人工耳蜗植入术中耳蜗内耳蜗电图监测耳蜗健康:一项国际临床研究的结果
这项国际多中心前瞻性观察性分组研究纳入78例成人人工耳蜗候选者(80耳),术中记录耳蜗内耳蜗电图,发现电极插入期间耳蜗微音电位幅度下降≥30%者术后4-6周低频听力损失中位数26.9 dB,显著差于保持稳定者的12.6 dB(p<0.014)。术前低频听阈≤80 dB HL是获得可测量耳蜗微音电位信号的关键条件。
要点
- 电极插入期间耳蜗微音电位幅度下降≥30%与术后4-6周低频听力损失显著相关。
- 保持稳定的耳蜗微音电位组术后听力损失中位数12.6 dB,下降组为26.9 dB。
- 术前低频听阈≤80 dB HL是术中记录到>1 µV耳蜗微音电位信号的重要条件。
核心结论
- 电极插入期间耳蜗微音电位下降与术后听力保留显著较差相关,而全程稳定的耳蜗微音电位则对应更好的术后听力。
- 术中测量耳蜗电图的可行性取决于术前低频听阈,尤其是刺激频率0.5 kHz处的听阈。
研究设计与样本
- 国际多中心前瞻性观察性分组研究,目标每组41例成人,比较耳蜗微音电位下降组与保持组。
- 招募计划接受Cochlear Nucleus CI(Slim Straight或Slim Modiolar电极阵列)且术耳500、750、1000 Hz低频平均听阈≤80 dB HL的成人候选者,来自8个国际植入中心。
- 术前及术后随访1(4-6周)和随访2(3个月)行纯音测听,植入术中及术后即刻记录耳蜗电图。
- 共78例(80耳)入组,77例(79耳)接受手术;因方案偏差排除18耳(23%)。
- 61耳有耳蜗电图反应,其中18耳(23%)全程幅度<1 µV被归为「不明确」,43耳(55%)幅度>1 µV。
主要发现
- 43耳耳蜗微音电位>1 µV中,8耳全程稳定,35耳下降。
- 主要终点随访1:保持组7/41耳(17%)听力损失中位数12.6 dB,下降组34/41耳(83%)听力损失中位数26.9 dB(p<0.014)。
- 术中可测量耳蜗微音电位(>1 µV)的存在取决于术前低频听阈,尤其是刺激频率0.5 kHz处。
- 术前高频听阈也与术中可测量耳蜗微音电位>1 µV相关。
关键数值与推荐
- 耳蜗微音电位下降定义为电极插入期间幅度下降≥30%,无论暂时或永久。
- 可测量耳蜗微音电位信号定义为幅度>1 µV。
- 建议术前低频听阈<80 dB HL作为获得耳蜗微音电位信号>1 µV的必要条件。
临床怎么用
- 术中耳蜗内耳蜗电图可用于监测电极插入期间的耳蜗功能,耳蜗微音电位下降提示术后听力保留可能较差。
- 术前低频听阈≤80 dB HL的成人人工耳蜗候选者适合术中记录耳蜗微音电位。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: Electrocochleography (ECochG) is emerging as a tool for monitoring cochlear function during cochlear implant (CI) surgery. ECochG may be recorded directly from electrodes on the implant array intraoperatively. For low-frequency stimulation, its amplitude tends to rise or may plateau as the electrode is inserted. The aim of this study was to explore whether compromise of the ECochG signal, defined as a fall in its amplitude of 30% or more during insertion, whether transient or permanent, is associated with poorer postoperative acoustic hearing, and to examine how preoperative hearing levels may influence the ability to record ECochG. The specific hypotheses tested were threefold: (a) deterioration in the pure-tone average of low-frequency hearing at the first postoperative follow-up interval (follow-up visit 1 [FUV1], 4 to 6 weeks) will be associated with compromise of the cochlear microphonic (CM) amplitude during electrode insertion (primary hypothesis); (b) an association is observed at the second postoperative follow-up interval (FUV2, 3 months) (secondary hypothesis 1); and (c) the CM response will be recorded earlier during electrode array insertion when the preoperative high-frequency hearing is better (secondary hypothesis 2). DESIGN: International, multi-site prospective, observational, between groups design, targeting 41 adult participants in each of two groups, (compromised CM versus preserved CM). Adult CI candidates who were scheduled to receive a Cochlear Nucleus CI with a Slim Straight or a Slim Modiolar electrode array and had a preoperative audiometric low-frequency average thresholds of ≤80 dB HL at 500, 750, and 1000 Hz in the ear to be implanted, were recruited from eight international implant sites. Pure tone audiometry was measured preoperatively and at postoperative visits (FUV1 and follow-up visit 2 [FUV2]). ECochG was measured during and immediately after the implantation of the array. RESULTS: From a total of 78 enrolled individuals (80 ears), 77 participants (79 ears) underwent surgery. Due to protocol deviations, 18 ears (23%) were excluded. Of the 61 ears with ECochG responses, amplitudes were < 1 µV throughout implantation for 18 ears (23%) and deemed "unclear" for classification. EcochG responses >1 µV in 43 ears (55%) were stable throughout implantation for 8 ears and compromised in 35 ears. For the primary endpoint at FUV1, 7/41 ears (17%) with preserved CM had a median hearing loss of 12.6 dB versus 34/41 ears (83%) with compromised CM and a median hearing loss of 26.9 dB ( p < 0.014). In assessing the practicalities of measuring intraoperative ECochG, the presence of a measurable CM (>1 µV) during implantation was dependent on preoperative, low-frequency thresholds, particularly at the stimulus frequency (0.5 kHz). High-frequency, preoperative thresholds were also associated with a measurable CM > 1 µV during surgery. CONCLUSIONS: Our data shows that CM drops occurring during electrode insertion were correlated with significantly poorer hearing preservation postoperatively compared to CMs that remained stable throughout the electrode insertion. The practicality of measuring ECochG in a large cohort is discussed, regarding the suggested optimal preoperative low-frequency hearing levels ( < 80 dB HL) considered necessary to obtain a CM signal >1 µV.