2.1%)发现传感导线故障。
  • MAUDE数据库共122份与传感导线组件相关的不良事件报告,其中72%(88/122)需翻修手术。
  • 最常报告的不良事件为传感导线尖端与导线主体分离,占这些报告的46%(56/122)。

临床怎么用

  • 原文未据此给出推荐;原文结论指出传感导线功能对UAS装置完整性至关重要、可通过波形分析评估,并强调需全面了解该组件失效的场景、发生率及恰当处理。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

OBJECTIVES: Upper airway stimulation (UAS) is an effective treatment for patients with obstructive sleep apnea. The device consists of a stimulating electrode, an internal pulse generator, and a respiratory sensing lead. The purpose of this study is to characterize the incidence of sensing lead malfunction necessitating revision surgery in a high-volume center in conjunction with a review of the FDA Manufacturer and User Facility Device Experience (MAUDE) database reports on adverse events associated with the sensing lead component. METHODS: Patients age ≥18 implanted with the device were reviewed between July 2017 and June 2022. Those determined to have sensing lead malfunction were analyzed. The FDA MAUDE database was reviewed for reports associated with "Inspire Model 4323 Respiratory" and "Inspire Model 4340 Respiratory" between January 2014 and September 2022. RESULTS: One hundred ninety patients underwent UAS at our institution during the study period and four (2.1%) patients were found to have sensing lead malfunction. Analysis of the MAUDE database revealed 122 reports of adverse events associated with the sensing lead component, with 72% (88/122) of these requiring revision surgery. The most frequently reported adverse event was sensing lead tip separation from lead body, noted in 46% of these reports (56/122). CONCLUSION: The functionality of the sensing lead component is vital to the integrity of the UAS device and is measurable through waveform analysis. To provide optimal care after device implantation, it is essential to have a comprehensive understanding of the scenarios in which this component fails, the incidence of such events, and appropriate management. LEVEL OF EVIDENCE: 4 Laryngoscope, 134:1479-1484, 2024.