基于阻抗的耳蜗植入电极插入深度术后评估
研究利用56例相同侧壁电极阵列患者的术后CT作为真实插入深度,从植入日至最长60个月的阻抗遥测记录建立现象学模型估算线性与角度插入深度,结果显示全部电极插入深度估算绝对误差为0.9 mm ± 0.6 mm或22° ± 18°,术后组织电阻除最基底两个电极外保持稳定。
要点
- 基于阻抗遥测的现象学模型可估算耳蜗植入电极插入深度,绝对误差0.9 mm ± 0.6 mm或22° ± 18°。
- 术后随访期间组织电阻总体稳定,仅电极11约10 Ω/年、电极12约30 Ω/年随时间显著升高。
- 早期与晚期阻抗记录推断的现象学模型无差异,提示估算随时间可靠。
核心结论
- 基于阻抗的电极位置估算方法可应用于术后阻抗遥测记录,插入深度估算在同耳两次术后CT比较中随时间可靠。
- 未来工作需解决耳蜗外电极检测以提高方法性能。
研究设计与样本
- 从56例使用相同侧壁电极阵列的病例记录中获取术后CT扫描测量真实插入深度。
- 每例检索从植入当日到最长60个月观察期的阻抗遥测记录,基于记录用现象学模型估算线性与角度电极插入深度,并与真实值比较计算模型准确度。
- 使用线性混合效应模型分析长期记录。
主要发现
- 长期记录分析显示术后组织电阻在整个随访期保持稳定,但最基底两个电极随时间显著升高:电极11约10 Ω/年,电极12约30 Ω/年。
- 早期与晚期阻抗遥测记录推断的现象学模型无差异。
- 全部电极插入深度估算绝对误差为0.9 mm ± 0.6 mm或22° ± 18°角度(均值 ± 标准差)。
局限性
- 未来工作需解决耳蜗外电极检测以提高方法性能。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: Reliable determination of cochlear implant electrode positions shows promise for clinical applications, including anatomy-based fitting of audio processors or monitoring of electrode migration during follow-up. Currently, electrode positioning is measured using radiography. The primary objective of this study is to extend and validate an impedance-based method for estimating electrode insertion depths, which could serve as a radiation-free and cost-effective alternative to radiography. The secondary objective is to evaluate the reliability of the estimation method in the postoperative follow-up over several months. DESIGN: The ground truth insertion depths were measured from postoperative computed tomography scans obtained from the records of 56 cases with an identical lateral wall electrode array. For each of these cases, impedance telemetry records were retrieved starting from the day of implantation up to a maximum observation period of 60 mo. Based on these recordings, the linear and angular electrode insertion depths were estimated using a phenomenological model. The estimates obtained were compared with the ground truth values to calculate the accuracy of the model. RESULTS: Analysis of the long-term recordings using a linear mixed-effects model showed that postoperative tissue resistances remained stable throughout the follow-up period, except for the two most basal electrodes, which increased significantly over time (electrode 11: ~10 Ω/year, electrode 12: ~30 Ω/year). Inferred phenomenological models from early and late impedance telemetry recordings were not different. The insertion depth of all electrodes was estimated with an absolute error of 0.9 mm ± 0.6 mm or 22° ± 18° angle (mean ± SD). CONCLUSIONS: Insertion depth estimations of the model were reliable over time when comparing two postoperative computed tomography scans of the same ear. Our results confirm that the impedance-based position estimation method can be applied to postoperative impedance telemetry recordings. Future work needs to address extracochlear electrode detection to increase the performance of the method.
引用
- Schraivogel S, Aebischer P, Wagner F, et al. Postoperative Impedance-Based Estimation of Cochlear Implant Electrode Insertion Depth. Ear Hear. 2023. PMID: 37157125. doi:10.1097/AUD.0000000000001379
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。