用于全植入式人工耳蜗的中耳新型悬浮压电麦克风
本研究通过猫体内实验和人工耳蜗术中测试,评估新型悬浮压电麦克风(NFPM)的灵敏度和有效性。结果显示,与鼓室腔放置相比,NFPM夹持于听骨链时能更灵敏地检测振动;术中测试中信号输出随声刺激强度降低而下降。结论认为NFPM术中测试有效,作为全植入式人工耳蜗的中耳植入麦克风可行。
要点
- NFPM在猫实验和术中测试中,夹持于听骨链比置于鼓室腔更灵敏地检测振动。
- 术中测试显示NFPM信号输出水平随声刺激强度降低而下降。
- NFPM术中测试有效,作为全植入式人工耳蜗的中耳植入麦克风可行。
核心结论
- NFPM在术中测试中有效,作为全植入式人工耳蜗的中耳植入麦克风可行。
研究设计与样本
- 研究采用动物实验和术中测试,在猫和人类患者中分析NFPM从0.25至10 kHz、90 dB SPL的不同频率响应。
- NFPM被夹持于听骨链或置于猫和人类患者的鼓室腔不同位置进行测试。
- 两名志愿者的长砧骨脚和四只猫的锤骨颈被夹持NFPM,记录并比较不同位置的输出电信号。
- 测试后移除NFPM,未对猫的中耳结构造成损伤;术中测试在人工耳蜗手术期间进行,所有测试后完成人工耳蜗手术。
主要发现
- 与鼓室腔结果相比,NFPM在猫实验和术中测试中能更灵敏地检测来自听骨链的振动。
- 术中测试发现,NFPM的信号输出水平随声刺激强度降低而下降。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVE: Our team designed a long-lasting, well-sealed microphone, which uses laser welding and vacuum packaging technology. This study examined the sensitivity and effectiveness of this new floating piezoelectric microphone (NFPM) designed for totally implantable cochlear implants (TICIs) in animal experiments and intraoperative testing. METHODS: Different NFPM frequency responses from 0.25 to 10 kHz at 90 dB SPL were analyzed using in vivo testing of cats and human patients. The NFPM was tested in different positions that were clamped to the ossicular chains or placed in the tympanic cavity of cats and human patients. Two volunteers' long incus foot and four cats' malleus neck of the ossicular chain were clamped with the NSFM. The output electrical signals from different locations were recorded, analyzed, and compared. The NFPM was removed after the test without causing any damage to the middle-ear structure of the cats. Intraoperative tests of the NFPM were performed during the cochlear implant surgery and the cochlear implant surgery was completed after all tests. RESULTS: Compared with the results in the tympanic cavity, the NFPM could detect the vibration from the ossicular chain more sensitively in cat experiments and intraoperative testing. We also found that the signal output level of the NFPM decreased as the acoustic stimulation strength decreased in the intraoperative testing. CONCLUSION: The NFPM is effective in the intraoperative testing, making it feasible as an implantable middle-ear microphone for TICIs. LEVEL OF EVIDENCE: 4 Laryngoscope, 134:937-944, 2024.
引用
- Xu XD, Zhang WX, Jia XH, et al. A New Floating Piezoelectric Microphone for Fully Implantable Cochlear Implants in Middle Ear. Laryngoscope. 2023. PMID: 37421255. doi:10.1002/lary.30861
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。