人工耳蜗植入者言语感知表现与电诱发复合动作电位衍生的兴奋听神经纤维数量及同步性相关
这项回顾性研究纳入124例成人语后聋人工耳蜗植入者,术中记录电诱发复合动作电位(eCAP),用反卷积法得到复合放电潜伏期分布(CDLD),分析其8个参数与单音节词识别的关联。结果显示CDLD两个成分波幅较大、曲线下面积(AUCD)较大及AUCD增长函数斜率较陡均与较好言语感知显著相关,早期成分潜伏期方差较小也与较好言语识别显著相关。
要点
- CDLD两个成分波幅较大、AUCD较大及AUCD增长斜率较陡与较好言语感知显著相关。
- 早期CDLD成分潜伏期方差较小与较好言语识别显著相关,晚期成分无此关联。
- 言语识别与CDLD潜伏期无显著相关。
核心结论
- CDLD揭示的兴奋听神经纤维数量与神经同步性可提示人工耳蜗植入后的言语感知表现。
- 作者认为CDLD可作为一种临床预测指标,预测听神经纤维存活及术后言语感知表现。
研究设计与样本
- 回顾性研究,纳入124例成人语后聋患者,均植入Advanced Bionics HiRes 90K装置。
- 获取单音节词识别得分与术中eCAP波幅增长函数,并用反卷积法确定每个eCAP波形的CDLD。
主要发现
- CDLD两个成分波幅较大、AUCD较大、AUCD增长函数斜率较陡均与较好言语感知显著相关。
- 早期CDLD成分潜伏期方差较小与较好言语识别得分显著相关,晚期成分潜伏期方差无显著关联。
- 言语识别与CDLD潜伏期无显著相关。
- AUCD及其增长函数斜率对言语感知方差的解释度(R²)与eCAP波幅、波幅增长函数斜率、第一CDLD成分波幅及方差相似。
临床怎么用
- 作者认为CDLD可作为一种临床预测指标,用于预测听神经纤维存活及术后言语感知表现。
- 作者认为值得在未来的临床应用中把CDLD纳入eCAP测量。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
OBJECTIVES: Many studies have assessed the performance of individuals with cochlear implants (CIs) with electrically evoked compound action potentials (eCAPs). These eCAP-based studies have focused on the amplitude information of the response, without considering the temporal firing properties of the excited auditory nerve fibers (ANFs), such as neural latency and synchrony. These temporal features have been associated with neural health in animal studies and, consequently, could be of importance to clinical CI outcomes. With a deconvolution method, combined with a unitary response, the eCAP can be mathematically unraveled into the compound discharge latency distribution (CDLD). The CDLD reflects both the number and the temporal firing properties of excited ANFs. The present study aimed to determine to what extent the CDLD derived from intraoperatively recorded eCAPs is related to speech perception in individuals with CIs. DESIGN: This retrospective study acquired data on monosyllabic word recognition scores and intraoperative eCAP amplitude growth functions from 124 adult patients with postlingual deafness that received the Advanced Bionics HiRes 90K device. The CDLD was determined for each recorded eCAP waveform by deconvolution. Each of the two Gaussian components of the CDLD was described by three parameters: the amplitude, the firing latency (the average latency of each component of the CDLD), and the variance of the CDLD components (an indication of the synchronicity of excited ANFs). Apart from these six CDLD parameters, the area under the CDLD curve (AUCD) and the slope of the AUCD growth function were determined as well. The AUCD was indicative of the total number of excited ANFs over time. The slope of the AUCD growth function indicated the increases in the number of excited ANFs with stimulus level. Associations between speech perception and each of these eight CDLD-related parameters were investigated with linear mixed modeling. RESULTS: In individuals with CIs, larger amplitudes of the two CDLD components, greater AUCD, and steeper slopes of the AUCD growth function were all significantly associated with better speech perception. In addition, a smaller latency variance in the early CDLD component, but not in the late, was significantly associated with better speech recognition scores. Speech recognition was not significantly dependent on CDLD latencies. The AUCD and the slope of the AUCD growth function provided a similar explanation of the variance in speech perception (R 2 ) as the eCAP amplitude, the slope of the amplitude growth function, the amplitude, and variance of the first CDLD component. CONCLUSION: The results demonstrate that both the number and the neural synchrony of excited ANFs, as revealed by CDLDs, are indicative of postimplantation speech perception in individuals with a CI. Because the CDLD-based parameters yielded a higher significance than the eCAP amplitude or the AGF slope, the authors conclude that CDLDs can serve as a clinical predictor of the survival of ANFs and that they have predictive value for postoperative speech perception performance. Thus, it would be worthwhile to incorporate the CDLD into eCAP measures in future clinical applications.
引用
- Dong Y, Briaire JJ, Stronks HC, et al. Speech Perception Performance in Cochlear Implant Recipients Correlates to the Number and Synchrony of Excited Auditory Nerve Fibers Derived From Electrically Evoked Compound Action Potentials. Ear Hear. 2022. PMID: 36253905. doi:10.1097/AUD.0000000000001279
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。