自动电极阵列插入装置的研制及其耳蜗内压力研究
本研究开发了一种由滚轮机构驱动的紧凑型自动电极阵列插入装置,并用3D打印直形耳蜗模型和模型电极测试不同插入速度下的耳蜗内压力变化。结果显示压力随插入深度增加而升高,压力变化率随插入速度呈指数增长;插入速度0.15 mm/s时峰值压力为133 Pa,作者认为可进一步在解剖模型中评估临床场景。
要点
- 成功研制滚轮驱动的自动电极阵列插入装置,用于人工耳蜗植入。
- 在3D打印耳蜗模型中,耳蜗内压力随电极插入深度增加而升高,与插入速度无关。
- 压力变化率随插入速度呈指数增长;0.15 mm/s时峰值压力为133 Pa。
核心结论
- 插入速度0.15 mm/s时观察到峰值压力为133 Pa,作者认为可进一步在解剖模型中评估临床场景。
研究设计与样本
- 本研究开发了一种由滚轮机构驱动电极阵列的自动插入装置。
- 采用3D打印机开发了具有鼓阶尺寸的直形耳蜗和模型电极用于测试。
- 利用压力传感器观察不同插入条件下的压力变化。
主要发现
- 电极以不同速度无暂停地插入原型耳蜗,观察到压力随电极插入深度增加而升高,与电极插入速度无关。
- 观察到压力变化率随插入速度呈指数增长。
- 插入速度为0.15 mm/s时,观察到峰值压力为133 Pa。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
UNLABELLED: Cochlear implantation is the most successful approach for people with profound sensorineural hearing loss. Manual insertion of the electrode array may result in damaging the soft tissue structures and basilar membrane. An automated electrode array insertion device is reported to be less traumatic in cochlear implant surgery. OBJECTIVES: The present work develops a simple, reliable, and compact device for automatically inserting the electrode array during cochlear implantation and test the device to observe intracochlear pressure during simulated electrode insertion. METHODS: The device actuates the electrode array by a roller mechanism. For testing the automated device, a straight cochlea having the dimension of the scala tympani and a model electrode is developed using a 3D printer. A pressure sensor is utilized to observe the pressure change at different insertional conditions. RESULTS: The electrode is inserted into a prototype cochlea at different speeds without any pause, and it is noticed that the pressure is increased with the depth of insertion of the electrode irrespective of the speed of electrode insertion. The rate of pressure change is observed to be increased exponentially with the speed of insertion. CONCLUSION: At an insertion speed of 0.15 mm/s, the peak pressure is observed to be 133 Pa, which can be further evaluated in anatomical models for clinical scenarios. LEVEL OF EVIDENCE: N/A Laryngoscope, 134:1388-1395, 2024.
引用